Investigation of gated cone-beam CT to reduce respiratory motion blurring

被引:17
|
作者
Kincaid, Russell E., Jr. [1 ]
Yorke, Ellen D. [1 ]
Goodman, Karyn A. [2 ]
Rimner, Andreas [2 ]
Wu, Abraham J. [2 ]
Mageras, Gig S. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10065 USA
关键词
image guided radiation therapy; cone-beam CT; respiratory motion; motion management; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; LUNG-TUMORS; QUALITY-ASSURANCE; CANCER; LIVER;
D O I
10.1118/1.4795336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Methods of reducing respiratory motion blurring in cone-beam CT (CBCT) have been limited to lung where soft tissue contrast is large. Respiration-correlated cone-beam CT uses slow continuous gantry rotation but image quality is limited by uneven projection spacing. This study investigates the efficacy of a novel gated CBCT technique. Methods: In gated CBCT, the linac is programmed such that gantry rotation and kV image acquisition occur within a gate around end expiration and are triggered by an external respiratory monitor. Standard CBCT and gated CBCT scans are performed in 22 patients (11 thoracic, 11 abdominal) and a respiration-correlated CT (RCCT) scan, acquired on a standard CT scanner, from the same day serves as a criterion standard. Image quality is compared by calculating contrast-to-noise ratios (CNR) for tumors in lung, gastroesophageal junction (GEJ) tissue, and pancreas tissue, relative to surrounding background tissue. Congruence between the object in the CBCT images and that in the RCCT is measured by calculating the optimized normalized cross-correlation (NCC) following CBCT-to-RCCT rigid registrations. Results: Gated CBCT results in reduced motion artifacts relative to standard CBCT, with better visualization of tumors in lung, and of abdominal organs including GEJ, pancreas, and organs at risk. CNR of lung tumors is larger in gated CBCT in 6 of 11 cases relative to standard CBCT. A paired two-tailed t-test of lung patient mean CNR shows no statistical significance (p = 0.133). In 4 of 5 cases where CNR is not increased, lung tumor motion observed in RCCT is small (range 1.3-5.2 mm). CNR is increased and becomes statistically significant for 6 out of 7 lung patients with > 5 mm tumor motion (p = 0.044). CNR is larger in gated CBCT in 5 of 7 GEE cases and 3 of 4 pancreas cases (p = 0.082 and 0.192). Gated CBCT yields improvement with lower NCC relative to standard CBCT in 10 of 11, 7 of 7, and 3 of 4 patients for lung, GEJ, and pancreas images, respectively (p = 0.0014, 0.0030, 0.165). Conclusions: Gated CBCT reduces image blurring caused by respiratory motion. The gated gantry rotation yields uniformly and closely spaced projections resulting in improved reconstructed image quality. The technique is shown to be applicable to abdominal sites, where image contrast of soft tissues is low. (C) 2013 American Association of Physicists in Medicine. [http ://dx.doi.org/10.1118/1.4795336]
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Investigation of Gated Cone-Beam CT to Reduce Respiratory Motion Blurring
    Kincaid, R.
    Yorke, E.
    Rimner, A.
    Mageras, G.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [2] Investigation of Gated Cone-Beam CT to Reduce Respiratory Motion Blurring in Images of Thorax and Abdomen
    Kincaid, R.
    Yorke, E.
    Goodman, K.
    Rimner, A.
    Wu, A.
    Mageras, G.
    MEDICAL PHYSICS, 2012, 39 (06) : 3904 - 3904
  • [3] INVESTIGATION OF A GATED CONE-BEAM CT TECHNIQUE TO REDUCE RESPIRATORY MOTION BLURRING IN ABDOMINAL SITES
    Kincaid, R.
    Yorke, E.
    Goodman, K.
    Wu, A.
    Mageras, G.
    RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S152 - S153
  • [4] Respiratory gated cone-beam CT volumetric imaging for external beam radiotherapy
    Moseley, DJ
    Hawkins, M
    Eccles, C
    Euler, C
    White, EA
    Bissonnette, J
    Dawson, LA
    Jaffray, DA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S27 - S28
  • [5] Breathing motion in cone-beam CT
    Rit, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S119 - S120
  • [6] Preliminary Evaluation of the Dosimetric Accuracy of the Cone-Beam CT for the Cases with the Respiratory Motion
    Kim, D.
    Chung, W.
    Lee, Y.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Characterisation of cardiac and respiratory motion in cone-beam CT images for cardiac SABR
    Daniel, J.
    Burke, K.
    Whitehurst, P.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S385 - S386
  • [8] Respiratory motion models from Cone-Beam CT for lung tumour tracking
    Fassi, A.
    Tagliabue, E.
    Tirindelli, M.
    Sarrut, D.
    Riboldi, M.
    Baroni, G.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S424 - S424
  • [9] Influence of respiratory motion on cone-beam CT (CBCT) imaging of thorax and abdomen
    Kim, G
    Li, T
    Yang, Y
    Schreibmann, E
    Thorndyke, B
    Boyer, A
    Xing, L
    MEDICAL PHYSICS, 2005, 32 (06) : 1934 - 1934
  • [10] A Hybrid Approach to Reduce Cone-Beam Artifacts for a Circular Orbit Cone-Beam CT System
    Choi, Shinkook
    Kim, Jinsung
    Baek, Jongduk
    IEEE ACCESS, 2018, 6 : 54595 - 54606