A new G-M counter hybrid dead-time correction model

被引:7
|
作者
Hou, Guojing [1 ]
Gardner, Robin P. [1 ]
机构
[1] N Carolina State Univ, Dept Nucl Engn, Ctr Engn Applicat Radioisotopes, Raleigh, NC 27695 USA
关键词
G-M counter; Dead-time; Paralyze;
D O I
10.1016/j.radphyschem.2015.05.014
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
A new G-M counter dead-time model based on the hybrid model is proposed and studied. The new model contains the idealized paralyzable and non-paralyzable dead-time models, but also allows the non-paralyzable dead-time to vary with the true counting rate. To validate this model, the decaying source experiment with a very pure Mn-56 source was used. The calculated true counting rates obtained by fitting the experimental data to the model were compared to the data obtained by table lookup. In general, the fitting was in good agreement (within 4.8% up to 5.35 x 10(6) counts/minute) for a G-M counter with a paralyzable dead-time component of 85.9 mu s and a non-paralyzable dead-time component ranging from 438 mu s to 24.3 ms. (c) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:125 / 129
页数:5