Evaluation of a vertical piezoelectric transducer stage using a large range metrological atomic force microscope

被引:0
|
作者
Wang, Shihua [1 ]
Tan, Siew Leng [1 ]
Xu, Gan [1 ]
机构
[1] Agcy Sci Technol & Res, Natl Metrol Ctr, Singapore 118221, Singapore
来源
关键词
atomic force microscope; large range metrological atomic force microscope; displacement calibration; traceability; PZT stage; TRACEABLE CALIBRATION; SYSTEM; INTERFEROMETER; ACCURACY; AFM;
D O I
10.1117/1.JMM.11.1.011005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A method is described for the calibration of vertical piezoelectric transducer (PZT) stages using a large range metrological atomic force microscope (LRM-AFM). A vertical PZT stage is mounted onto the system and an optical-flat sample is attached on the top of the PZT stage. The AFM probe is operated in contact mode and used as a null indicator to measure the movement of the optical-flat as it is driven by the PZT stage. As the PZT stage is scanned vertically, the AFM probe is in contact with the test surface without lateral scanning. The displacement of the vertical stage of the LRM-AFM is measured by the integrated laser interferometer, and the corresponding laser interferometer readings and the nominal position of the PZT stage are recorded. This body of data is then used to establish the relationship between the laser interferometer reading and the nominal displacement of the PZT stage. Our results show that this method can be used to calibrate PZT stages up to a range of 250 mu m with an expanded uncertainty of less than 5 nm. (C) 2012 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.JMM.11.1.011005]
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Design of a sample approach mechanism for a metrological atomic force microscope
    Piot, J.
    Qian, J.
    Piree, H.
    Kotte, G.
    Petry, J.
    Kruth, J. -P.
    Vanherck, P.
    Van Haesendonck, C.
    Reynaerts, D.
    MEASUREMENT, 2013, 46 (01) : 739 - 746
  • [32] Development and Metrological Evaluation of a Force Transducer for Industrial Application
    Singh, Ajay Pratap
    Ghoshal, Sanjoy K.
    Kumar, Harish
    Yoon, Jonghun
    Yadav, Poonam
    IEEE ACCESS, 2021, 9 : 33299 - 33312
  • [33] Metrological large range magnetic force microscopy
    Dai, Gaoliang
    Hu, Xiukun
    Sievers, Sibylle
    Scarioni, Alexander Fernandez
    Neu, Volker
    Fluegge, Jens
    Schumacher, Hans Werner
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (09):
  • [34] Piezoelectric measurements of granite as composite material using atomic force microscope
    Matsuda, T
    Yamanaka, C
    Ikeya, M
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2005, 44 (02): : 968 - 971
  • [35] Linewidth calibration using a metrological atomic force microscope with a tip-tilting mechanism
    Kizu, Ryosuke
    Misumi, Ichiko
    Hirai, Akiko
    Kinoshita, Kazuto
    Gonda, Satoshi
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2019, 30 (01)
  • [36] Piezoelectric measurements of granite as composite material using atomic force microscope
    Matsuda, Tomonori
    Yamanaka, Chihiro
    Ikeya, Motoji
    Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 2005, 44 (02): : 968 - 971
  • [37] Modeling and control of an atomic force microscope using a piezoelectric tuning fork for force sensing
    Yeh, T. -J.
    Wang, Chai-De
    Wu, Ting-Ying
    SIMULATION MODELLING PRACTICE AND THEORY, 2008, 16 (07) : 768 - 782
  • [38] Feedforward control of a closed-loop piezoelectric translation stage for atomic force microscope
    Li, Yang
    Bechhoefer, John
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (01):
  • [39] A standard used for probe-tip diameter evaluation in surface roughness measurements using metrological atomic force microscope
    Misumi, Ichiko
    Kizu, Ryosuke
    Sugawara, Kentaro
    Hirai, Akiko
    Gonda, Satoshi
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2020, 31 (09)
  • [40] Auto leveling control technique for large-range atomic force microscope
    Chen, Daixie
    Yin, Bohua
    Lin, Yunsheng
    Chu, Mingzhang
    Han, Li
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2011, 32 (01): : 225 - 229