Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing

被引:30
|
作者
Lima-Oliveira, Gabriel [1 ,2 ,3 ,4 ]
Lippi, Giuseppe [5 ]
Salvagno, Gian Luca [1 ]
Montagnana, Martina [1 ]
Picheth, Geraldo [2 ]
Guidi, Gian Cesare [1 ,2 ]
机构
[1] Univ Verona, Lab Clin Biochem, Dept Life & Reprod Sci, I-37100 Verona, Italy
[2] Univ Fed Parana, Dept Med Pathol, Postgrad Program Pharmaceut Sci, BR-80060000 Curitiba, Parana, Brazil
[3] MERCOSUL Sect Comm Clin Anal & Vitro Diagnost CSM, Rio De Janeiro, Brazil
[4] Brazilian Soc Clin Anal Sao Paulo State, Sao Paulo, Brazil
[5] Acad Hosp Parma, Dept Pathol & Lab Med, Clin Chem & Hematol Lab, Parma, Italy
关键词
blood collection; plasma samples; preanalytical variability; vacuum tubes; validation process; BIOLOGICAL VARIATION; CLINICAL-CHEMISTRY; QUALITY; COLLECTION; STASIS;
D O I
10.1097/MBC.0b013e32835b72ea
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Sometimes in-vitro diagnostic devices (e. g. blood collection tubes) are not validated before use or when the producer's brand is changed. The aim of this study was to validate five brands of sodium citrate vacuum tubes. Blood specimens from 50 volunteers were collected in five different tube brands (I: Venosafe, II: VACUETTE, III: BD Vacutainer, IV: LABOR IMPORT and V: S-Monovette). Routine coagulation tests [activated partial thromboplastin time (aPTT), prothrombin time (PT), and fibrinogen (FIB)] were performed on ACL TOP instrument using HemosIL reagents. The significance of the differences between samples was assessed by paired Student's t-test, set at P<0.005. Significant differences were observed for: PT when comparing I vs. II, I vs. III, I vs. V, II vs. III, II vs. IV, II vs. V, III vs. IV, III vs. V and IV vs. V; aPTT when comparing I vs. II, I vs. III, I vs. IV, II vs. IV, III vs. IV and IV vs. V. No differences were observed among brands for FIB determination. We suggest that every laboratory management should both standardize the procedures and frequently evaluate the quality of in-vitro diagnostic devices. Blood Coagul Fibrinolysis 24:252-255 (C) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:252 / 255
页数:4
相关论文
共 27 条
  • [1] Comparison of Low Voleme and Conventional Sodium Citrate Tubes for Routine Coagulation Testing
    Chang, Soon Hee
    CLINICAL LABORATORY, 2021, 67 (07) : 1555 - 1560
  • [2] Preanalytical management: serum vacuum tubes validation for routine clinical chemistry
    Lima-Oliveira, Gabriel
    Lippi, Giuseppe
    Salvagno, Gian Luca
    Montagnana, Martina
    Picheth, Geraldo
    Guidi, Gian Cesare
    BIOCHEMIA MEDICA, 2012, 22 (02) : 180 - 186
  • [3] Guidance on the critical shortage of sodium citrate coagulation tubes for hemostasis testing
    Gosselin, Robert C.
    Bowyer, Annette
    Favaloro, Emmanuel J.
    Johnsen, Jill M.
    Lippi, Giuseppe
    Marlar, Richard A.
    Neeves, Keith
    Rollins-Raval, Marian A.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2021, 19 (11) : 2857 - 2861
  • [4] Routine coagulation testing in Vacutainer® Citrate Plus tubes filled at minimum or optimal volume
    Balboni, Fiamma
    Barbui, Stefania
    Gallo, Marco
    Berardi, Margherita
    Vezzosi, Marco
    Lippi, Giuseppe
    DIAGNOSIS, 2020, 7 (01) : 55 - 60
  • [5] Effect of 3.2% vs 3.8% sodium citrate concentration on routine coagulation testing
    Adcock, DM
    Kressin, DC
    Marlar, RA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1997, 107 (01) : 105 - 110
  • [6] Minimum citrate tube fill volume for routine coagulation testing
    Pretorius, L.
    van Rensburg, W. J. Janse
    Conradie, C.
    Coetzee, M. J.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (04) : 493 - 495
  • [7] Comparison of samples obtained from 3.2% sodium citrate glass and two 3.2% sodium citrate plastic blood collection tubes used in coagulation testing
    Gosselin, RC
    Janatpour, K
    Larkin, EC
    Lee, YP
    Owings, JT
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2004, 122 (06) : 843 - 848
  • [8] Implementation of lipidomics in clinical routine: Can fluoride/citrate blood sampling tubes improve preanalytical stability?
    Hahnefeld, Lisa
    Gurke, Robert
    Thomas, Dominique
    Schreiber, Yannick
    Schaefer, Stephan M. G.
    Trautmann, Sandra
    Snodgrass, Isabel Faria
    Kratz, Daniel
    Geisslinger, Gerd
    Ferreiros, Nerea
    TALANTA, 2020, 209
  • [9] Filling accuracy and imprecision of commercial evacuated sodium citrate coagulation tubes
    Lippi, Giuseppe
    Salvagno, Gian Luca
    Biljak, Vanja Radisic
    Kralj, Ana-Katarina
    Kuktic, Ivona
    Gelati, Matteo
    Simundic, Ana-Maria
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2019, 79 (04): : 276 - 279
  • [10] Sodium citrate blood contamination by K2-ethylenediaminetetraacetic acid (EDTA): impact on routine coagulation testing
    Lima-Oliveira, G.
    Salvagno, G. L.
    Danese, E.
    Favaloro, E. J.
    Guidi, G. C.
    Lippi, G.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (03) : 403 - 409