Electromagnetic interference from welding and motors on implantable cardioverter-defibrillators as tested in the electrically hostile work site

被引:32
|
作者
Fetter, JG
Benditt, DG
Stanton, MS
机构
[1] UNIV MINNESOTA,DEPT MED,CARDIAC ARRHYTHMIA CTR,MINNEAPOLIS,MN 55455
[2] MAYO CLIN,ROCHESTER,MN
关键词
D O I
10.1016/0735-1097(96)00147-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This study was designed to determine the susceptibility of an implanted cardioverter-defibrillator to electromagnetic interference in an electrically hostile work site environment, with the ultimate goal of allowing the patient to return to work. Background. Normal operation of an implanted cardioverter-defibrillator depends on reliable sensing of the heart's electrical activity. Consequently, there is concern that external electromagnetic interference from external sources in the work place, especially welding equipment or motor-generator systems, may be sensed and produce inappropriate shocks or abnormal reed switch operation, temporarily suspending detection of ventricular tachycardia or ventricular fibrillation. Methods. The effects of electromagnetic interference on the operation of one type of implantable cardioverter-defibrillator (Medtronic models 7217 and 7219) was measured by using internal event counter monitoring in 10 patients operating are welders at up to 900 A or working near 200-hp motors and 1 patient close to a locomotive starter drawing up to 400 A. Results. The electromagnetic interference produced two sources of potential interference on the sensing circuit or reed switch operation, respectively: 1) electrical fields with measured frequen cies up to 50 MHz produced by the high currents during welding electrode activation, and 2) magnetic fields produced by the current in the welding electrode and cable. The defibrillator sensitivity was programmed to the highest (most sensitive) value: 0.15 mV (model 7219) or 0.3 mV (model 7217). The ventricular tachycardia and ventricular fibrillation therapies were temporarily turned off but the detection circuits left on. Conclusions. None of the implanted defibrillators tested were affected by oversensing of the electric field as verified by telemetry from the detection circuits. The magnetic field from 225-A welding current produced a flux density of 1.2 G; this density was not adequate to close the reed switch, which requires similar to 10 G. Our testing at the work site revealed no electrical interference with this type of defibrillator. Patients were allowed to return to work, The following precautions should be observed by the patient: 1) maintain a minimal distance of 2 ft (61 cm) from the melding are and cables or large motors, 2) do not exceed tested currents with the welding equipment, 3) wear insulated gloves while operating electrical equipment, 4) verify that electrical equipment is properly grounded, and 5) stop welding and leave the work area immediately if a therapy is delivered or a feeling of lightheadedness is experienced.
引用
收藏
页码:423 / 427
页数:5
相关论文
共 50 条
  • [21] Implantation of implantable cardioverter-defibrillators from an ileofemoral approach
    Perzanowski, C
    Timothy, P
    McAfee, M
    McDaniel, M
    Meyer, D
    Torres, V
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 11 (02) : 155 - 159
  • [22] Correction to: Electromagnetic interference in implantable cardioverter defibrillators: present but rare
    Gesa von Olshausen
    Ina-Christine Rondak
    Carsten Lennerz
    Verena Semmler
    Christian Grebmer
    Tilko Reents
    Sonia Ammar-Busch
    Alessandra Buiatti
    Felix Bourier
    Isabel Deisenhofer
    Christof Kolb
    Clinical Research in Cardiology, 2020, 109 : 652 - 652
  • [23] Primary Prevention Implantable Cardioverter-Defibrillators They Work in the Real World, Too
    Lampert, Rachel
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (04): : 624 - 625
  • [24] Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators
    Wolber, Thomas
    Ryf, Salome
    Binggeli, Christian
    Holzmeister, Johannes
    Brunckhorst, Corinna
    Luechinger, Roger
    Duru, Firat
    HEART RHYTHM, 2007, 4 (01) : 1 - 4
  • [25] Position of Subcutaneous Implantable Cardioverter-Defibrillators and Possible Interference on Myocardial Perfusion Imaging
    Kahaly, Omar Ray
    Shafiei, Fereidoon
    Hardebeck, Charles
    Houmsse, Mahmoud
    TEXAS HEART INSTITUTE JOURNAL, 2017, 44 (03): : 223 - 225
  • [26] Interference of two common European digital cellular phones with implantable cardioverter-defibrillators
    Niehaus, M
    Gille, R
    Cierpka, R
    Korte, T
    Tebbenjohans, J
    EUROPEAN HEART JOURNAL, 2002, 23 (07) : 586 - 588
  • [27] In vitro tests of electromagnetic interference of electromagnetic navigational bronchoscopy to implantable cardioverter defibrillators
    Magnani, Andrea
    Matheoud, Roberta
    Brambilla, Marco
    Valzano, Serena
    Occhetta, Eraldo
    Marino, Paolo
    Balbo, Piero
    EUROPACE, 2012, 14 (07): : 1054 - 1059
  • [28] Should patients with implantable cardioverter-defibrillators refrain from driving?
    Anderson, Mark
    NATURE CLINICAL PRACTICE CARDIOVASCULAR MEDICINE, 2008, 5 (06): : 304 - 305
  • [29] Should patients with implantable cardioverter-defibrillators refrain from driving?
    Mark Anderson
    Nature Clinical Practice Cardiovascular Medicine, 2008, 5 : 304 - 305
  • [30] Beam profile disturbances from interactions with implantable pacemakers and implantable cardioverter-defibrillators
    Gossman, Michael S.
    Blohm, Christopher M.
    MEDICAL DOSIMETRY, 2013, 38 (02) : 109 - 109