Dose response of selected ion chambers in applied homogeneous transverse and longitudinal magnetic fields

被引:95
|
作者
Reynolds, M. [1 ]
Fallone, B. G. [2 ,3 ,4 ]
Rathee, S. [1 ,2 ]
机构
[1] Univ Alberta, Dept Oncol, Med Phys Div, Edmonton, AB T6G 1Z2, Canada
[2] Cross Canc Inst, Dept Med Phys, Edmonton, AB T6G 1Z2, Canada
[3] Univ Alberta, Dept Oncol, Edmonton, AB T6G 1Z2, Canada
[4] Univ Alberta, Dept Phys, Edmonton, AB T6G 1Z2, Canada
关键词
Monte Carlo; Monte Carlo methods; linac-MR; longitudinal and transverse linac-MR; linear accelerators; biomedical MRI; magnetic field; dosimetry; ionization chambers; MV RADIOTHERAPY ACCELERATOR; MONTE-CARLO-SIMULATION; MRI SCANNER; REFERENCE DOSIMETRY; IONIZATION-CHAMBER; PHOTON; ELECTRON; SYSTEM; LINAC; DEPOSITION;
D O I
10.1118/1.4794496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The magnetic fields of an integrated MR-Linac system will alter the paths of electrons that produce ions in the ionization chambers. The dose response of selected ion chambers is evaluated in the presence of varying transverse and longitudinal magnetic fields. The investigation is useful in calibration of therapeutic x-ray beams associated with MR-Linac systems. Methods: The Monte Carlo code PENELOPE was used to model the irradiation of NE2571, and PR06C ionization chambers in the presence of a transverse and longitudinal (with respect to the photon beam) magnetic fields of varying magnitude. The long axis of each chamber was simulated both parallel and perpendicular to the incident photon beam for each magnetic field case. The dose deposited in each chamber for each case was compared to the case with zero magnetic field by means of a ratio. The PR06C chamber's response was measured in the presence of a transverse magnetic field with field strengths ranging from 0.0 to 0.2 T to compare to simulated results. Results: The simulations and measured data show that in the presence of a transverse magnetic field there is a considerable dose response (maximum of 11% near 1.0 T in the ion chambers investigated, which depends on the magnitude of magnetic field, and relative orientation of the magnetic field, radiation beam, and ion chamber. Measurements made with the PR06C chamber verify these results in the region of measurement. In contrast, a longitudinal magnetic field produces only a slight increase in dose response (2% at 1.5 T) that rises slowly with increasing magnetic field and is seemingly independent of chamber orientation. Response trends were similar for the two ion chambers and relative orientations considered, but slight variations are present from chamber to chamber. Conclusions: Care must be taken when making ion chamber measurements in a transverse magnetic field. Ion chamber responses vary not only with transverse field strength, but with chamber orientation and type, and can be considerable. Longitudinal magnetic fields influence ion chamber responses relatively little (2% at 1.5 T), and only at field strengths in excess of 1.0 T. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4794496]
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Dose response of selected solid state detectors in applied homogeneous transverse and longitudinal magnetic fields
    Reynolds, M.
    Fallone, B. G.
    Rathee, S.
    MEDICAL PHYSICS, 2014, 41 (09)
  • [2] Dose Response of Common Solid State Detectors in Homogeneous Transverse and Longitudinal Magnetic Fields
    Reynolds, M.
    Fallone, B.
    Rathee, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 166 - 166
  • [3] Radiation Detector Responses to Applied Homogeneous Transverse and Parallel Magnetic Fields
    Reynolds, M.
    Rathee, S.
    Vidakovic, S.
    Fallone, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3704 - 3705
  • [4] Impact of Transverse Magnetic Fields On Dose Response of a NanoDot OSLD
    Ito, S.
    Araki, F.
    Ohno, T.
    MEDICAL PHYSICS, 2019, 46 (06) : E240 - E241
  • [5] Impact of magnetic fields on dose measurement with small ion chambers illustrated in high-resolution response maps
    Lehmann, Joerg
    Beveridge, Toby
    Oliver, Chris
    Bailey, Tracy E.
    Lye, Jessica E.
    Livingstone, Jayde
    Stevenson, Andrew W.
    Butler, Duncan J.
    MEDICAL PHYSICS, 2019, 46 (07) : 3298 - 3305
  • [6] Perturbation effect of parallel-plate ionization chambers on buildup dose measurements in transverse magnetic fields
    Matsuoka, Takanori
    Araki, Fujio
    Ohno, Takeshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 59 : 112 - 116
  • [7] Radiation Detector Responses to Applied Homogeneous Magnetic Fields
    Reynolds, M.
    Fallone, B.
    Rathee, S.
    Vidakovic, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [8] Response of Parallel-Plate Ionization Chambers On the Buildup Dose Measurement in Inline Magnetic Fields
    Anami, N.
    Araki, F.
    Ohno, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E844 - E844
  • [9] Impact of transverse magnetic fields on dose response of a nanoDot OSLD in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Hoshida, Kento
    Ohno, Takeshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 153 - 160
  • [10] Simulation of MR response to transverse magnetic fields
    Fujimoto, K
    Totsuka, K
    Uesaka, Y
    Saitoh, I
    Fukui, H
    Akagi, F
    Yoshihara, I
    Nakatani, Y
    Hayashi, N
    IEEE TRANSACTIONS ON MAGNETICS, 1997, 33 (03) : 2386 - 2391