Does VMAT for treatment of NSCLC patients increase the risk of pneumonitis compared to IMRT? - A planning study

被引:19
|
作者
Bertelsen, Anders [1 ,2 ]
Hansen, Olfred [3 ]
Brink, Carsten [1 ,2 ]
机构
[1] Univ So Denmark, Inst Clin Res, Odense, Denmark
[2] Odense Univ Hosp, Lab Radiat Phys, DK-5000 Odense, Denmark
[3] Odense Univ Hosp, Dept Oncol, DK-5000 Odense, Denmark
关键词
DOSE CALCULATION ALGORITHMS; 3-DIMENSIONAL CONFORMAL RADIOTHERAPY; CONCURRENT CHEMOTHERAPY; LUNG; THERAPY; PHOTON; IRRADIATION; TOXICITY; HEART; ARC;
D O I
10.3109/0284186X.2011.648341
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. Volumetric modulated arc therapy (VMAT) for treatment of non-small cell lung cancer (NSCLC) patients potentially changes the risk of radiation-induced pneumonitis (RP) compared to intensity modulated radiation therapy (IMRT) if the dose to the healthy lung is changed significantly. In this study, clinical IMRT plans were used as starting point for VMAT optimization and differences in risk estimates of RP between the two plan types were evaluated. Material and methods. Fifteen NSCLC patients prescribed 66 Gy in 2 Gy fractions were planned with IMRT and subsequently with single arc VMAT. Dose metrics were evaluated for target and lung together with population averaged dose volume histograms. The risk of RP was calculated using normal tissue complication probability (NTCP) models. Finally, applicability of the plans was tested through delivery on an Elekta accelerator. Results. When changing from IMRT to VMAT only modest differences were observed in the dose to the lung and target volume. On average, fractions of lung irradiated to doses between 18 Gy and 48 Gy were statistically signifi cant reduced using VMAT compared to IMRT. For the fraction of lung receiving more than 20 Gy the reduction was 1.2% percentage points: (range -0.6 2.6%). The evaluated toxicity were smaller with VMAT compared to IMRT, however only modest differences were observed in the NTCP values. The plans were delivered without any problems. The average beam on time with VMAT was 83 s. This was a reduction of 141 s (ranging from 37 s to 216 s) compared to IMRT. Conclusions. Using IMRT as reference for the VMAT optimization it was possible to implement VMAT in the clinic with no increase in estimated risk of RP. Thus, toxicity is not expected to be a hindrance to using VMAT and will profit from the shorter delivery time with VMAT compared to IMRT.
引用
收藏
页码:752 / 758
页数:7
相关论文
共 50 条
  • [1] VMAT for Unresectable Locally Advanced NSCLC Does Not Increase the Risk of Radiation Pneumonitis Compared with IMRT
    Zhang, T.
    Zhou, Z.
    Bi, N.
    Wang, J.
    Wang, L.
    Deng, L.
    Feng, Q.
    Liang, J.
    Xiao, Z.
    Chen, D.
    Lv, J.
    Wang, X.
    Wang, W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E543 - E543
  • [2] Dosimetric Comparison of IMRT and VMAT for Retinoblastoma Treatment: A treatment Planning Study
    Patil, Deepali
    Madhawi, Richa
    Zope, Mukesh K. U. M. A. R.
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2025, 40 (01): : 51 - 59
  • [3] Automatic Treatment Planning for IMRT, VMAT and IMPT
    Zhang, X.
    MEDICAL PHYSICS, 2012, 39 (06) : 4008 - 4008
  • [4] Tangential VMAT and IMRT for WBI compared using automated multi-criterial treatment planning
    Redapi, L.
    Rossi, L.
    Marrazzo, L.
    Penninkhof, J. J.
    Pallotta, S.
    Heijmen, B.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S579 - S580
  • [5] Difference in the Incidence of RP in Elderly NSCLC Patients between VMAT & IMRT: a Prospective Study
    Fang, Rongyao
    Xiao, Ningping
    Lu, Wenhui
    Abdilim, Albert
    Pei, Zhengda
    Yu, Tingjie
    Xiang, Jiayu
    Zhang, Yihui
    Dai, Jie
    Jin, Yi
    Jin, Hekun
    Yang, Pei
    RADIOTHERAPY AND ONCOLOGY, 2024, 197 : S406 - S407
  • [6] Automated VMAT Treatment Planning for Stage III Lung Cancer: How Does It Compare with IMRT?
    Quan, E.
    Chang, J.
    Liao, Z.
    Xia, T.
    Liu, H.
    Li, X.
    Balter, P.
    Zhang, X.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [7] Fully automated VMAT treatment planning for advanced-stage NSCLC patients
    Della Gala, Giuseppe
    Dirkx, Maarten L. P.
    Hoekstra, Nienke
    Fransen, Dennie
    Lanconelli, Nico
    van de Pol, Marjan
    Heijmen, Ben J. M.
    Petit, Steven F.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 (05) : 402 - 409
  • [8] Automatic IMRT and VMAT Treatment Planning for Head and Neck Cancer
    Shen, Z. L.
    Murray, E.
    Yu, N.
    Kolar, M. D.
    Greskovich, J. F., Jr.
    Koyfman, S.
    Bzdusek, K.
    Xia, P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E560 - E561
  • [9] Comparison of Treatment Planning and Delivery Performance of VMAT versus IMRT
    Bergman, A.
    Vollans, E.
    Duzenli, C.
    MEDICAL PHYSICS, 2012, 39 (07) : 4630 - 4630
  • [10] Progressive Knowledge Modeling for Pelvic IMRT/VMAT Treatment Planning
    Lu, S.
    Yuan, L.
    Craciunescu, O.
    Ge, Y.
    Yin, F.
    Wu, Q.
    MEDICAL PHYSICS, 2014, 41 (06) : 460 - 460