Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids

被引:79
|
作者
Fleming, Cherina [1 ]
Brouwer, Rob [1 ]
Lindemans, Jan [1 ]
de Jonge, Robert [1 ]
机构
[1] Univ Med Ctr Rotterdam, Dept Clin Chem, Erasmus MC, NL-3015 CE Rotterdam, Netherlands
关键词
body fluid; cerebrospinal fluid; counting chamber; hematology analyzer; leukocytes; Sysmex XN-1000; FLOW-CYTOMETRIC ANALYSIS; AUTOMATED HEMATOLOGY; PLEURAL FLUID; MODE; ANALYZERS; XE-5000;
D O I
10.1515/cclm-2011-0927
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: We evaluated the body fluid (BF) module on the new Sysmex XN-1000 for counting blood cells. Methods: One hundred and eighty-seven BF samples [73 cerebrospinal fluid (CSF), 48 continuous ambulatory peritoneal dialysis (CAPD), 46 ascites, and 20 pleural fluid] were used for method comparison between the XN-1000 and manual microscopy (Fuchs-Rosenthal chamber and stained cytospin slides) for counting red blood cells (RBCs) and white blood cells (WBCs) (differential). Results: Good agreement was found for counting WBCs (y=1.06x+0.09, n=67, R-2=0.96) and mononuclear cells (MNs) (y=1.04x-0.01, n=40, R-2=0.93) in CSF. However, the XN-1000 systematically counted more polymorphonuclear cells (PMNs) (y=1.48x+0.18, n=40, R-2=0.99) compared to manual microscopy. Excellent correlation for RBCs >1x109/L (y=0.99x+116.56, n=26, R-2=0.99) in CSF was found. For other fluids (CAPD, ascites and pleural fluid) excellent agreement was found for counting WBCs (y=1.06x+0.26, n=109, R-2=0.98), MNs (y=1.06x-0.41, n=93, R-2=0.96), PMNs (y=1.06x+0.81, n=93, R-2=0.98) and RBCs (y=1.04x+110.04, n=43, R-2=0.98). By using BF XN-check, the lower limit of quantitation (LLoQ) for WBC was defined at 5x10(6)/L. Linearity was excellent for both the WBCs (R-2=0.99) and RBCs (R-2=0.99) and carry-over never exceeded 0.05%. Conclusions: The BF module on the XN-1000 is a suitable tool for fast and accurate quantification of WBC (differential) and RBC counts in CSF and other BFs in a diagnostic setting.
引用
收藏
页码:1791 / 1798
页数:8
相关论文
共 50 条
  • [1] VALIDATION OF THE BODY FLUID MODULE ON THE NEW SYSMEX XN-1000 FOR COUNTING BLOOD CELLS IN CEREBROSPINAL FLUID AND OTHER BODY FLUIDS
    Fleming, Cherina
    Brouwer, Rob
    Lindemans, Jan
    de Jonge, Robert
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 112 - 113
  • [2] BLOOD CELLS ANALYSIS IN CEREBROSPINAL FLUID USING SYSMEX® XN-1000 HEMATOLOGY ANALYSER
    Alic, Elizabeta Boznar
    Kase, Simona
    Skitek, Milan
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 81 - 81
  • [3] Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
    de Jonge, Robert
    Brouwer, Rob
    de Graaf, Marieke T.
    Luitwieler, Ronald L.
    Fleming, Cherina
    de Frankrijker-Merkestijn, Magda
    Smitt, Peter A. E. Sillevis
    Boonstra, Joke G.
    Lindemans, Jan
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (05) : 665 - 675
  • [4] Characterization of Candida spp. interference on the Sysmex XN-1000 body fluid mode
    Favresse, J.
    van Dievoet, M. -A.
    De Wolf, H.
    Rodriguez-Villalobos, H.
    Defour, J. -P.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (02) : E28 - E32
  • [5] Two-site evaluation of a new workflow for the detection of malignant cells on the Sysmex XN-1000 body fluid analyzer
    Favresse, Julien
    Boland, Lidvine
    Schellen, Marie
    Fervaille, Caroline
    Wuestenberghs, Fabien
    Camboni, Alessandra
    Chatelain, Bernard
    Mullier, Francois
    Defour, Jean-Philippe
    Jacqmin, Hugues
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (05) : 544 - 551
  • [6] Evaluation of Sysmex XN-1000 High-Sensitive Analysis (hsA) Research Mode for Counting and Differentiating Cells in Cerebrospinal Fluid
    Fleming, Cherina
    Russcher, Henk
    Brouwer, Rob
    Lindemans, Jan
    de Jonge, Robert
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2016, 145 (03) : 299 - 307
  • [7] VALIDATION AND IMPLEMENTATION OF THE SYSMEX XN 2000 BODY FLUID MODULE FOR WHITE BLOOD CELL DIFFERENTIAL COUNT IN SEROUS FLUIDS
    Labaere, Delphine
    Boeckx, Nancy
    Geerts, Inge
    Moens, Marc
    Van den Driessche, Marleen
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 85 - 85
  • [8] Cell counting chamber vs. Sysmex XN-1000 for determining white blood cell count and differentiation for body fluids
    Evren, Ebru
    Us, Ebru
    Hekimoglu, Can Huseyin
    Karahan, Zeynep Ceren
    TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI, 2023, 48 (01): : 19 - 25
  • [9] Performance of the Sysmex XN-V body fluid module for canine cerebrospinal fluid cell count
    Ortiz-Nisa, Sonia
    Sanz, Alba
    Pastor, Josep
    De la Fuente, Cristian
    Anor, Sonia
    VETERINARY CLINICAL PATHOLOGY, 2021, 50 (03) : 359 - 368
  • [10] UF-1000i: validation of the body fluid mode for counting cells in body fluids
    Fleming, Cherina
    Brouwer, Rob
    van Alphen, Adriaan
    Lindemans, Jan
    de Jonge, Robert
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (12) : 1781 - 1790