Biomimetic compliant lower limb exoskeleton (BioComEx) and its experimental evaluation

被引:27
|
作者
Baser, Ozgur [1 ]
Kizilhan, Hasbi [1 ]
Kilic, Ergin [1 ]
机构
[1] Suleyman Demirel Univ, Dept Mech Engn, Isparta, Turkey
关键词
Lower limb exoskeleton; Variable stiffness actuator; Series elastic actuator; Robot-in-charge; Human-in-charge; ROBOT; ACTUATOR; DESIGN; REHABILITATION; WALKING; JOINT;
D O I
10.1007/s40430-019-1729-4
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Exoskeleton robots are generally used for rehabilitation and load-carrying applications. Stable and flexible walking with minimal energy consumption of the human body is achieved by the compliant operation of the human joints. Essentially, the stiffness of the human ankle joint varies continuously, while the stiffness of the knee and hip joints remains nearly constant during loading phases of the walking cycles. With inspiration from the human leg biomechanics, a new biomimetic compliant lower limb exoskeleton robot (BioComEx) was constructed and its preliminary experimental tests were carried out in this paper. A variable stiffness actuator was developed for the ankle joint of BioComEx, and series elastic actuator designs were employed in the knee and hip joints of the robot. The joint actuators of BioComEx were designed longitudinally in order to be embedded inside of the thigh and shank leg segments for the compactness. Separate force sensors were employed in each segment of the robot for the modularity of the control strategies. Since the exoskeletons can be used for both load-carrying and rehabilitation, the developed robot was tested in both human-in-charge and robot-in-charge modes, respectively. The robot needs to mimic movement of healthy user joints in the human-in-charge mode and maximize the compliance between the user and robot; thus, the interaction forces should be reduced as possible. Hence, a modular closed-loop impedance control algorithm was developed for this mode. On the other hand, in the robot-in-charge mode, the robot joints need to track predefined gait references to make the patient walk. PID position control algorithm was chosen for preliminary test of the robot in this mode. Experimental results show that BioComEx is sufficiently satisfactory for walking applications of healthy and paralyzed users.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Biomimetic compliant lower limb exoskeleton (BioComEx) and its experimental evaluation
    Ozgur Baser
    Hasbi Kizilhan
    Ergin Kilic
    Journal of the Brazilian Society of Mechanical Sciences and Engineering, 2019, 41
  • [2] Mechanical Design of a Biomimetic Compliant Lower Limb Exoskeleton (BioComEx)
    Baser, Ozgur
    Kizilhan, Hasbi
    Kilic, Ergin
    2016 IEEE INTERNATIONAL CONFERENCE ON AUTONOMOUS ROBOT SYSTEMS AND COMPETITIONS (ICARSC 2016), 2016, : 60 - 65
  • [3] Design of a Compliant Joint Actuator for Lower-limb Exoskeleton Robot
    Huang, Chao
    Chen, Weihai
    Liu, Jingmeng
    Zhang, Jianbin
    PROCEEDINGS OF THE 2017 12TH IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS (ICIEA), 2017, : 1522 - 1527
  • [4] IMPLEMENTATION OF AN ADJUSTABLE COMPLIANT KNEE JOINT IN A LOWER-LIMB EXOSKELETON
    Sanz-Merodio, D.
    Cestari, M.
    Arevalo, J. C.
    Garcia, E.
    NATURE INSPIRED MOBILE ROBOTICS, 2013, : 223 - 231
  • [5] Lower-Limb Exoskeleton With Compliant Actuators: Human Cooperative Control
    Bergmann, Lukas
    Voss, Daniel
    Leonhardt, Steffen
    Ngo, Chuong
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2023, 5 (03): : 717 - 729
  • [6] Compliant Control of Lower Limb Rehabilitation Exoskeleton Robot Based on Flexible Transmission
    Liu, Keping
    Li, Li
    Li, Wanting
    Gu, Jian
    Sun, Zhongbo
    JOURNAL OF BIONIC ENGINEERING, 2023, 20 (03) : 1021 - 1035
  • [7] Compliant Control of Lower Limb Rehabilitation Exoskeleton Robot Based on Flexible Transmission
    Keping Liu
    Li Li
    Wanting Li
    Jian Gu
    Zhongbo Sun
    Journal of Bionic Engineering, 2023, 20 : 1021 - 1035
  • [8] Lower Limb Exoskeleton With Compliant Actuators: Design, Modeling, and Human Torque Estimation
    Bergmann, Lukas
    Lueck, Oliver
    Voss, Daniel
    Buschermoehle, Philipp
    Liu, Lin
    Leonhardt, Steffen
    Ngo, Chuong
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2023, 28 (02) : 758 - 769
  • [9] Biomimetic Viscoelastic Compliance Control for Self-Balancing Lower Limb Exoskeleton
    Tian, Dingkui
    Wang, Wanxiang
    Li, Feng
    Chen, Ziqiang
    He, Yong
    Li, Jinke
    Zhang, Li
    Wu, Xinyu
    IEEE TRANSACTIONS ON INDUSTRIAL INFORMATICS, 2024, : 13502 - 13512
  • [10] Design and Evaluation of a Modular Lower Limb Exoskeleton for Rehabilitation
    dos Santos, Wilian M.
    Nogueira, Samuel L.
    de Oliveira, Gustavo C.
    Pena, Guido G.
    Siqueira, Adriano A. G.
    2017 INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS (ICORR), 2017, : 447 - 451