Influence of the implant abutment types and the dynamic loading on initial screw loosening

被引:15
|
作者
Kim, Eun-Sook [1 ]
Shin, Soo-Yeon [1 ]
机构
[1] Dankook Univ, Coll Dent, Dept Prosthodont, Cheonan, South Korea
来源
JOURNAL OF ADVANCED PROSTHODONTICS | 2013年 / 5卷 / 01期
关键词
Computer-aided design; Dental implantation; Prosthesis failure; Removal torque; Screw loosening; INTERNAL CONICAL INTERFACE; 3-YEAR FOLLOW-UP; DENTAL IMPLANTS; FRAMEWORKS; SYSTEMS; FIT; CONNECTION; STRENGTH; DESIGN;
D O I
10.4047/jap.2013.5.1.21
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS. Three groups of abutments were produced using different types of fabrication methods; stock abutment, gold cast abutment, and CAD/CAM custom abutment. A customized jig was fabricated to apply the load at 300 to the long axis. The implant fixtures were fixed to the jig, and connected to the abutments with a 30 Ncm tightening torque. A sine curved dynamic load was applied for 10 cycles between 25 and 250 N at 14 Hz. Removal torque before loading and after loading were evaluated. The SPSS was used for statistical analysis of the results. A Kruskal-Wallis test was performed to compare screw loosening between the abutment systems. A Wilcoxon signed-rank test was performed to compare screw loosening between before and after loading in each group (alpha=0.05). RESULTS. Removal torque value before loading and after loading was the highest in stock abutment, which was then followed by gold cast abutment and CAD/CAM custom abutment, but there were no significant differences. CONCLUSION. The abutment types did not have a significant influence on short term screw loosening. On the other hand, after 105 cycles dynamic loading, CAD/CAM custom abutment affected the initial screw loosening, but stock abutment and gold cast abutment did not. [J Adv Prosthodont 2013;5:21-8]
引用
收藏
页码:21 / 28
页数:8
相关论文
共 50 条
  • [31] Zirconium implant abutments: Fracture strength and influence of cyclic loading on retaining-screw loosening
    Gehrke, P
    Dhom, G
    Brunner, J
    Wolf, D
    Degidi, M
    Piattelli, D
    QUINTESSENCE INTERNATIONAL, 2006, 37 (01): : 19 - 26
  • [32] Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit
    Velez, Jorge
    Pelaez, Jesus
    Lopez-Suarez, Carlos
    Agustin-Panadero, Ruben
    Tobar, Celia
    Suarez, Maria J.
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (08) : 1 - 13
  • [33] Influence of Abutment Surface Treatments on Screw Loosening of Morse Taper Implants
    Asanuma Hirayama, Patricia Mitiko
    Lima Bohner, Lauren Oliveira
    Marotti, Juliana
    Steagall, Washington, Jr.
    Lagana, Dalva Cruz
    Tortamano, Pedro
    IMPLANT DENTISTRY, 2017, 26 (05) : 718 - 722
  • [34] Preventing loosening of implant abutment screws
    Cavazos, E
    Bell, FA
    JOURNAL OF PROSTHETIC DENTISTRY, 1996, 75 (05): : 566 - 569
  • [35] Effect of dynamic cyclic loading on screw loosening of retightened versus new abutment screw in both narrow and standard implants (in-vitro study)
    Attiah, Eman Mohammed Nasr
    AlGendy, Attiah Ali
    Mostafa, Tamer Mohamed Nasr
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2020, 6 (01)
  • [36] Effect of dynamic cyclic loading on screw loosening of retightened versus new abutment screw in both narrow and standard implants (in-vitro study)
    Eman Mohammed Nasr Attiah
    Attiah Ali AlGendy
    Tamer Mohamed Nasr Mostafa
    International Journal of Implant Dentistry, 6
  • [37] In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening
    Kose, Onur Dincer
    Karatasli, Burcin
    Demircan, Sabit
    Kose, Taha Emre
    Cene, Erhan
    Aya, Serhan Aydjn
    Erdem, Mehmet Ali
    Cankaya, Abdulkadir Burak
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [38] Abutment screw loosening in implants: A literature review
    Alsubaiy, Ebrahim F.
    JOURNAL OF FAMILY MEDICINE AND PRIMARY CARE, 2020, 9 (11) : 5490 - 5494
  • [39] Success of Dental Implant Influenced by Abutment Types and Loading Protocol
    Mayuri, Sneha
    Irfan, Ahamed K. A.
    Raj, Ritu
    Sen, Amrita
    Malik, Rajvir
    Bandgar, Subhash
    Rangari, Priyadarshini
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2022, 14 : 1019 - 1022
  • [40] An abutment screw loosening study of a Diamond Like Carbon-coated CP titanium implant
    Kim, SK
    Lee, JB
    Koak, JY
    Heo, SJ
    Lee, KR
    Cho, LR
    Lee, SS
    JOURNAL OF ORAL REHABILITATION, 2005, 32 (05) : 346 - 350