Use of the Open-Ended Probe Technique for the Dielectric Characterization of Biological Tissues at Low Frequencies.

被引:0
|
作者
Liporace, Flavia [1 ]
Cavagnaro, Marta [1 ]
机构
[1] Univ Roma La Sapienza, Dept Informat Engn Elect & Telecommun, Rome, Italy
关键词
dielectric properties; open-ended probe; accuracy; radial dimensions; MICROWAVE-FREQUENCIES; PERMITTIVITY; RADIO;
D O I
暂无
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
The open-ended probe is a popular technique adopted for the measurement of the dielectric properties of biological tissues. Commercial probes are commonly used for measurements at frequencies higher than 500 MHz because under this frequency limit the error of reconstruction increases and reaches not acceptable values. The aim of this paper is to understand how to improve the accuracy of this technique for the dielectric characterization of samples at low frequencies. Theorical studies have shown that the measurement accuracy is related to a capacitive parameter of the probe which is dependent on the radial dimensions of the probe itself. Starting from this assumption, in this work some experimental measurements have been done on known liquids finding that the dimensions of the probe can influence the results. It is demonstrated that the choice of the probe with the optimal structure can improve the measurement accuracy at low frequencies.
引用
收藏
页码:66 / 69
页数:4
相关论文
共 50 条
  • [41] Broad Band Microwave Dielectric Measurement of Liquids with an Open-ended Coaxial Probe
    Shi, Jiaming
    Balzano, Quirino
    Davis, Christopher C.
    MEASUREMENT TECHNOLOGY AND ENGINEERING RESEARCHES IN INDUSTRY, PTS 1-3, 2013, 333-335 : 191 - 198
  • [42] Open-ended waveguide dielectric probe using time-domain measurements
    Chan, Kevin
    Rambabu, Karumudi
    Boulanger, Pierre
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2018, 60 (05) : 1108 - 1112
  • [43] FEM Analysis of Conical Type Coaxial Open-ended Probe for Dielectric Measurement
    Arab, H.
    Akyel, C.
    PIERS 2014 GUANGZHOU: PROGRESS IN ELECTROMAGNETICS RESEARCH SYMPOSIUM, 2014, : 2223 - 2227
  • [44] Distributed parameters for low-frequency dielectric characterization of liquids with open-ended coaxial cell
    Iglesias, TP
    Pereira, SA
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2006, 55 (01) : 176 - 179
  • [45] VASECTOMY FAILURE AFTER USE OF AN OPEN-ENDED TECHNIQUE - REPLY
    GOLDSTEIN, M
    FERTILITY AND STERILITY, 1985, 43 (04) : 667 - 668
  • [46] Dielectric properties of rubber wood at microwave frequencies measured with an open-ended coaxial line
    Kabir, MF
    Khalid, KB
    Daud, WM
    Aziz, SHA
    WOOD AND FIBER SCIENCE, 1997, 29 (04): : 319 - 324
  • [47] Corrections on "Precision open-ended coaxial probes for in vivo and ex vivo dielectric spectroscopy of biological tissues at microwave frequencies" (vol 53, pg 1713, 2005)
    Popovic, D
    McCartney, L
    Beasley, C
    Lazebnik, M
    Okoniewski, M
    Hagness, SC
    Booske, JH
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2005, 53 (09) : 3053 - 3053
  • [48] A new technique for measuring electromagnetic properties of materials with an open-ended coaxial probe
    Wang, SJ
    Xu, DM
    Niu, M
    1997 ASIA-PACIFIC MICROWAVE CONFERENCE PROCEEDINGS, VOLS I-III, 1997, : 921 - 924
  • [49] A new technique for measuring electromagnetic properties of materials with an open-ended coaxial probe
    Wang, SJ
    Xu, DN
    Niu, M
    1997 ASIA-PACIFIC MICROWAVE CONFERENCE PROCEEDINGS, VOLS I-III, 1997, : 765 - 768
  • [50] A nomogram technique for microwave permittivity determination using an open-ended coaxial probe
    Ivanov, V. K.
    Silin, O. O.
    Stadnyk, O. M.
    SIXTH INT KHARKOV SYMPOSIUM ON PHYSICS AND ENGINEERING OF MICROWAVES, MILLIMETER AND SUBMILLIMETER WAVES/WORKSHOP ON TERAHERTZ TECHNOLOGIES, VOLS 1 AND 2, 2007, : 910 - +