Usefulness of the Total Thrombus-Formation Analysis System (T-TAS) in the diagnosis and characterization of von Willebrand disease

被引:18
|
作者
Daidone, V. [1 ]
Barbon, G. [2 ]
Cattini, M. G. [1 ]
Pontara, E. [1 ]
Romualdi, C. [3 ]
Di Pasquale, I. [4 ]
Hosokawa, K. [5 ]
Casonato, A. [4 ]
机构
[1] Univ Padua, Med Sch, Thrombohemorrhag Disorders Unit, Dept Cardiol Thorac & Vasc Sci, Padua, Italy
[2] Univ Padua, Veneto Inst Oncol IRCCS, Familial Canc Clin & Oncoendocrinol, Padua, Italy
[3] Univ Padua, Dept Biol, Padua, Italy
[4] Univ Padua, Dept Med, Padua, Italy
[5] Res Inst Fujimori Kogyo Co, Yokohama, Kanagawa, Japan
关键词
high-molecular-weight von Willebrand factor multimers; thrombus formation; total thrombus-formation analysis system; von Willebrand disease; von Willebrand disease diagnosis; von Willebrand factor; FLOW-CHAMBER SYSTEM; ABO BLOOD-GROUP; VONWILLEBRAND-FACTOR; PATHOPHYSIOLOGY; ABNORMALITIES; INDIVIDUALS; SURVIVAL; BINDING; TYPE-1;
D O I
10.1111/hae.12971
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionThe heterogeneity of von Willebrand disease (VWD) makes its diagnosis a difficult task. MethodsWe report here on the usefulness of a microchip-based flow-chamber system, the total thrombus-formation analysis system (T-TAS), in the identification and characterization of VWD. Thirty VWD patients and 20 healthy subjects were studied with the T-TAS platelet (PL) and atherome (AR) microchips developed for the in vitro assessment of platelet thrombus formation and fibrin-rich platelet thrombus formation respectively. ResultsSamples from severe type 1 VWD, characterized by von Willebrand factor (VWF) levels below 10 U dL(-1), failed to occlude either the PL or the AR chip capillaries, while the occlusion times were normal in patients with mild type 1 VWD (VWF above 25 U dL(-1)). PL and/or AR chip occlusion occurred, but took longer than normal, for samples from type Vicenza and type 1 VWD patients, whose VWF levels ranged between 10 and 25 U dL(-1). No PL or AR chip capillary occlusion was seen for samples from patients with type 2A or 2B VWD featuring the absence of large VWF multimers, whereas no abnormalities emerged for type 2B patients with normal multimer patterns. ConclusionThe T-TAS appears to be sensitive mainly to plasma VWF concentration and the presence of large multimers. Failure of the PL and AR chips to become occluded points to a lack of large VWF multimers, or type 1 VWD with VWF levels below 10 U dL(-1). Although the T-TAS does not assure a precise VWD diagnosis, it does point us in the right direction, and thus seems a useful global preliminary test.
引用
收藏
页码:949 / 956
页数:8
相关论文
共 45 条
  • [41] A novel haemodilution chip mounted on the total thrombus-formation analysis system, a flow chamber system, enables stable analysis of platelet products under low platelet concentration/high shear conditions
    Fuchizaki, Akihiro
    Yasui, Kazuta
    Hayashi, Tomoya
    Tanaka, Mitsunobu
    Oyamada, Chiaki
    Ohnishi-Wada, Tomoko
    Hosokawa, Kazuya
    Fujimura, Yoshihiro
    Shimogaki, Kazushige
    Hirayama, Fumiya
    Kimura, Takafumi
    Takihara, Yoshihiro
    VOX SANGUINIS, 2024, 119 (08) : 883 - 887
  • [42] Elevated Leukocyte Count and Platelet-Derived Thrombogenicity Measured Using the Total Thrombus-Formation Analysis System in Patients with ST-Segment Elevation Myocardial Infarction
    Kikuchi, Shinnosuke
    Tsukahara, Kengo
    Ichikawa, Shinya
    Abe, Takeru
    Nakahashi, Hidefumi
    Minamimoto, Yugo
    Kimura, Yuichiro
    Akiyama, Eiichi
    Okada, Kozo
    Matsuzawa, Yasushi
    Konishi, Masaaki
    Maejima, Nobuhiko
    Iwahashi, Noriaki
    Kosuge, Masami
    Ebina, Toshiaki
    Tamura, Kouichi
    Kimura, Kazuo
    Hibi, Kiyoshi
    JOURNAL OF ATHEROSCLEROSIS AND THROMBOSIS, 2024, 31 (09) : 1277 - 1292
  • [43] Combination of Two Different Microchips in Total Thrombus-Formation Analysis System Predicts High Risk Patients With Periprocedural Bleeding Events Events After Percutaneous Coronary Intervention
    Oimatsu, Yu
    Kaikita, Koichi
    Ishii, Masanobu
    Mitsuse, Tatsuro
    Ito, Miwa
    Arima, Yuichiro
    Sueta, Daisue
    Kanazawa, Hisanori
    Takashio, Seiji
    Yamamoto, Eiichiro
    Sakamoto, Kenji
    Izumiya, Yasuhiro
    Kojima, Sunao
    Tsujita, Kenichi
    CIRCULATION, 2017, 136
  • [44] Platelet-Derived Thrombogenicity Measured by Total Thrombus-Formation Analysis System in Patients With ST-Segment Elevation Myocardial Infarction Undergoing Primary Percutaneous Coronary Intervention
    Kikuchi, Shinnosuke
    Tsukahara, Kengo
    Ichikawa, Shinya
    Abe, Takeru
    Minamimoto, Yugo
    Kimura, Yuichiro
    Akiyama, Eiichi
    Nakayama, Naoki
    Okada, Kozo
    Matsuzawa, Yasushi
    Konishi, Masaaki
    Maejima, Nobuhiko
    Iwahashi, Noriaki
    Hibi, Kiyoshi
    Kosuge, Masami
    Ebina, Toshiaki
    Tamura, Kouichi
    Kimura, Kazuo
    CIRCULATION JOURNAL, 2020, 84 (06) : 975 - +
  • [45] Total Thrombus Formation Analysis System Can Predict High Bleeding Risk in Coronary Artery Disease Patients Undergoing Percutaneous Coronary Intervention
    Nakanishi, Nobuhiro
    Kaikita, Koichi
    Tsujita, Kenichi
    CIRCULATION, 2020, 142