A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

被引:24
|
作者
Qi, Zhen-Yu [1 ,2 ]
Deng, Xiao-Wu [1 ,2 ]
Cao, Xin-ping [1 ]
Huang, Shao-Min [1 ]
Lerch, Michael [2 ]
Rosenfeld, Anatoly [2 ]
机构
[1] Sun Yat Sen Univ, Ctr Canc, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
中国国家自然科学基金;
关键词
MOSFET; in vivo dosimetry; real time dosimetry; intracaviatry brachytherapy; quality assurance; AVOID MISADMINISTRATIONS; SPATIAL-RESOLUTION; MOSFET DOSIMETRY; RADIOTHERAPY; CALIBRATION; TLD;
D O I
10.1118/1.4758067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 +/- 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% +/- 3.8% and -0.1% +/- 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4758067]
引用
收藏
页码:6757 / 6763
页数:7
相关论文
共 50 条
  • [1] In vivo real-time dosimetric verification in high dose rate prostate brachytherapy
    Seymour, Erin L.
    Downes, Simon J.
    Fogarty, Gerald B.
    Izard, Michael A.
    Metcalfe, Peter
    MEDICAL PHYSICS, 2011, 38 (08) : 4785 - 4794
  • [2] High dose rate intracavitary brachytherapy in the treatment of nasopharyngeal carcinoma
    Leung, TW
    Tung, SY
    Wong, VYW
    Lui, CMM
    Sze, WK
    Cheung, KL
    Lau, WH
    O, SK
    ACTA ONCOLOGICA, 1996, 35 (01) : 43 - 47
  • [3] Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera
    Duan, J
    Macey, DJ
    Pareek, PN
    Brezovich, IA
    MEDICAL PHYSICS, 2001, 28 (02) : 167 - 173
  • [4] Real-time in vivo dosimetry in high dose rate prostate brachytherapy
    Mason, Josh
    Mamo, Arielle
    AI-Qaisieh, Bashar
    Henry, Ann M.
    Bownes, Peter
    RADIOTHERAPY AND ONCOLOGY, 2016, 120 (02) : 333 - 338
  • [5] INTRACAVITARY RADIATION TREATMENT OF NASOPHARYNGEAL CARCINOMA BY THE HIGH-DOSE RATE AFTERLOADING TECHNIQUE
    ZHANG, YW
    LIU, TF
    FI, CX
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 16 (02): : 315 - 318
  • [6] Real-Time Source and Plan Verification for High-Dose Rate (HDR) Brachytherapy Using Flat Panel Detector
    Rong, Y.
    Hernandez, D.
    Zhang, J.
    Benedict, S.
    Lu, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E424 - E424
  • [7] High-dose rate intracavitary brachytherapy pretreatment dwell position verification using a transparent applicator
    Otani, Yuki
    Sumida, Iori
    Nose, Takayuki
    Shimamoto, Shigetoshi
    Okubo, Hirofumi
    Ogawa, Kazuhiko
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 428 - 434
  • [8] Vaginal Applicators in High-Dose Rate Brachytherapy for Endometrial Carcinoma: A Dosimetric Study
    Khan, Aijaz Ahmad
    Vijay, Anuj
    Wani, Shaqul Qamar
    Haq, Malik Mohib-ul
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2023, 38 (02): : 191 - 200
  • [9] In vivo dose verification method in catheter based high dose rate brachytherapy
    Jaselske, Evelina
    Adliene, Diana
    Rudzianskas, Viktoras
    Urbonavicius, Benas Gabrielis
    Inciura, Arturas
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 44 : 1 - 10
  • [10] High-dose-rate intracavitary brachytherapy boost for early T stage nasopharyngeal carcinoma{private}
    Leung, To-Wai
    Wong, Victy Y. W.
    Sze, Wing-Kin
    Luj, Collin M. M.
    Tung, Stewart Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 361 - 367