Sinogram affirmed iterative reconstruction in head CT: Improvement of objective and subjective image quality with concomitant radiation dose reduction

被引:43
|
作者
Korn, A. [1 ]
Bender, B. [1 ]
Fenchel, M. [1 ]
Spira, D. [2 ]
Schabel, C. [2 ]
Thomas, C. [2 ]
Flohr, T. [3 ]
Claussen, C. D. [2 ]
Bhadelia, R. [4 ]
Ernemann, U. [1 ]
Brodoefel, H. [2 ,4 ]
机构
[1] Dept Diagnost & Intervent Neuroradiol, D-72076 Tubingen, Germany
[2] Dept Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[3] Siemens HealthCare, Computed Tomog Div, D-91301 Forchheim, Germany
[4] Beth Israel Deaconess Med Ctr, Boston, MA 02115 USA
关键词
Head CT; Iterative reconstruction; Image quality; FILTERED BACK-PROJECTION; ART;
D O I
10.1016/j.ejrad.2013.03.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Iterative reconstruction has recently been revisited as a promising concept for substantial CT dose reduction. The purpose of this study was to assess the potential benefit of sinogram affirmed iterative reconstruction (SAFIRE) in head CT by comparing objective and subjective image quality at reduced tube current with standard dose filtered back projection (FBP). Materials and methods: Non-contrast reduced dose head CT (255 mA s, CTDIvol 47.8 mGy) was performed in thirty consecutive patients and reconstructed with SAFIRE and FBP. Images were assessed in terms of quantitative and qualitative image quality and compared with FBP of standard dose acquisitions (320 mA s, CTDIvol 59.7 mGy). Results: In reduced dose CT examinations, use of SAFIRE versus FBP resulted in 47% increase in contrast-to-noise ratio (CNR) (2.49 vs. 1.69; p<0.0001). While reduction of tube current was associated with 13% decrease in CNR, quantitative degradation of image quality at lower dose was more than compensated through SAFIRE (2.49 vs. 1.96; p = 0.0004). Objective measurements of image sharpness were comparable between FBP and SAFIRE reconstructions (575.9 +/- 74.1 vs. 583.4 +/- 74.7 change in HU/Pixel; p = 0.28). Compared to standard dose FBP, subjective grading of noise as well as overall image quality scores were significantly improved when SAFIRE was used in reduced dose exams (1.3 vs. 1.6, p = 0.006; 1.3 vs. 1.7, p = 0.026). Conclusion: At 20% dose reduction, reconstruction of head CT by SAFIRE provides above standard objective and subjective image quality, suggesting potential for more vigorous dose savings in neuroradiology CT applications. (C) 2013 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1431 / 1435
页数:5
相关论文
共 50 条
  • [1] Sinogram-Affirmed Iterative Reconstruction of Low-Dose Chest CT: Effect on Image Quality and Radiation Dose
    Kalra, Mannudeep K.
    Woisetschlaeger, Mischa
    Dahlstroem, Nils
    Singh, Sarabjeet
    Digumarthy, Subbarao
    Do, Synho
    Pien, Homer
    Quick, Petter
    Schmidt, Bernhard
    Sedlmair, Martin
    Shepard, Jo-Anne O.
    Persson, Anders
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 201 (02) : W235 - W244
  • [2] Cerebral bone subtraction CT angiography using 80 kVp and sinogram-affirmed iterative reconstruction: contrast medium and radiation dose reduction with improvement of image quality
    Nagayama, Yasunori
    Nakaura, Takeshi
    Tsuji, Akinori
    Urata, Joji
    Furusawa, Mitsuhiro
    Yuki, Hideaki
    Hirarta, Kenichiro
    Oda, Seitaro
    Kidoh, Masafumi
    Utsunomiya, Daisuke
    Yamashita, Yasuyuki
    [J]. NEURORADIOLOGY, 2017, 59 (02) : 127 - 134
  • [3] Cerebral bone subtraction CT angiography using 80 kVp and sinogram-affirmed iterative reconstruction: contrast medium and radiation dose reduction with improvement of image quality
    Yasunori Nagayama
    Takeshi Nakaura
    Akinori Tsuji
    Joji Urata
    Mitsuhiro Furusawa
    Hideaki Yuki
    Kenichiro Hirarta
    Seitaro Oda
    Masafumi Kidoh
    Daisuke Utsunomiya
    Yasuyuki Yamashita
    [J]. Neuroradiology, 2017, 59 : 127 - 134
  • [4] Radiation Dose Reduction With Sinogram Affirmed Iterative Reconstruction Technique for Abdominal Computed Tomography
    Kalra, Mannudeep K.
    Woisetschlager, Mischa
    Dahlstrom, Nils
    Singh, Sarabjeet
    Lindblom, Maria
    Choy, Garry
    Quick, Petter
    Schmidt, Bernhard
    Sedlmair, Martin
    Blake, Michael A.
    Persson, Anders
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2012, 36 (03) : 339 - 346
  • [5] Image quality, diagnostic accuracy, and potential for radiation dose reduction in thoracoabdominal CT, using Sinogram Affirmed Iterative Reconstruction (SAFIRE) technique in a longitudinal study
    Scharf, Michael
    Brendel, Stephanie
    Melzer, Katja
    Hentschke, Christian
    May, Matthias
    Uder, Michael
    Lell, Michael M.
    [J]. PLOS ONE, 2017, 12 (07):
  • [6] Image quality and radiation dose of low dose coronary CT angiography in obese patients: Sinogram affirmed iterative reconstruction versus filtered back projection
    Wang, Rui
    Schoepf, U. Joseph
    Wu, Runze
    Reddy, Ryan P.
    Zhang, Chuanchen
    Yu, Wei
    Liu, Yi
    Zhang, Zhaoqi
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) : 3141 - 3145
  • [7] Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation
    Schmid, Andrea I.
    Uder, Michael
    Lell, Michael M.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2016, 46 (01)
  • [8] Image quality assessment of ultra low-dose chest CT using sinogram-affirmed iterative reconstruction
    So Won Lee
    Yookyung Kim
    Sung Shine Shim
    Jeong Kyong Lee
    Seok Jeong Lee
    Yon Ju Ryu
    Jung Hyun Chang
    [J]. European Radiology, 2014, 24 : 817 - 826
  • [9] Objective assessment of image quality and dose reduction in CT iterative reconstruction
    Vaishnav, J. Y.
    Jung, W. C.
    Popescu, L. M.
    Zeng, R.
    Myers, K. J.
    [J]. MEDICAL PHYSICS, 2014, 41 (07)
  • [10] Image quality assessment of ultra low-dose chest CT using sinogram-affirmed iterative reconstruction
    Lee, So Won
    Kim, Yookyung
    Shim, Sung Shine
    Lee, Jeong Kyong
    Lee, Seok Jeong
    Ryu, Yon Ju
    Chang, Jung Hyun