Robotic hip joint testing: Development and experimental protocols

被引:11
|
作者
El Daou, Hadi [1 ]
Ng, K. C. Geoffrey [1 ]
Van Arkel, Richard [1 ]
Jeffers, Jonathan R. T. [1 ]
Rodriguez y Baena, Ferdinando [1 ]
机构
[1] Imperial Coll London, Dept Mech Engn, South Kensington Campus, London SW7 2AZ, England
基金
英国工程与自然科学研究理事会;
关键词
Robotics; Biomechanics; In vitro; Hip; Optical tracking; LIGAMENT RECONSTRUCTION; ANTEROLATERAL LIGAMENT; PART; SYSTEM; ARTHROPLASTY; BIOMECHANICS; CAPSULE; LAXITY; MOTION;
D O I
10.1016/j.medengphy.2018.10.006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of robotic systems combined with force sensing is emerging as the gold standard for in vitro biomechanical joint testing, due to the advantage of controlling all six degrees of freedom independently of one another. This paper describes a novel robotic platform and the experimental protocol used for hip joint testing. An experimental protocol implemented optical tracking and registration techniques in order to define the position of the hip joint centre of rotation (COR) in the coordinate system of the robot's end effector. The COR coordinates defined the origin of the task-related coordinate system used to control the robot, with a hybrid force/position law to simulate standard clinical tests. The axes of this frame were defined using the International Society of Biomechanics (ISB) anatomical coordinate system. Experiments were carried out on two cadaveric hip joint specimens using the robotic testing platform and a mechanical testing rig previously developed and described by our group. Simulated internal-external and adduction/abduction laxity tests were carried out with both systems and the resulting peak range of motion (ROM) was measured. Similarities and differences were observed in these experiments, which were used to highlight some of the limitations of conventional systems and the corresponding advantages of robotics, further emphasising their added value in vitro testing. (C) 2018 The Authors. Published by Elsevier Ltd on behalf of IPEM.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 50 条
  • [1] Development and validation of a robotic system for ankle joint testing
    El Daou, Hadi
    Calder, James D.
    Stephen, Joanna M.
    MEDICAL ENGINEERING & PHYSICS, 2018, 62 : 53 - 57
  • [2] EXPERIMENTAL TESTING OF HUMAN HIP-PROSTHESIS IN THE SHOULDER JOINT OF THE DOG
    KUPPER, W
    ZEITSCHRIFT FUR VERSUCHSTIERKUNDE, 1982, 24 (1-2): : 46 - 46
  • [3] DEVELOPMENT OF THE HIP JOINT
    GRAY, DJ
    ANATOMICAL RECORD, 1949, 103 (03): : 460 - 460
  • [5] RESULTS OF THE EXPERIMENTAL TESTING OF A SCREWABLE POLYETHYLENE HIP-JOINT SOCKET IN SHEEP
    ENDLER, M
    PLENK, H
    GRUNDSCHOBER, F
    GIRTLER, D
    SCHNABL, H
    ACTA MEDICA AUSTRIACA, 1982, 9 (03) : 26 - 27
  • [6] Development of loading protocols for experimental testing of double skin composite wall
    Wang, Shiye
    Wang, Wei
    JOURNAL OF BUILDING ENGINEERING, 2022, 51
  • [7] A Surgical Robotic System for Osteoporotic Hip Augmentation: System Development and Experimental Evaluation
    Bakhtiarinejad, Mahsan
    Gao, Cong
    Farvardin, Amirhossein
    Zhu, Gang
    Wang, Yu
    Oni, Julius K.
    Taylor, Russell H.
    Armand, Mehran
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2023, 5 (01): : 18 - 29
  • [8] Development of collapse-consistent loading protocols for experimental testing of steel columns
    Suzuki, Yusuke
    Lignos, Dimitrios G.
    EARTHQUAKE ENGINEERING & STRUCTURAL DYNAMICS, 2020, 49 (02): : 114 - 131
  • [9] THE DEVELOPMENT OF OSTEOARTHRITIS OF THE HIP JOINT
    PRIDIE, KH
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1952, 34 (01): : 153 - 153
  • [10] DEVELOPMENT OF HIP-JOINT
    ENGELHARDT, P