A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

被引:24
|
作者
Qi, Zhen-Yu [1 ,2 ]
Deng, Xiao-Wu [1 ,2 ]
Cao, Xin-ping [1 ]
Huang, Shao-Min [1 ]
Lerch, Michael [2 ]
Rosenfeld, Anatoly [2 ]
机构
[1] Sun Yat Sen Univ, Ctr Canc, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
中国国家自然科学基金;
关键词
MOSFET; in vivo dosimetry; real time dosimetry; intracaviatry brachytherapy; quality assurance; AVOID MISADMINISTRATIONS; SPATIAL-RESOLUTION; MOSFET DOSIMETRY; RADIOTHERAPY; CALIBRATION; TLD;
D O I
10.1118/1.4758067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 +/- 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% +/- 3.8% and -0.1% +/- 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4758067]
引用
收藏
页码:6757 / 6763
页数:7
相关论文
共 50 条
  • [21] Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
    Jamalludin, Zulaikha
    Jong, Wei Loong
    Malik, Rozita Abdul
    Rosenfeld, Anatoly
    Ung, Ngie Min
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 58 : 1 - 7
  • [22] Factors affecting the dosimetry of high-dose rate intracavitary brachytherapy in cervical cancer
    Mahapatra, Bikash Ranjan
    Barik, Sandip Kumar
    Das, Deepak Kumar
    Das Majumdar, Saroj Kumar
    Parida, Dillip Kumar
    Ramasubbu, Mathan Kumar
    Badajena, Avinash
    Barik, Bijay Kumar
    Mishra, Minakshi
    Pattanaik, Ashutosh
    Kanungo, Satyabrata
    Muraleedharan, Anupam
    Ahmed, Sk Soel
    Mukherjee, Priyanka
    Abdulla, Shaha Sheik
    Mahajan, Ankur
    Sarkar, Arnab
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2024, 23
  • [23] Modern Brachytherapy Techniques - Real-Time High Dose Rate Brachytherapy for Prostate Cancer
    Miszczak, D.
    ACTA PHYSICA POLONICA A, 2009, 115 (02) : 583 - 585
  • [24] Additional rectal and sigmoid mucosal points and doses in high dose rate intracavitary brachytherapy for carcinoma cervix: A dosimetric study
    Mahantshetty, Umesh
    Tiwana, Manpreet Singh
    Jamema, Swamidas
    Mishra, Sanjib
    Engineer, Reena
    Deshpande, Deepak
    Shrivastava, Shyamkishore
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2011, 7 (03) : 298 - 303
  • [25] INTRACAVITARY RADIATION IN THE TREATMENT OF ADVANCED ESOPHAGEAL-CARCINOMA - A COMPARISON OF HIGH-DOSE RATE VS LOW-DOSE RATE BRACHYTHERAPY
    HARVEY, JC
    FLEISCHMAN, EH
    BELLOTTI, JE
    KAGAN, AR
    JOURNAL OF SURGICAL ONCOLOGY, 1993, 52 (02) : 101 - 104
  • [26] BEST IN PHYSICS (THERAPY) - A Real-Time Applicator Position Monitoring System (RAPS) for Intracavitary High-Dose-Rate Brachytherapy
    Xia, J.
    Waldron, T.
    Kim, Y.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [27] HIGH-DOSE RATE INTRACAVITARY BRACHYTHERAPY FOR CARCINOMA OF THE CERVIX - THE MADISON SYSTEM .1. CLINICAL AND RADIOBIOLOGICAL CONSIDERATIONS
    STITT, JA
    FOWLER, JF
    THOMADSEN, BR
    BUCHLER, DA
    PALIWAL, BP
    KINSELLA, TJ
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (02): : 335 - 348
  • [28] Treatment of Cervical Carcinoma with High-Dose Rate Intracavitary Brachytherapy: Two Years Follow-Up Study
    Das, Diptimay
    Chaudhuri, Snehamay
    Deb, Asit Ranjan
    Aich, Ranen Kanti
    Gangopadhyay, Subir
    Ray, Amitava
    ASIAN PACIFIC JOURNAL OF CANCER PREVENTION, 2011, 12 (03) : 807 - 810
  • [29] Long-term results of high-dose rate intracavitary brachytherapy for squamous cell carcinoma of the uterine cervix
    Nakano, T
    Kato, S
    Ohno, T
    Tsujii, H
    Sato, S
    Fukuhisa, K
    Arai, T
    CANCER, 2005, 103 (01) : 92 - 101
  • [30] Variation in applicator positions during interfraction high-dose rate brachytherapy in carcinoma cervix: dosimetric evaluation
    Passi, Kamlesh
    Aggarwal, Lalit M.
    Vashistha, Rajesh
    Singh, Bikramjit
    Kaur, Parveen
    Gupta, Sureshchandra J.
    Kehwar, Than S.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2013, 12 (04) : 305 - 317