An optical tracker based robot registration and servoing method for ultrasound guided percutaneous renal access

被引:20
|
作者
Zhang, Dongwen [1 ,2 ]
Li, Zhicheng [1 ]
Chen, Ken [1 ]
Xiong, Jing [1 ]
Zhang, Xuping [1 ]
Wang, Lei [1 ]
机构
[1] Chinese Acad Sci, Shenzhen Inst Adv Technol, Key Lab Hlth Informat, Shenzhen Key Lab Lowcost Healthcare, Shenzhen 518055, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
ABLATION;
D O I
10.1186/1475-925X-12-47
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Robot-assisted needle steering facilitates the percutaneous renal access (PRA) for their accuracy and consistency over manual operation. However, inaccurate image-robot correspondence and uncertainties in robot parameters make the needle track deviate from the intrarenal target. This paper aims to simplify the image-tracker-robot registration procedure and improves the accuracy of needle alignment for robot assisted ultrasound-guided PRA. Methods: First, a semi-automatic rigid registration is used for the alignment of the preoperative MR volume and the intraoperative orthogonal US slices. Passive markers are mounted both on US probe and robot end-effector, the planned puncture path is transferred from the MR volume frame into optical tracker frame. Tracker-robot correspondence and robot calibration are performed iteratively using a simplified scheme, both position and orientation information are incorporated to estimate the transformation matrix, only several key structural robot parameters and joint zero positions are calibrated for simplicity in solving the inverse kinematic. Furthermore, an optical tracker feedback control is designed for compensating inaccuracies in robot parameters and tracker-robot correspondence, and improving the accuracy of needle alignment. The intervention procedure was implemented by a telemanipulated 5R1P robot, two experiments were conducted to validate the efficiency of robot-tracker registration method and the optical tracker feedback control, robot assisted needle insertion experiment was conducted on kidney phantom to evaluate the system performance. Results: The relative positioning accuracy of needle alignment is 0.24 +/- 0.08 mm, the directional accuracy is 6.78 +/- 1.65 x 10(-4)rad; the needle-target distance of needle insertion is 2.15 +/- 0. 17 mm. The optical tracker feedback control method performs stable against wide range of angular disturbance over (0 similar to 0.4) radians, and the length disturbance over (0 similar to 100) mm. Conclusions: The proposed optical tracker based robot registration and servoing method is capable of accurate three dimension needle operation for PRA procedure with improved precision and shortened time.
引用
下载
收藏
页数:16
相关论文
共 50 条
  • [1] An optical tracker based robot registration and servoing method for ultrasound guided percutaneous renal access
    Dongwen Zhang
    Zhicheng Li
    Ken Chen
    Jing Xiong
    Xuping Zhang
    Lei Wang
    BioMedical Engineering OnLine, 12
  • [2] ULTRASOUND GUIDED PERCUTANEOUS RENAL ACCESS
    Patel, Manish
    Gutierrez, Jorge
    JOURNAL OF UROLOGY, 2016, 195 (04): : E659 - E659
  • [3] ULTRASOUND-GUIDED SUPRACOSTAL PERCUTANEOUS RENAL ACCESS
    Ahn, Justin
    Bayne, David
    Stoller, Marshall
    Chi, Tom
    JOURNAL OF UROLOGY, 2020, 203 : E387 - E387
  • [4] AUGMENTED REALITY SIMULATOR FOR ULTRASOUND GUIDED PERCUTANEOUS RENAL ACCESS
    Wang, Zhan Tao
    Mu, Yanyu
    Hocking, David
    Eagleson, Roy
    Peters, Terry
    JOURNAL OF UROLOGY, 2020, 203 : E691 - E692
  • [5] An Optical Tracker Based Registration Method Using Feedback for Robot-Assisted Insertion Surgeries
    Li, Zhuo
    Liu, Xingtong
    Xie, Xiang
    Li, Guolin
    Mai, Songping
    Wang, Zhihua
    2017 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2017, : 670 - 673
  • [6] Augmented reality simulator for ultrasound-guided percutaneous renal access
    Yanyu Mu
    David Hocking
    Zhan Tao Wang
    Gregory J. Garvin
    Roy Eagleson
    Terry M. Peters
    International Journal of Computer Assisted Radiology and Surgery, 2020, 15 : 749 - 757
  • [7] Augmented reality simulator for ultrasound-guided percutaneous renal access
    Mu, Yanyu
    Hocking, David
    Wang, Zhan Tao
    Garvin, Gregory J.
    Eagleson, Roy
    Peters, Terry M.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2020, 15 (05) : 749 - 757
  • [8] Effective evaluation of using ultrasound-guided renal access for percutaneous nephrolithotomy
    Pham Hong Duc
    Tran Vinh Hung
    Hung Quang Huy
    Nguyen Thuy Linh
    Bui Van Giang
    MEDICAL SCIENCE, 2020, 24 (104) : 2195 - 2199
  • [9] Endoscopic guided percutaneous renal access
    Borin, JF
    Khan, F
    McDougall, EM
    Clayman, RV
    JOURNAL OF UROLOGY, 2006, 175 (04): : 393 - 394
  • [10] Ultrasound-Guided Renal Access for Percutaneous Nephrolithotomy: A Description of Three Novel Ultrasound-Guided Needle Techniques
    Chu, Carissa
    Masic, Selma
    Usawachintachit, Manint
    Hu, Weiguo
    Yang, Wenzeng
    Stoller, Marshall
    Li, Jianxing
    Chi, Thomas
    JOURNAL OF ENDOUROLOGY, 2016, 30 (02) : 153 - 158