Design of a high-resolution small-animal SPECT-CT system sharing a CdTe semiconductor detector

被引:2
|
作者
Ryu, Hyun-Ju [1 ]
Lee, Young-Jin
Lee, Seung-Wan
Cho, Hyo-Min
Choi, Yu-Na
Kim, Hee-Joung
机构
[1] Yonsei Univ, Dept Radiol Sci, Wonju 220710, South Korea
基金
新加坡国家研究基金会;
关键词
CdTe detector; SPECT-CT; Pinhole collimator;
D O I
10.3938/jkps.61.130
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
A single photon emission computed tomography (SPECT) system with a co-registered X-y computed tomography (CT) system allows the convergence of functional information and morphologic information. The localization of radiopharmaceuticals on a SPECT can be enhanced by combining the SPECT with an anatomical modality, such as X-ray CT. Gamma-ray imaging for nuclear medicine devices and X-ray imaging systems for diagnostics has recently been developed based on semiconductor detectors, and semiconductor detector materials such as cadmium telluride (CdTe) or cadmium zinc telluride (CZT) are available for both X-ray and gamma-ray systems for small-animal imaging. CdTe or CZT detectors provide strong absorption and high detection efficiency of high energy X-ray and gamma-ray photons because of their large atomic numbers. In this study, a pinhole collimator SPECT system sharing a cadmium telluride (CdTe) detector with a CT was designed. The GEANT4 application for tomographic emission (GATE) v.6.1 was used for the simulation. The pinhole collimator was designed to obtain a high spatial resolution of the SPECT system. The acquisition time for each projection was 40 seconds, and 60 projections were obtained for tomographic image acquisition. The reconstruction was performed using ordered subset expectation maximization (OS-EM) algorithms. The sensitivity and the spatial resolution were measured on the GATE simulation to evaluate the system characteristics. The spatial resolution of the system calculated from the FWHM of Gaussian fitted PSF curve was 0.69 mm, and the sensitivity of the system was measured to be 0.354 cps/kBq by using a Tc-99m point source of 1 MBq for 800 seconds. A phantom study was performed to verify the design of the dual imaging modality system. The system will be built as designed, and it can be applied as a pre-clinical imaging system.
引用
收藏
页码:130 / 134
页数:5
相关论文
共 50 条
  • [41] Comparison between pixel array CdTe detector and NaI(Tl) detector for ultra-high resolution small animal SPECT imaging
    Park, S.
    Lee, C.
    Kim, H.
    [J]. MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [42] HIGH-RESOLUTION 4-HEAD PINHOLE SPECT SYSTEM FOR SMALL ANIMAL STUDIES
    ISHIZU, K
    MUKAI, T
    YONEKURA, Y
    FUJITA, T
    MAGATA, Y
    NISHIZAWA, S
    TAMAKI, N
    SHIBASAKI, H
    KONISHI, J
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (05): : P196 - P196
  • [43] Ultra-high-resolution SPECT with CdTe for small-animal imaging applications: A Monte Carlo simulation study using a voxelized phantom
    Su-Jin Park
    Chang-Lae Lee
    Hyo-Min Cho
    Hee-Joung Kim
    [J]. Journal of the Korean Physical Society, 2012, 60 : 1145 - 1149
  • [44] Quasi-continous detector module design for high-resolution small animal PET
    Correia, JA
    Burnham, CA
    Kaufman, D
    Fischman, A
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2063 - 2066
  • [45] FastSPECT II: First static and dynamic images from a new high-resolution small-animal SPECT imager.
    Furenlid, LR
    Chen, Y
    Wilson, DW
    Lehovich, A
    Kim, H
    Stevenson, G
    Fesler, L
    Liu, Z
    Patton, DD
    Barrett, HH
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05): : 161P - 162P
  • [46] Ultra-high-resolution SPECT with CdTe for Small-animal Imaging Applications: A Monte Carlo Simulation Study Using A Voxelized Phantom
    Park, Su-Jin
    Lee, Chang-Lae
    Cho, Hyo-Min
    Kim, Hee-Joung
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 60 (07) : 1145 - 1149
  • [47] A Prototype Ultra-High-Resolution Small-Animal PET System
    Liu, Jiguo
    Liu, Qiong
    Chen, Jyh-Cheng
    Wang, Deguang
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [48] Animals In Synchrotrons: Overcoming Challenges For High-Resolution, Live, Small-Animal Imaging
    Donnelley, Martin
    Parsons, David
    Morgan, Kaye
    Siu, Karen
    [J]. 6TH INTERNATIONAL CONFERENCE ON MEDICAL APPLICATIONS OF SYNCHROTRON RADIATION, 2010, 1266 : 30 - +
  • [49] Simulation of HyperSPECT: a high-resolution small-animal system with in-line x-ray optics
    Tibbelin, Sandra
    Nillius, Peter
    Danielsson, Mats
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (06): : 1617 - 1629
  • [50] State-of-the-art of small animal imaging with high-resolution SPECT
    Nikolaus, S
    Wirrwar, A
    Antke, C
    Kley, K
    Müller, HW
    [J]. NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2005, 44 (06): : 257 - 266