Arterial wall histology in chronic pulsatile-flow and continuous-flow device circulatory support

被引:26
|
作者
Potapov, Evgenij V. [1 ]
Dranishnikov, Nikolay [1 ]
Morawietz, Lars [1 ]
Stepanenko, Alexander [1 ]
Rezai, Sajjad [1 ]
Blechschmidt, Cristiane [2 ]
Lehmkuhl, Hans B. [1 ]
Weng, Yuguo [1 ]
Pasic, Miralem [1 ]
Huebler, Michael [1 ]
Hetzer, Roland [1 ]
Krabatsch, Thomas [1 ]
机构
[1] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, D-13353 Berlin, Germany
[2] Univ Hosp Charite, Inst Neuropathol, Berlin, Germany
来源
关键词
Ventricular assist device; long-term support; continuous flow; pulsatile flow; pathology; arterial wall; LEFT-VENTRICULAR ASSIST; END-ORGAN FUNCTION; DESTINATION THERAPY; SYSTEM; PUMP;
D O I
10.1016/j.healun.2012.08.013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Continuous-flow (CF) ventricular assist devices (VAD) are an established option for treatment of end-stage heart failure. However, the effect of long-term CF with lack of peripheral arterial wall motions on blood pressure regulation and end-organ arterial wall sclerosis, especially in the case of long-term support (> 3 years), remains unclear. METHODS: Tissue samples obtained at autopsy from liver, kidney, coronary arteries, and brain from 27 VAD recipients supported for > 180 days between 2000 and 2010 were histologically examined to assess vascular alterations, including perivascular infiltrate, intravascular infiltrate, wall thickness, thrombosis, endothelial cell swelling, vessel wall necrosis, and pen-vascular fibrosis. Pulsatile-flow (PF) devices had been inserted in 9 patients and CF devices had been inserted in 16. The pathologist was blinded to the group distribution. Demographic, pharmacologic, and clinical data were retrospectively analyzed before surgery and during the follow-up period of up to 24 months. RESULTS: Median duration of support was 467 days (range, 235-1,588 days) in the PF group and 263 days (range, 182-942 days) in the CF group. Demographic and clinical data before and after surgery were similar. Amiodarone was more often used during follow-up in CF group than in the PF group (61% vs 10%, p = 0.009). Throughout the follow-up period, mean arterial pressure did not differ between recipients of the 2 pump types, nor did systolic and diastolic pressure, except at 2 weeks after VAD implantation, when systolic blood pressure was higher (p = 0.05) and diastolic lower (p = 0.03) in the PF group. Histologic studies did not identify any relevant differences in arterial wall characteristics between the 2 groups. CONCLUSION: Long-term mechanical circulatory support with CF devices does not adversely influence arterial wall properties of the end-organ vasculature. J Heart Lung Transplant 2012; 31: 1171-1176 (C) 2012 International Society for Heart and Lung Transplantation. All rights reserved.
引用
收藏
页码:1171 / 1176
页数:6
相关论文
共 50 条
  • [41] Design and Drive Control of Novel Pulsatile-flow Left Ventricular Assist Device.
    Yu, H.
    Lai, K.
    Lin, S.
    Lee, S.
    2015 IEEE MAGNETICS CONFERENCE (INTERMAG), 2015,
  • [42] Contemporary Mechanical Circulatory Support With Continuous-Flow Biventricular Assist Devices: A Systematic Review
    Meece, Lauren E.
    Yu, Julia
    Jeng, Eric, I
    Ahmed, Mustafa M.
    CARDIOLOGY IN REVIEW, 2022, 30 (04) : 197 - 205
  • [43] Mechanical Circulatory Support for the Failing Heart: Continuous-Flow Left Ventricular Assist Devices
    Englert, Joseph A. R., III
    Davis, Jennifer A.
    Krim, Selim R.
    OCHSNER JOURNAL, 2016, 16 (03): : 263 - 269
  • [44] Arterial Compliance and Continuous-Flow Left Ventricular Assist Device Pump Function
    Adji, Audrey
    Shehab, Sajad
    Jain, Pankaj
    Robson, Desiree
    Jansz, Paul
    Hayward, Christopher S.
    ASAIO JOURNAL, 2022, 68 (07) : 925 - 931
  • [45] MEASURING DEVICE FOR THE MEASURING OF THE WALL THICKNESS OF HOT ROLLED TUBES IN THE CONTINUOUS-FLOW
    不详
    NEUE HUTTE, 1985, 30 (02): : 55 - 56
  • [46] The Lavare Cycle: A Novel Pulsatile Feature of the HVAD Continuous-Flow Left Ventricular Assist Device
    Kumar, Julia
    Elhassan, Amir
    Dimitrova, Galina
    Essandoh, Michael
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2019, 33 (04) : 1170 - 1171
  • [47] Pulsatile operation of a continuous-flow right ventricular assist device (RVAD) to improve vascular pulsatility
    Ng, Boon C.
    Kleinheyer, Matthias
    Smith, Peter A.
    Timms, Daniel
    Cohn, William E.
    Lim, Einly
    PLOS ONE, 2018, 13 (04):
  • [48] Pulsatile-flow mechanical circulatory support (MCS) as a bridge to transplantation or recovery. Single-centre experience with the POLCAS system in 2014
    Kusmierczyk, Mariusz
    Kuc, Mateusz
    Szymanski, Jaroslaw
    Juraszek, Andrzej
    Kolsut, Piotr
    Kusmierski, Krzysztof
    Zielinski, Tomasz
    Sobieszczanska-Malek, Malgorzata
    Sitkowska-Rysiak, Ewa
    KARDIOCHIRURGIA I TORAKOCHIRURGIA POLSKA, 2015, 12 (03): : 228 - 232
  • [49] Myocardial Recovery after Ventricular Unloading with Pulsatile- Versus Continuous-Flow Assist Device
    Wagner, F. M.
    Hakmi, S.
    Oelschner, C.
    Kubik, M.
    Deuse, T.
    Reichenspurner, H.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S283 - S284
  • [50] Influence of Pulsatile Flow on LDL Transport in the Arterial Wall
    Nanfeng Sun
    Nigel B. Wood
    Alun D. Hughes
    Simon A. M. Thom
    X. Yun Xu
    Annals of Biomedical Engineering, 2007, 35 : 1782 - 1790