Development of a novel cardiopulmonary resuscitation measurement tool using real-time feedback from wearable wireless instrumentation

被引:7
|
作者
Ward, Sarah R. [1 ]
Scott, Bronwyn C. [1 ]
Rubin, David M. [1 ]
Pantanowitz, Adam [1 ]
机构
[1] Univ Witwatersrand, Sch Elect & Informat Engn, Biomed Engn Res Grp, Private Bag 3, ZA-2050 Johannesburg, South Africa
关键词
Cardiopulmonary resuscitation (CPR); Quality; Dual-quaternions; Electromyogram (EMG); Inertial Measurement Unit (IMU); MYO; Machine learning; IMPROVES CPR QUALITY; AUDIOVISUAL FEEDBACK; PERFORMANCE; COMPRESSION;
D O I
10.1016/j.resuscitation.2019.02.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: The design and implementation of a wearable training device to improve cardiopulmonary resuscitation (CPR) is presented. Methods: The MYO contains both Electromyography (EMG) and Inertial Measurement Unit (IMU) sensors which are used to detect effective CPR, and the four common incorrect hand and arm positions viz. relaxed fingers; hands too low on the sternum; patient too close; or patient too far. The device determines the rate and depth of compressions calculated using a Fourier transform and dual-quaternions respectively. In addition, common positional mistakes are determined using classification algorithms (six machine learning algorithms are considered and tested). Feedback via Graphical User Interface (GUI) and audio is integrated. Results: The system is tested by performing CPR on a mannequin and comparing real-time results to theoretical values. Tests show that although the classification algorithm performed well in testing (98%), in real time, it had low accuracy for certain categories (60%) which are attributable to the MYO calibration, sampling rate and misclassification of similar hand positions. Combining these similar incorrect positions into more general categories significantly improves accuracy, and produces the same improved outcome of improved CPR. The rate and depth measures have a general accuracy of 97%. Conclusion: The system allows for portable, real-time feedback for use in training and in the field, and shows promise toward classifying and improving the administration of CPR.
引用
收藏
页码:183 / 189
页数:7
相关论文
共 50 条
  • [41] REAL-TIME FEEDBACK DERIVED FROM WEARABLE SENSORS TO IMPROVE GAIT IN PARKINSON'S DISEASE
    Bartels, Brandon M.
    Moreno, Arianna
    Quezada, Maria Jose
    Sivertson, Haley
    Abbas, James
    Krishnamurthi, Narayanan
    TECHNOLOGY AND INNOVATION, 2018, 20 (1-2) : 37 - 46
  • [42] The effect of automatic external defibrillator with a real-time feedback on quality of bystander cardiopulmonary resuscitation: A before-and-after simulation study
    Kim, Chu Hyun
    Kim, Tae Han
    Shin, Sang Do
    Song, Kyoung Jun
    Ro, Young Sun
    Ahn, Ki Ok
    Hong, Ki Jeong
    Lee, Yu Jin
    Lee, Eui Jung
    Ha, So Young
    HEALTH & SOCIAL CARE IN THE COMMUNITY, 2019, 27 (05) : E744 - E751
  • [43] Association of Real-Time Feedback and Cardiopulmonary-Resuscitation Quality Delivered by Ambulance Personnel for Out-of-Hospital Cardiac Arrest
    Lyngby, Rasmus Meyer
    Quinn, Tom
    Oelrich, Roselil Maria
    Nikoletou, Dimitra
    Gregers, Mads Christian Tofte
    Kjolbye, Julie Samsoe
    Ersboll, Annette Kjaer
    Folke, Fredrik
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2023, 12 (20):
  • [44] Wearable shear and normal force sensing glove development for real-time feedback on assembly line processes
    Kerner, Scott
    Krugh, Matthew
    Mears, Laine
    JOURNAL OF MANUFACTURING SYSTEMS, 2022, 64 : 668 - 675
  • [45] Do cardiopulmonary resuscitation real-time audiovisual feedback devices improve patient outcomes? A systematic review and meta-analysis
    Sood, Nitish
    Sangari, Anish
    Goyal, Arnav
    Sun, Christina
    Horinek, Madison
    Hauger, Joseph Andy
    Perry, Lane
    WORLD JOURNAL OF CARDIOLOGY, 2023, 15 (10): : 531 - 541
  • [46] TCP/IP connection management using a real-time development tool
    Gray, A
    Whitelock, R
    Peytchev, E
    Al-Dabass, D
    ESM 2003: 17TH EUROPEAN SIMULATION MULTICONFERENCE: FOUNDATIONS FOR SUCCESSFUL MODELLING & SIMULATION, 2003, : 636 - 643
  • [47] Confident Head Circumference Measurement from Ultrasound with Real-Time Feedback for Sonographers
    Budd, Samuel
    Sinclair, Matthew
    Khanal, Bishesh
    Matthew, Jacqueline
    Lloyd, David
    Gomez, Alberto
    Toussaint, Nicolas
    Robinson, Emma C.
    Kainz, Bernhard
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2019, PT IV, 2019, 11767 : 683 - 691
  • [48] Development of wireless communication system in real-time internal radiation dose measurement system using magnetic field
    Sato, Fumihiro
    Shinohe, Kohta
    Takura, Tetsuya
    Matsuki, Hidetoshi
    Yamada, Syogo
    Sato, Tadakuni
    JOURNAL OF APPLIED PHYSICS, 2009, 105 (07)
  • [49] Wearable E-Textile and CNT Sensor Wireless Measurement System for Real-Time Penile Erection Monitoring
    Heo, Yongki
    Kim, Jinhyung
    Cha, Cheolung
    Shin, Kyusik
    Roh, Jihyoung
    Jo, Jungki
    SENSORS, 2022, 22 (01)
  • [50] Cerebral Oximetry as a Real-Time Monitoring Tool to Assess Quality of In-Hospital Cardiopulmonary Resuscitation and Post Cardiac Arrest Care
    Ibrahim, Akram W.
    Trammell, Antoine R.
    Austin, Harland
    Barbour, Kenya
    Onuorah, Emeka
    House, Dorothy
    Miller, Heather L.
    Tutt, Chandila
    Combs, Deborah
    Phillips, Roger
    Dickert, Neal W.
    Zafari, A. Maziar
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2015, 4 (08): : e001859