What is inverse-geometry CT?

被引:7
|
作者
Schmidt, Taly Gilat [1 ]
机构
[1] Marquette Univ, Dept Biomed Engn, Milwaukee, WI 53201 USA
关键词
Volumetric CT; Inverse geometry; CT systems; X-RAY SCATTER; SYSTEM; TOMOGRAPHY; PROTOTYPE; MAGNITUDE;
D O I
10.1016/j.jcct.2011.04.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inverse-geometry computed tomography (IGCT) systems are being developed to provide improved volumetric imaging. In conventional multislice CT systems, x-rays are emitted from a small area and irradiate a large-area detector. In an IGCT system, x-ray sources are distributed over a large area, with each beam irradiating a small-area detector. Therefore, in the inverse geometry, a series of narrow x-ray beams are switched on and off while the gantry rotates. In conventional CT geometry, cone-beam and scatter artifacts increase with the imaged volume thickness. An inverse geometry may be less susceptible to scatter effects, because only a fraction of the field of view is irradiated at one time. The distributed source in the inverse geometry potentially improves sampling, leading to reduced cone-beam artifacts. In the inverse geometry, the tube current may be adjusted separately for each source location, which potentially reduces dose. Multiple IGCT prototypes have been constructed and tested on phantoms. A gantry-based IGCT system with one-second gantry rotation was developed, and images of phantoms and small animals were successfully acquired. Clinical feasibility with acceptable noise levels and scan times has not yet been shown. Overall, results from prototype systems suggest that the inverse geometry will enable imaging of a thick volume (similar to 16 cm) while potentially reducing cone-beam artifacts, scatter effects, and radiation dose. The magnitude of these benefits will depend on the specific IGCT implementation and need to be quantified relative to comparable multislice scanners. (C) 2011 Society of Cardiovascular Computed Tomography. All rights reserved.
引用
收藏
页码:145 / 148
页数:4
相关论文
共 50 条
  • [1] Noise simulations for an inverse-geometry volumetric CT system
    Schmidt, TG
    Fahrig, R
    Pelc, NJ
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 420 - 427
  • [2] Multisource Inverse-Geometry CT Prototype System Integration
    Uribe, Jorge
    Reynolds, Joseph L.
    Inzinna, Louis P.
    Longtin, Randy
    Harrison, Daniel D.
    De Man, Bruno
    Neculaes, Bogdan
    Caiafa, Antonio
    Waters, William
    Frutschy, Kristopher J.
    Senzig, Robert
    Baek, Jongduk
    Pelc, Norbert
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2578 - 2581
  • [3] A volumetric reconstruction algorithm for stationary source inverse-geometry CT
    Hsieh, Scott S.
    Pelc, Norbert J.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [4] Geometry analysis of an inverse-geometry volumetric CT system with multiple detector arrays
    Mazin, S
    Schmidt, TG
    Solomon, E
    Fahrig, R
    Pelc, N
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 320 - 329
  • [5] Dynamic X-ray Flux Modulation of Inverse-Geometry CT
    Shen, Liuxing
    Windsor, Haydon
    Zhou, Shuang
    Jiang, Hao
    Zhang, Tiezhi
    MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [6] A prototype table-top inverse-geometry volumetric CT system
    Schmidt, Taly Gilat
    Star-Lack, Josh
    Bennett, N. Robert
    Mazin, Samuel R.
    Solomon, Edward G.
    Fahrig, Rebecca
    Pelc, Norbert J.
    MEDICAL PHYSICS, 2006, 33 (06) : 1867 - 1878
  • [7] Initial results with a multi-source inverse-geometry CT system
    Baek, Jongduk
    Pelc, Norbert J.
    Deman, Bruno
    Uribe, Jorge
    Harrison, Daniel
    Reynolds, Joseph
    Neculaes, Bogdan
    Inzinna, Louis
    Caiafa, Antonio
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [8] Simulated scatter performance of an inverse-geometry dedicated breast CT system
    Bhagtani, Reema
    Schmidt, Taly Gilat
    MEDICAL PHYSICS, 2009, 36 (03) : 788 - 796
  • [9] Two-dimensional reconstruction algorithm of an inverse-geometry volumetric CT system
    Baek, Jongduk
    Pelc, Norbert J.
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [10] Direct two-dimensional reconstruction algorithm for an inverse-geometry CT system
    Baek, Jongduk
    Pelc, Norbert J.
    MEDICAL PHYSICS, 2009, 36 (02) : 394 - 401