Assessment of the influence of display veiling glare on observer and model performance

被引:0
|
作者
Krupinski, Elizabeth A. [1 ]
Lubin, Jeffrey [2 ]
Roehrig, Hans [1 ]
Johnson, Jeffrey [3 ]
Nafziger, John [3 ]
机构
[1] Univ Arizona, Dept Radiol, 1609 N Warren Ave, Tucson, AZ 85724 USA
[2] Sarnoff Corp, Princeton, NJ 80540 USA
[3] Siemens Corp Res, Princeton, NJ 80540 USA
来源
MEDICAL IMAGING 2006: IMAGE PERCEPTION, OBSERVER PERFORMANCE AND TECHNOLOGY ASSESSMENT | 2006年 / 6146卷
关键词
veiling glare; physical characterization; observer performance;
D O I
10.1117/12.650677
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We evaluated human observer and model (JNDmetrix) performance to assess whether the veiling glare of a digital display influences performance in softcopy interpretation of mammographic images. 160 mammographic images, half with a single mass, were processed to simulate four levels of veiling glare: none, comparable to a typical cathode ray tube (CRT) display, double a CRT and quadruple a CRT. Six radiologist observers were shown the images in a randomized presentation order on a liquid crystal display (LCD) that had relatively no veiling glare. The JNDmetrix human visual system model also analyzed the images. Receiver Operating Characteristic (ROC) techniques showed that performance declined with increasing veiling glare (F = 6.884, p = 0.0035). Quadruple veiling glare yielded significantly lower performance than the lower veiling glare levels. The JNDmetrix model did not predict a reduction in performance with changes in veiling glare, and correlation with the human observer data was modest (0.588). Display veiling glare may influence observer performance, but only at very high levels.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Contrast and glare testing in the assessment of visual performance of candidate eyes for penetrating keratoplasty
    Fontaine, N
    Boisjoly, H
    Gresset, J
    Charest, M
    Brunette, I
    Le François, M
    Deschênes, J
    Ponomarenko, S
    CORNEA, 2000, 19 (04) : 433 - 438
  • [42] Performance model for tempered anti-glare LCD cover glass
    Jeongju Choi
    Journal of Mechanical Science and Technology, 2019, 33 : 4951 - 4956
  • [43] Correlation between human observer performance and model observer performance in differential phase contrast CT
    Li, Ke
    Garrett, John
    Chen, Guang-Hong
    MEDICAL PHYSICS, 2013, 40 (11)
  • [44] Performance model for tempered anti-glare LCD cover glass
    Choi, Jeongju
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2019, 33 (10) : 4951 - 4956
  • [45] An interactive method of assessing the characteristics of softcopy display using observer performance tests
    Wang, JH
    Peng, Q
    MEDICAL IMAGING 2002: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2002, 4686 : 189 - 197
  • [46] Three Digital Mammography Display Configurations: Observer Performance in a Pilot ROC Study
    Tabar, Laszlo K.
    Nino, Alfredo S.
    Schreiber, Eric
    Ellis, Richard
    Goralnik, Cynthia
    Scislaw, Steven
    Cheruvu, Raja
    Masrani, Deepa
    Zhang, Daoxian
    Muller, Steve
    Heffernan, Patrick
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 280 - +
  • [47] Assessment of flux observer performance for induction motor drive
    Luiz, ASA
    Silva, SR
    Menezes, BR
    IECON '97 - PROCEEDINGS OF THE 23RD INTERNATIONAL CONFERENCE ON INDUSTRIAL ELECTRONICS, CONTROL, AND INSTRUMENTATION, VOLS. 1-4, 1997, : 494 - 499
  • [48] Assessment of flux observer performance for induction motor drive
    Dep Acad de Eng. Eletrica CEFET-MG, Belo Horizonte, Brazil
    IECON Proceedings (Industrial Electronics Conference), 1997, 2 : 494 - 499
  • [49] Influence of observer position on visual optical instrument performance
    Jozwicki, R
    OPHTHALMIC MEASUREMENTS AND OPTOMETRY, 1998, 3579 : 40 - 47