Management of arterial puncture site after catheterization procedures: Evaluating a suture-mediated closure device

被引:94
|
作者
Gerckens, U [1 ]
Cattelaens, N [1 ]
Lampe, EG [1 ]
Grube, E [1 ]
机构
[1] Herzentrum Siegburg, Siegburg, Germany
来源
AMERICAN JOURNAL OF CARDIOLOGY | 1999年 / 83卷 / 12期
关键词
D O I
10.1016/S0002-9149(99)00174-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To overcome the challenge associated with achievement in hemostasis after a catheterization procedure, a suture-based closure device was compared with manual compression in a 600-patient randomized trial. The major study end points included the incidence of vascular complications and the time to ambulation after the procedure. The study included diagnostic or interventional procedures. The suture-mediated closure was performed immediately after the procedure independent of the anticoagulation level, whereas manual compression was performed per hospital protocol with sheath removal relying on normalization of patient's anticoagulation status. A significant reduction in time to achieve hemostasis (7.8 +/- 4.8 vs 19.6 +/- 13.2 minutes, p < 0001) and time to ambulation (4.5 +/- 6.5 vs 17.8 +/- 5 hours, p < 0001) was associated with use of the suture-mediated closure device. The incidence of vascular complications was similar in the overall population (5.7% for suturing device vs 11.3% for compression) or in the interventional patient subset (8.4% for suturing device vs 9.6% for compression). There was a significant reduction in the incidence of vascular complications in the diagnostic procedure subset (4.4% for suturing device vs 12.1% for compression, p < 0.05). Thus, the use of a suture-mediated closure device represents a safe alternative to manual compression. Hemostasis and ambulation can be achieved faster with the suturing device than with manual compression, with a potential reduction in access site complications. (C) 1999 by Excerpta Medico, Inc.
引用
收藏
页码:1658 / 1663
页数:6
相关论文
共 50 条
  • [1] Femoral Artery Occlusion After Using a Suture-Mediated Arterial Puncture Closure Device
    Park, Sung-Min
    Kwak, Young-Tae
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (04) : 522 - 524
  • [2] Suture-mediated closure of the arterial puncture site with a percutaneous vascular surgical device: technical note
    Wiskirchen, J
    Erb, M
    Albes, JM
    Pereira, PL
    Claussen, CD
    Duda, SH
    JOURNAL OF INTERVENTIONAL RADIOLOGY, 1998, 13 (02): : 31 - 33
  • [3] Suture-mediated percutaneous closure of the arterial puncture site after abdominal aortic stent implantation
    Wiskirchen, J
    Duda, SH
    Steger, V
    Schott, U
    Ziemer, G
    Claussen, CD
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 1998, 7 (04) : 321 - 324
  • [4] Initial experience using The Closer: A new device for percutaneous suture-mediated closure of arterial puncture sites
    Hahn, U
    Betsch, A
    Tepe, G
    Koenig, CW
    Wiskirchen, J
    Duda, SH
    RADIOLOGY, 2000, 217 : 380 - 380
  • [5] Suture-mediated arterial access site closure after transfemoral aortic valve implantation
    Kahlert, Philipp
    Al-Rashid, Fadi
    Plicht, Bjoern
    Konorza, Thomas
    Neumann, Till
    Thielmann, Matthias
    Wendt, Daniel
    Erbel, Raimund
    Eggebrecht, Holger
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (02) : E139 - E150
  • [6] Initial experience using Prostar(TM): A new device for percutaneous suture-mediated closure of arterial puncture sites
    Carere, RG
    Webb, JG
    Ahmed, T
    Dodek, AA
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1996, 37 (04): : 367 - 372
  • [7] Percutaneous Approach in Endovascular Aortic Procedures Using a Suture-Mediated Closure Device
    Oikonomou, Kyriakos
    Kvataia, Akaki
    Pfister, Karin
    Zygouridou, Evgenia
    Betz, Thomas
    Schierling, Wilma
    Sachsamanis, Georgios
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (22)
  • [8] Antegrade Femoral Puncture Using a Suture-Mediated Closure Device in Infrainguinal Endovascular Interventions
    Mastrorilli, Davide
    Forcella, Edoardo
    Miccoli, Tommaso
    De Massari, Chiara
    Garriboli, Luca
    ANNALS OF VASCULAR SURGERY, 2025, 112 : 246 - 252
  • [9] Analysis of Determinants for Suture-mediated Closure Device Failure During EVAR Procedures
    Garabet, W.
    Arnautovic, A.
    Meurer, L.
    Mulorz, J.
    Rembe, J. D.
    Duran, M.
    Suess, J. D.
    Schelzig, H.
    Wagenhaeuser, M. U.
    VASCULAR AND ENDOVASCULAR SURGERY, 2024, 58 (02) : 129 - 135
  • [10] Iatrogenic stenosis following suture-mediated closure device
    Tam, J.
    Given, M.
    Lutjen, P.
    Thomson, K.
    Lyon, S.
    AUSTRALASIAN RADIOLOGY, 2007, 51 : B319 - B323