Incidence and predictors of phantom shocks in implantable cardioverter defibrillator recipients

被引:21
|
作者
Kraaier, K. [1 ]
Starrenburg, A. H. [2 ]
Verheggen, R. M. [1 ]
van der Palen, J. [3 ,4 ]
Scholten, M. F. [1 ]
机构
[1] Med Spectrum Twente, Thoraxctr Twente, Dept Cardiol, NL-7513 ER Enschede, Netherlands
[2] Med Spectrum Twente, Dept Psychol, NL-7513 ER Enschede, Netherlands
[3] Med Spectrum Twente, Dept Epidemiol, NL-7513 ER Enschede, Netherlands
[4] Univ Twente, Dept Res Methodol Methods & Data Anal, NL-7500 AE Enschede, Netherlands
关键词
Implantable defibrillator; Primary prevention; Secondary prevention; Phantom shock; Psychological consequences; ATRIAL-FIBRILLATION; SYMPTOMS;
D O I
10.1007/s12471-012-0345-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Implantable cardioverter defibrillators (ICDs) are designed to deliver shocks or antitachycardia pacing (ATP) in the event of ventricular arrhythmias. During follow-up, some ICD recipients experience the sensation of ICD discharge in the absence of an actual discharge (phantom shock). The aim of this study was to evaluate the incidence and predictors of phantom shocks in ICD recipients. Medical records of 629 consecutive patients with ischaemic or dilated cardiomyopathy and prior ICD implantation were studied. With a median follow-up of 35 months, phantom shocks were reported by 5.1 % of ICD recipients (5.7 % in the primary prevention group and 3.7 % for the secondary prevention group; p=NS). In the combined group of primary and secondary prevention, there were no significant predictors of the occurrence of phantom shocks. However, in the primary prevention group, phantom shocks were related to a history of atrial fibrillation (p=0.03) and NYHA class < III (p=0.05). In the secondary prevention group, there were no significant predictors for phantom shocks. Phantom shocks occur in approximately 5 % of all ICD recipients. In primary prevention patients, a relation with a history of atrial fibrillation and NYHA class < III were significant predictors for the occurrence of phantom shocks. In the secondary prevention patients, no significant predictors were found.
引用
收藏
页码:191 / 195
页数:5
相关论文
共 50 条
  • [21] Incidence of and predictors for appropriate implantable cardioverter-defibrillator therapy in patients with a secondary preventive implantable cardioverter-defibrillator indication
    Schaer, Beat
    Kuehne, Michael
    Reichlin, Tobias
    Osswald, Stefan
    Sticherling, Christian
    EUROPACE, 2016, 18 (02): : 227 - 231
  • [22] CLINICAL PREDICTORS OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SHOCKS (RESULTS OF THE CASCADE TRIAL)
    DOLACK, GL
    GREENE, HL
    POOLE, JE
    KUDENCHUK, PJ
    BARDY, GH
    COBB, LA
    GRAHAMRENFROE, EL
    POWELL, JL
    GALLOWAY, AC
    KELLIE, J
    FELLOWS, CL
    MAIN, CL
    MCMAHONBUSCH, M
    BROUDY, DR
    SANDERS, J
    GARNI, JE
    GERITY, D
    MAYNARD, C
    HALLSTROM, AP
    MCBRIDE, R
    AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (04): : 237 - 241
  • [23] Experiences of shocks in patients with implantable cardioverter defibrillator
    Stromberg, A.
    Hammargard, J.
    Fant, D.
    Johansson, I.
    EUROPEAN HEART JOURNAL, 2007, 28 : 390 - 390
  • [24] Biventricular implantable cardioverter defibrillator and inappropriate shocks
    Srivathsan, K
    Bazzell, JL
    Lee, RW
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (01) : 88 - 89
  • [25] Inappropriate implantable cardioverter defibrillator shocks—incidence, effect, and implications for driver licensing
    Eiichi Watanabe
    Katsunori Okajima
    Akira Shimane
    Tomoya Ozawa
    Tetsuyuki Manaka
    Itsuro Morishima
    Toru Asai
    Masahiko Takagi
    Toshihiro Honda
    Atsunobu Kasai
    Eitaro Fujii
    Kohei Yamashiro
    Ritsuko Kohno
    Haruhiko Abe
    Takashi Noda
    Takashi Kurita
    Shigeyuki Watanabe
    Hiroya Ohmori
    Takashi Nitta
    Yoshifusa Aizawa
    Ken Kiyono
    Ken Okumura
    Journal of Interventional Cardiac Electrophysiology, 2017, 49 : 271 - 280
  • [26] Audit of implantable cardioverter defibrillator (ICD) shocks
    Llewellyn, Chris
    Chieng, Vui Yung
    Affolter, Jonathan
    SCOTTISH MEDICAL JOURNAL, 2016, 61 (04) : NP39 - NP40
  • [27] Implantable cardioverter defibrillator shocks induced by showering
    Spurrell, P.
    Patel, N.
    Sulke, N.
    INTERNATIONAL JOURNAL OF CLINICAL PRACTICE, 2008, 62 (03) : 505 - 507
  • [28] INCIDENCE OF LOSS OF CONSCIOUSNESS DURING AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SHOCKS
    KOU, WH
    CALKINS, H
    LEWIS, RR
    BOLLING, SF
    KIRSCH, MM
    LANGBERG, JJ
    DEBUITLEIR, M
    SOUSA, J
    ELATASSI, R
    MORADY, F
    ANNALS OF INTERNAL MEDICINE, 1991, 115 (12) : 942 - 945
  • [29] Incidence, Risk Factors and Predictors of Phantom Shocks in Patients with Implantable Cardioverter Defibrillators: State-of-the-art Review
    Elenizi, Khaled
    Alharthi, Rasha
    ARRHYTHMIA & ELECTROPHYSIOLOGY REVIEW, 2024, 13
  • [30] Phantom Shocks and Automated Implantable Cardioverter Defibrillators
    Lundberg, Andrea B.
    Bowen, Kerstin P.
    Baumgart, Patrick M.
    Caplan, Jason P.
    PSYCHOSOMATICS, 2015, 56 (01) : 94 - 97