Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study

被引:24
|
作者
Herrero-Climent, Mariano [1 ]
Lemos, Bernardo Ferreira [1 ,2 ]
Herrero-Climent, Federico [3 ]
Falcao, Carlos [1 ,2 ]
Oliveira, Helder [2 ]
Herrera, Manuela [4 ]
Javier Gil, Francisco [5 ]
Rios-Carrasco, Blanca [6 ]
Rios-Santos, Jose-Vicente [6 ]
机构
[1] Porto Dent Inst, P-4150518 Porto, Portugal
[2] Fernando Pessoa Univ, Fac Hlth Sci, P-4249004 Porto, Portugal
[3] Univ Complutense Madrid, Dept Periodontol, Madrid 28040, Spain
[4] Univ Seville, Dept Stomatol, Seville 41009, Spain
[5] Univ Int Catalunya, Fac Dent, Barcelona 08017, Spain
[6] Univ Seville, Dept Periodontol, Seville 41009, Spain
关键词
dental implant; tapered; under-preparation; implant preparation; implant stability; insertion torque; ISQ; RFA; RESONANCE FREQUENCY-ANALYSIS; LOW-DENSITY BONE; DENTAL IMPLANTS; SURGICAL TECHNIQUE; SURFACE-ROUGHNESS; STEPPED OSTEOTOMY; INSERTION TORQUE; HUMAN CADAVERS; OSSEOINTEGRATION; RELIABILITY;
D O I
10.3390/ijerph17124436
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this study was to evaluate the effects of different implant sites an under-preparation sequence associated with two different implant designs on implant primary stability measured by two parameters: insertion torque (IT) and implant stability quotient (ISQ). It used two different implants: one cylindrical as a control and another one with a tapered design. The implants were inserted in type III fresh humid bovine bone and four drilling sequences were used: one control, the one proposed by the implant company (P1), and three different undersized (P2, P3 and P4). P2 was the same as P1 without the cortical drill, P3 was without the last pilot drill and P4 was without both of them. The sample size wasn= 40 for each of the eight groups. Final IT was measured with a torquemeter and the ISQ was measured with Penguin resonance frequency analysis. Results showed that both ISQ and IT have a tendency to increase as the preparation technique reduces the implant site diameter when compared with the standard preparation, P1. The preparations without cortical drill, P2 and P4, showed the best results when compared with the ones with a cortical drill. Tapered implants always showed higher or the same ISQ and IT values when compared with the cylindrical implants. Giving the limitations of this study, it can be concluded that reducing implant preparation can increase IT and ISQ values. Removing the cortical drill and the use of a tapered design implant are also effective methods of increasing primary implant stability.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 50 条
  • [21] Macrogeometric Differentiation of Dental Implant Primary Stability: An In Vitro Study
    Al-Tarawneh, Sandra K.
    Thalji, Ghadeer
    Cooper, Lyndon F.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (06) : 1110 - 1118
  • [22] Effect of bone quality, implant design, and surgical technique on primary implant stability
    Karl, Matthias
    Grobecker-Karl, Tanja
    QUINTESSENCE INTERNATIONAL, 2018, 49 (03): : 189 - 198
  • [23] Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro
    Sakoh, Jun
    Wahlmann, Ulrich
    Stender, Elmar
    Al-Nawas, Bilal
    Wagner, Wilfried
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2006, 21 (04) : 560 - 566
  • [24] Evaluation of Implant Stability and Trephination Depth for Implant Removal-An In Vitro Study
    Meir, Haya
    Sebaoun, Alon
    Raz, Perry
    Levartovsky, Shifra
    Arieli, Adi
    Pilo, Raphel
    Dor, Zafar
    Beitlitum, Ilan
    MATERIALS, 2022, 15 (12)
  • [25] Ultrasonic implant site preparation vs. conventional drilling: in-vitro evaluation of implant site diameter
    Lahoud, Pierre
    Nasr, Joseph Abi
    Menhall, Abdallah
    Rebaudi, Alberto
    Makary, Christian
    CHIRURGIA-ITALY, 2023, 37 (04): : 224 - 231
  • [26] Primary implant stability of two implant macro-designs in different alveolar ridge morphologies: an in vitro study
    Jenner, Anna
    Sabatini, Gabriela P.
    Abou-Ayash, Samir
    Couso-Queiruga, Emilio
    Chappuis, Vivianne
    Raabe, Clemens
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2025, 11 (01)
  • [27] The effects of surgical preparation techniques and implant macro-geometry on primary stability: An in vitro study
    Falisi, Giovanni
    Severino, Marco
    Rastelli, Claudio
    Bernardi, Sara
    Caruso, Silvia
    Galli, Massimo
    Lamazza, Luca
    Di Paolo, Carlo
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2017, 22 (02): : E201 - E206
  • [28] Influence of simulated bone-implant contact and implant diameter on secondary stability: a resonance frequency in vitro study
    Veltri, Mario
    Gonzalez-Martin, Oscar
    Belser, Urs C.
    CLINICAL ORAL IMPLANTS RESEARCH, 2014, 25 (08) : 899 - 904
  • [29] In Vitro Assessment of Primary Stability of Straumann® Implant Designs
    Romanos, Georgios E.
    Ciornei, Gabriela
    Jucan, Adina
    Malmstrom, Hans
    Gupta, Bhumija
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2014, 16 (01) : 89 - 95
  • [30] Effect of implant neck design on primary and secondary implant stability in the posterior maxilla: A prospective randomized controlled study
    Eshkol-Yogev, Inbar
    Tandlich, Moshik
    Shapira, Lior
    CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (12) : 1220 - 1228