Simultaneous Measurement of Thermophysical Properties of Tissue-Mimicking Phantoms for High Intensity Focused Ultrasound (HIFU) Exposures

被引:3
|
作者
Gao, Jing [1 ]
You, Jiang [1 ]
Huang, Zhihong [1 ]
Cochran, Sandy [2 ]
Corner, George [3 ]
机构
[1] Univ Dundee, Sch Engn Phys & Math, Dundee DD1 4HN, Scotland
[2] Univ Dundee, Inst Med Sci & Technol, Dundee DD2 1FD, Scotland
[3] Ninewells Hosp, Dept Med Phys, Dundee DD1 1SY, Scotland
关键词
Egg white phantom; High intensity focused ultrasound (HIFU); Step-wise transient plane source (TPS) method; Thermophysical properties; Tissue-mimicking phantom; THERMAL-CONDUCTIVITY; DIFFUSIVITY; GEL;
D O I
10.1007/s10765-012-1158-4
中图分类号
O414.1 [热力学];
学科分类号
摘要
Tissue-mimicking phantoms, including bovine serum albumin phantoms and egg white phantoms, have been developed for, and in laboratory use for, real-time visualization of high intensity focused ultrasound-induced thermal coagulative necrosis since 2001. However, until now, very few data are available concerning their thermophysical properties. In this article, a step-wise transient plane source method has been used to determine the values of thermal conductivity, thermal diffusivity, and specific heat capacity of egg white phantoms with elevated egg white concentrations (0v/v% to 40v/v%, by 10v/v% interval) at room temperature (similar to 20 degrees C). The measured thermophysical properties were close to previously reported values; the thermal conductivity and thermal diffusivity were linearly proportional to the egg white concentration within the investigation range, while the specific heat capacity decreased as the egg white concentration increased. Taking account of large differences between real experiment and ideal model, data variations within 20% were accepted.
引用
收藏
页码:495 / 504
页数:10
相关论文
共 50 条
  • [1] Simultaneous Measurement of Thermophysical Properties of Tissue-Mimicking Phantoms for High Intensity Focused Ultrasound (HIFU) Exposures
    Jing Gao
    Jiang You
    Zhihong Huang
    Sandy Cochran
    George Corner
    International Journal of Thermophysics, 2012, 33 : 495 - 504
  • [2] Development and characterization of an innovative synthetic tissue-mimicking material for High Intensity Focused Ultrasound (HIFU) exposures
    Lafon, C
    Kaczkowski, PJ
    Vaezy, S
    Noble, M
    Sapozhnikov, OA
    2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2001, : 1295 - 1298
  • [3] Reusable tissue-mimicking hydrogel phantoms for focused ultrasound ablation
    Sun, Ming-Kuan
    Shieh, Jay
    Lo, Chia-Wen
    Chen, Chuin-Shan
    Chen, Ben-Ting
    Huang, Chang-Wei
    Chen, Wen-Shiang
    ULTRASONICS SONOCHEMISTRY, 2015, 23 : 399 - 405
  • [4] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Kang Il Lee
    Journal of the Korean Physical Society, 2018, 72 : 1313 - 1319
  • [5] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (11) : 1313 - 1319
  • [6] Boiling activity induced by high intensity focused ultrasound in a tissue-mimicking phantom
    Kang, Hwi Suk
    Lee, Kang Il
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2020, 59 (02)
  • [7] Development and Characterization of a Tissue-Mimicking Material for High-Intensity Focused Ultrasound
    King, Randy L.
    Liu, Yunbo
    Maruvada, Subha
    Herman, Bruce A.
    Wear, Keith A.
    Harris, Gerald R.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (07) : 1397 - 1405
  • [8] Prediction and Measurement of the Size of Thermal Lesion Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    Choi, Min Joo
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2009, 48 (02)
  • [9] Development of a tissue-mimicking phantom for evaluating the focusing performance of high intensity focused ultrasound
    Jing Zongyu
    Li Faqi
    Zou Hangzhong
    Wang Zhibiao
    THERAPEUTIC ULTRASOUND, 2006, 829 : 648 - +
  • [10] Measurement of Viscoelastic Properties of Polyacrylamide-Based Tissue-Mimicking Phantoms for Ultrasound Elastography Applications
    Kumar, Kishore
    Andrews, Maneesha E.
    Jayashankar, V.
    Mishra, Ashok K.
    Suresh, S.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (05) : 1224 - 1232