Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing
被引:30
|作者:
Lima-Oliveira, Gabriel
[1
,2
,3
,4
]
Lippi, Giuseppe
[5
]
Salvagno, Gian Luca
[1
]
Montagnana, Martina
[1
]
Picheth, Geraldo
[2
]
Guidi, Gian Cesare
[1
,2
]
机构:
[1] Univ Verona, Lab Clin Biochem, Dept Life & Reprod Sci, I-37100 Verona, Italy
[2] Univ Fed Parana, Dept Med Pathol, Postgrad Program Pharmaceut Sci, BR-80060000 Curitiba, Parana, Brazil
[3] MERCOSUL Sect Comm Clin Anal & Vitro Diagnost CSM, Rio De Janeiro, Brazil
[4] Brazilian Soc Clin Anal Sao Paulo State, Sao Paulo, Brazil
[5] Acad Hosp Parma, Dept Pathol & Lab Med, Clin Chem & Hematol Lab, Parma, Italy
关键词:
blood collection;
plasma samples;
preanalytical variability;
vacuum tubes;
validation process;
BIOLOGICAL VARIATION;
CLINICAL-CHEMISTRY;
QUALITY;
COLLECTION;
STASIS;
D O I:
10.1097/MBC.0b013e32835b72ea
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Sometimes in-vitro diagnostic devices (e. g. blood collection tubes) are not validated before use or when the producer's brand is changed. The aim of this study was to validate five brands of sodium citrate vacuum tubes. Blood specimens from 50 volunteers were collected in five different tube brands (I: Venosafe, II: VACUETTE, III: BD Vacutainer, IV: LABOR IMPORT and V: S-Monovette). Routine coagulation tests [activated partial thromboplastin time (aPTT), prothrombin time (PT), and fibrinogen (FIB)] were performed on ACL TOP instrument using HemosIL reagents. The significance of the differences between samples was assessed by paired Student's t-test, set at P<0.005. Significant differences were observed for: PT when comparing I vs. II, I vs. III, I vs. V, II vs. III, II vs. IV, II vs. V, III vs. IV, III vs. V and IV vs. V; aPTT when comparing I vs. II, I vs. III, I vs. IV, II vs. IV, III vs. IV and IV vs. V. No differences were observed among brands for FIB determination. We suggest that every laboratory management should both standardize the procedures and frequently evaluate the quality of in-vitro diagnostic devices. Blood Coagul Fibrinolysis 24:252-255 (C) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.