An optical tracker based robot registration and servoing method for ultrasound guided percutaneous renal access

被引:20
|
作者
Zhang, Dongwen [1 ,2 ]
Li, Zhicheng [1 ]
Chen, Ken [1 ]
Xiong, Jing [1 ]
Zhang, Xuping [1 ]
Wang, Lei [1 ]
机构
[1] Chinese Acad Sci, Shenzhen Inst Adv Technol, Key Lab Hlth Informat, Shenzhen Key Lab Lowcost Healthcare, Shenzhen 518055, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
ABLATION;
D O I
10.1186/1475-925X-12-47
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Robot-assisted needle steering facilitates the percutaneous renal access (PRA) for their accuracy and consistency over manual operation. However, inaccurate image-robot correspondence and uncertainties in robot parameters make the needle track deviate from the intrarenal target. This paper aims to simplify the image-tracker-robot registration procedure and improves the accuracy of needle alignment for robot assisted ultrasound-guided PRA. Methods: First, a semi-automatic rigid registration is used for the alignment of the preoperative MR volume and the intraoperative orthogonal US slices. Passive markers are mounted both on US probe and robot end-effector, the planned puncture path is transferred from the MR volume frame into optical tracker frame. Tracker-robot correspondence and robot calibration are performed iteratively using a simplified scheme, both position and orientation information are incorporated to estimate the transformation matrix, only several key structural robot parameters and joint zero positions are calibrated for simplicity in solving the inverse kinematic. Furthermore, an optical tracker feedback control is designed for compensating inaccuracies in robot parameters and tracker-robot correspondence, and improving the accuracy of needle alignment. The intervention procedure was implemented by a telemanipulated 5R1P robot, two experiments were conducted to validate the efficiency of robot-tracker registration method and the optical tracker feedback control, robot assisted needle insertion experiment was conducted on kidney phantom to evaluate the system performance. Results: The relative positioning accuracy of needle alignment is 0.24 +/- 0.08 mm, the directional accuracy is 6.78 +/- 1.65 x 10(-4)rad; the needle-target distance of needle insertion is 2.15 +/- 0. 17 mm. The optical tracker feedback control method performs stable against wide range of angular disturbance over (0 similar to 0.4) radians, and the length disturbance over (0 similar to 100) mm. Conclusions: The proposed optical tracker based robot registration and servoing method is capable of accurate three dimension needle operation for PRA procedure with improved precision and shortened time.
引用
下载
收藏
页数:16
相关论文
共 50 条
  • [21] Endoscopically guided percutaneous renal access: "Seeing is believing"
    Khan, Farhan
    Borin, James F.
    Pearle, Margaret S.
    McDougall, Elspeth M.
    Clayman, Ralph V.
    JOURNAL OF ENDOUROLOGY, 2006, 20 (07) : 451 - 455
  • [22] Fluoroscopy guided percutaneous renal access in prone position
    Gyanendra R Sharma
    Pankaj N Maheshwari
    Anshu G Sharma
    Reeta P Maheshwari
    Ritwik S Heda
    Sakshi P Maheshwari
    World Journal of Clinical Cases, 2015, (03) : 245 - 264
  • [23] AN IMPROVED TECHNIQUE FOR ULTRASOUND GUIDED PERCUTANEOUS RENAL BIOPSY
    BIRNHOLZ, JC
    KASINATH, BS
    CORWIN, HL
    KIDNEY INTERNATIONAL, 1985, 27 (01) : 80 - 82
  • [24] ULTRASOUND-GUIDED PERCUTANEOUS RENAL BIOPSY IN PATIENTS WITH RENAL IMPAIRMENT
    BACH, D
    HANRATHS, M
    GRABENSEE, B
    ULTRASCHALL IN DER MEDIZIN, 1987, 8 (03): : 149 - 151
  • [25] A NOVEL PREOPERATIVE SCORING SYSTEM FOR PREDICTING SUCCESSFUL RENAL ACCESS FOR ULTRASOUND-GUIDED PERCUTANEOUS NEPHROLITHOTOMY
    Rompsaithong, Ukrit
    Yang, Heiko
    Sui, Wilson
    Unno, Rei
    Bayne, David
    Ahn, Justin
    Stoller, Marshall
    Chi, Thomas
    JOURNAL OF UROLOGY, 2023, 209 : E474 - E475
  • [26] Middle calyx access is better for single renal pelvic stone in ultrasound-guided percutaneous nephrolithotomy
    Yan Song
    Wei Jin
    Shengyu Hua
    Xiang Fei
    Urolithiasis, 2016, 44 : 459 - 463
  • [27] Developing an Ex Vivo Pig Model for Teaching Ultrasound and Fluoroscopy-Guided Percutaneous Renal Access
    Du, Katie
    Tong, Steve
    De, Shubha
    JOURNAL OF ENDOUROLOGY, 2024, 38 (06) : 642 - 648
  • [28] Ultrasound-guided percutaneous ablation of a renal mass in a renal allograft
    Hruby, Gregory W.
    Fine, Joshua K.
    Landman, Jaime
    UROLOGY, 2006, 68 (04) : 891.e5 - 891.e6
  • [29] Middle calyx access is better for single renal pelvic stone in ultrasound-guided percutaneous nephrolithotomy
    Song, Yan
    Jin, Wei
    Hua, Shengyu
    Fei, Xiang
    UROLITHIASIS, 2016, 44 (05) : 459 - 463
  • [30] Method of robot calibration based on laser tracker
    State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China
    Jixie Gongcheng Xuebao, 2007, 9 (195-200): : 195 - 200