Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study

被引:24
|
作者
Herrero-Climent, Mariano [1 ]
Lemos, Bernardo Ferreira [1 ,2 ]
Herrero-Climent, Federico [3 ]
Falcao, Carlos [1 ,2 ]
Oliveira, Helder [2 ]
Herrera, Manuela [4 ]
Javier Gil, Francisco [5 ]
Rios-Carrasco, Blanca [6 ]
Rios-Santos, Jose-Vicente [6 ]
机构
[1] Porto Dent Inst, P-4150518 Porto, Portugal
[2] Fernando Pessoa Univ, Fac Hlth Sci, P-4249004 Porto, Portugal
[3] Univ Complutense Madrid, Dept Periodontol, Madrid 28040, Spain
[4] Univ Seville, Dept Stomatol, Seville 41009, Spain
[5] Univ Int Catalunya, Fac Dent, Barcelona 08017, Spain
[6] Univ Seville, Dept Periodontol, Seville 41009, Spain
关键词
dental implant; tapered; under-preparation; implant preparation; implant stability; insertion torque; ISQ; RFA; RESONANCE FREQUENCY-ANALYSIS; LOW-DENSITY BONE; DENTAL IMPLANTS; SURGICAL TECHNIQUE; SURFACE-ROUGHNESS; STEPPED OSTEOTOMY; INSERTION TORQUE; HUMAN CADAVERS; OSSEOINTEGRATION; RELIABILITY;
D O I
10.3390/ijerph17124436
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this study was to evaluate the effects of different implant sites an under-preparation sequence associated with two different implant designs on implant primary stability measured by two parameters: insertion torque (IT) and implant stability quotient (ISQ). It used two different implants: one cylindrical as a control and another one with a tapered design. The implants were inserted in type III fresh humid bovine bone and four drilling sequences were used: one control, the one proposed by the implant company (P1), and three different undersized (P2, P3 and P4). P2 was the same as P1 without the cortical drill, P3 was without the last pilot drill and P4 was without both of them. The sample size wasn= 40 for each of the eight groups. Final IT was measured with a torquemeter and the ISQ was measured with Penguin resonance frequency analysis. Results showed that both ISQ and IT have a tendency to increase as the preparation technique reduces the implant site diameter when compared with the standard preparation, P1. The preparations without cortical drill, P2 and P4, showed the best results when compared with the ones with a cortical drill. Tapered implants always showed higher or the same ISQ and IT values when compared with the cylindrical implants. Giving the limitations of this study, it can be concluded that reducing implant preparation can increase IT and ISQ values. Removing the cortical drill and the use of a tapered design implant are also effective methods of increasing primary implant stability.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 50 条
  • [1] Enhancing primary implant stability by undersizing implant site preparation: A human cadaver study
    Turkyilmaz, I
    Tozum, T. F.
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2020, 121 (01) : 58 - 62
  • [2] Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation
    Antonelli, Alessandro
    Barone, Selene
    Attanasio, Ferdinando
    Salviati, Marianna
    Cerra, Maria Giulia
    Calabria, Elena
    Bennardo, Francesco
    Giudice, Amerigo
    DENTISTRY JOURNAL, 2023, 11 (10)
  • [3] Implant Site Under-Preparation to Compensate the Remodeling of an Autologous Bone Block Graft
    Anitua, Eduardo
    Murias-Freijo, Alia
    Alkhraisat, Mohammad Hamdan
    JOURNAL OF CRANIOFACIAL SURGERY, 2015, 26 (05) : E374 - E377
  • [4] Analysis of the relationship between a modified method for implant site preparation and primary implant stability: A pilot study
    Unsal, Gokce Soganci
    Turkyilmaz, Ilser
    JOURNAL OF DENTAL SCIENCES, 2020, 15 (02) : 234 - 235
  • [5] Efficacy of biologically guided implant site preparation to obtain adequate primary implant stability
    Anitua, Eduardo
    Alkhraisat, Mohammad Hamdan
    Pinas, Laura
    Orive, Gorka
    ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2015, 199 : 9 - 15
  • [6] In-vitro investigation of initial implant stability.
    Strong, JT
    Greer, WD
    Boggan, RS
    George, PD
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 3301 - 3301
  • [7] Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study
    Romero, Marta
    Herrero-Climent, Mariano
    Rios-Carrasco, Blanca
    Brizuela, Aritza
    Romero, Manuel Maria
    Gil, Javier
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (13)
  • [8] Undersized Implant Site Preparation to Enhance Primary Implant Stability in Poor Bone Density: A Prospective Clinical Study
    Alghamdi, Hamdan
    Anand, Pradeep S.
    Anil, Sukumaran
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (12) : E506 - E512
  • [9] Piezoelectric vs. conventional implant site preparation: ex vivo implant primary stability
    Baker, Justin A.
    Vora, Shweta
    Bairam, Latifa
    Kim, Hyeong-il
    Davis, Elaine L.
    Andreana, Sebastiano
    CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (04) : 433 - 437
  • [10] DIFFERENCES IN IMPLANT STABILITY ASSOCIATED WITH VARIOUS METHODS OF PREPARATION OF THE IMPLANT BED: AN IN VITRO STUDY
    Ahn, Su-Jin
    Leesungbok, Richard
    Lee, Suk-Won
    Heo, Young-Ku
    Kang, Kyung Lhi
    JOURNAL OF PROSTHETIC DENTISTRY, 2012, 107 (06): : 366 - 372