Measurements and simulations analysing the noise behaviour of grating-based X-ray phase-contrast imaging

被引:10
|
作者
Weber, T. [1 ]
Bartl, P. [1 ]
Durst, J. [1 ]
Haas, W. [1 ,2 ]
Michel, T. [1 ]
Ritter, A. [1 ]
Anton, G. [1 ]
机构
[1] Univ Erlangen Nurnberg, ECAP, D-91058 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Pattern Recognit Lab, D-91058 Erlangen, Germany
关键词
Phase contrast; Grating interferometry; X-ray imaging; Photon-counting detector; Medipix; Probability density function; Noise;
D O I
10.1016/j.nima.2010.11.060
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the last decades, phase-contrast imaging using a Talbot-Lau grating interferometer is possible even with a low-brilliance X-ray source. With the potential of increasing the soft-tissue contrast, this method is on its way into medical imaging. For this purpose, the knowledge of the underlying physics of this technique is necessary. With this paper, we would like to contribute to the understanding of grating-based phase-contrast imaging by presenting results on measurements and simulations regarding the noise behaviour of the differential phases. These measurements were done using a microfocus X-ray tube with a hybrid, photon-counting, semiconductor Medipix2 detector. The additional simulations were performed by our in-house developed phase-contrast simulation tool "SPHINX", combining both wave and particle contributions of the simulated photons. The results obtained by both of these methods show the same behaviour. Increasing the number of photons leads to a linear decrease of the standard deviation of the phase. The number of used phase steps has no influence on the standard deviation, if the total number of photons is held constant. Furthermore, the probability density function (pdf) of the reconstructed differential phases was analysed. It turned out that the so-called von Mises distribution is the physically correct pdf, which was also confirmed by measurements. This information advances the understanding of grating-based phase-contrast imaging and can be used to improve image quality. (c) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:S273 / S275
页数:3
相关论文
共 50 条
  • [1] Noise in x-ray grating-based phase-contrast imaging
    Weber, Thomas
    Bartl, Peter
    Bayer, Florian
    Durst, Juergen
    Haas, Wilhelm
    Michel, Thilo
    Ritter, Andre
    Anton, Gisela
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4133 - 4140
  • [2] Spectroscopic Measurements Concerning Grating-Based X-Ray Phase-Contrast Imaging
    Weber, Thomas
    Bartl, Peter
    Bayer, Florian
    Durst, Juergen
    Haas, Wilhelm
    Michel, Thilo
    Pelzer, Georg
    Ritter, Andre
    Anton, Gisela
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [3] Contrast transfer function in grating-based x-ray phase-contrast imaging
    Huang, Jianheng
    Du, Yang
    Lin, Danying
    Liu, Xin
    Niu, Hanben
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 747 : 13 - 18
  • [4] Grating-based X-ray phase-contrast imaging at PETRA III
    Hipp, A.
    Beckmann, F.
    Lytaev, P.
    Greving, I.
    Lottermoser, L.
    Dose, T.
    Kirchhof, R.
    Burmester, H.
    Schreyer, A.
    Herzen, J.
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY IX, 2014, 9212
  • [5] Energy weighting in grating-based x-ray phase-contrast imaging
    Pelzer, Georg
    Weber, Thomas
    Anton, Gisela
    Sune, Rafael Ballabriga
    Bayer, Florian
    Campbell, Michael
    Haas, Wilhelm
    Horn, Florian
    Cudie, Xavi Llopart
    Michel, Norbert
    Mollenbauer, Uwe
    Rieger, Jens
    Ritter, Andre
    Ritter, Ina
    Woelfel, Stefan
    Wong, Winnie S.
    Zang, Andrea
    Michel, Thilo
    [J]. MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [6] Simulation and measurement of grating-based X-ray phase-contrast imaging
    Bartl, P.
    Bayer, F.
    Durst, J.
    Haas, W.
    Michel, T.
    Pelzer, G.
    Ritter, A.
    Weber, T.
    Anton, G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S269 - S272
  • [7] Inverse geometry for grating-based x-ray phase-contrast imaging
    Donath, Tilman
    Chabior, Michael
    Pfeiffer, Franz
    Bunk, Oliver
    Reznikova, Elena
    Mohr, Juergen
    Hempel, Eckhard
    Popescu, Stefan
    Hoheisel, Martin
    Schuster, Manfred
    Baumann, Joachim
    David, Christian
    [J]. JOURNAL OF APPLIED PHYSICS, 2009, 106 (05)
  • [8] Analysis of partial coherence in grating-based phase-contrast X-ray imaging
    Wang, Zhili
    Liu, Xiaosong
    Zhu, Peiping
    Huang, Wanxia
    Yuan, Qingxi
    Li, Enrong
    Liu, Yijin
    Zhang, Kai
    Hong, Youli
    Wu, Ziyu
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3): : 319 - 322
  • [9] Beam hardening effects in grating-based x-ray phase-contrast imaging
    Chabior, Michael
    Donath, Tilman
    David, Christian
    Bunk, Oliver
    Schuster, Manfred
    Schroer, Christian
    Pfeiffer, Franz
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1189 - 1195
  • [10] Noise analysis of grating-based x-ray differential phase-contrast imaging with angular signal radiography
    Faiz, Wali
    Bao, Yuan
    Gao, Kun
    Wu, Zhao
    Wei, Chen-Xi
    Zan, Gui-Bin
    Zhu, Pei-Ping
    Tian, Yang-Chao
    [J]. CHINESE PHYSICS B, 2017, 26 (04)