Comparison of hydrophone calibration by reciprocity and heterodyne interferometer in the frequency range 500 kHz to 15 MHz

被引:2
|
作者
Xing, Guangzhen [1 ]
Yang, Ping [1 ]
He, Longbiao [1 ]
机构
[1] Natl Inst Metrol, Beijing 100013, Peoples R China
关键词
Hydrophone calibration; Heterodyne interferometer; Two-transducer reciprocity; Arctangent demodulation; OPTICAL INTERFEROMETRY; FIELD;
D O I
10.4028/www.scientific.net/AMM.239-240.10
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents a comparison of high-frequency hydrophone calibration by a heterodyne interferometer and two-transducer reciprocity method respectively. For the calibration using the heterodyne interferometer, a vertical arrangement is applied to avoid the acousto-optical interaction which is a major uncertainty source in previous studies. For the demodulation of the acoustic displacement, a PC-based digital demodulation using arctangent algorithm is applied which has overcome the drawback of analog demodulation and achieved a higher demodulation accuracy. The demodulation is independent of the intensity of the carrier signal and thus can avoid the correction for the frequency response of the photodiode which is another uncertainty source. Comparison results between these two methods showed a good agreement. Owing to the optical measurement is independent of the acoustic field generating by the transducer, it has wider usage compared to the reciprocity method and can be easily extended to higher frequency and higher acoustic power application.
引用
收藏
页码:10 / 15
页数:6
相关论文
共 50 条
  • [41] Attenuation Measurement Using the IF Substitution Method in the Frequency Range of 9 kHz to 10 MHz
    Lee, Joo-Gwang
    Kang, Tae-Weon
    Hong, Young-Pyo
    Kwon, Jae-Yong
    [J]. JOURNAL OF ELECTROMAGNETIC ENGINEERING AND SCIENCE, 2024, 24 (04): : 370 - 376
  • [42] Calculable AC Voltage Standards for 10 kHz-1 MHz Frequency Range
    Grzenik, Michal
    Kampik, Marian
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1372 - 1378
  • [43] Measurement Uncertainty in Vibration Calibration in Frequency Range of 5 Hz to 10 kHz
    Garg, Naveen
    Chauhan, B. S.
    [J]. MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2020, 35 (03): : 397 - 405
  • [44] Measurement Uncertainty in Vibration Calibration in Frequency Range of 5 Hz to 10 kHz
    Naveen Garg
    B. S. Chauhan
    [J]. MAPAN, 2020, 35 : 397 - 405
  • [45] Primary reciprocity-based method for calibration of hydrophone magnitude and phase sensitivity: Complete tests at frequencies from 1 to 7 MHz
    Oliveira, E. G.
    Costa-Felix, R. P. B.
    Machado, J. C.
    [J]. ULTRASONICS, 2015, 58 : 87 - 95
  • [46] Capacitive micromachined ultrasonic transducer (cMUT) array for the frequency range below 500 kHz
    Buhrdorf, A
    Tebje, L
    Ahrens, O
    Glitza, O
    Binder, J
    [J]. 2000 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2000, : 915 - 918
  • [47] Dynamic range extension for heterodyne fibre-optic hydrophone by using instantaneous frequency-tracking method
    Yang, Sufan
    Zhang, Chunxi
    Wang, Xiaxiao
    [J]. ELECTRONICS LETTERS, 2023, 59 (16)
  • [48] The extrapolation factor for PLC radiation measurements in the 150-500 kHz frequency range
    Waechter, Markus
    Koch, Michael
    Schwing, Christoph
    Hirsch, Holger
    [J]. 2013 17TH IEEE INTERNATIONAL SYMPOSIUM ON POWER LINE COMMUNICATIONS AND ITS APPLICATIONS (ISPLC), 2013, : 220 - 224
  • [49] Electromagnetic Compatibility in Europe for OFDM based PLC access data transmission in the frequency range 150 kHz to 500 kHz
    Ponzelar, Sebastian
    Mengi, Anil
    Koch, Michael
    [J]. 2015 INTERNATIONAL SYMPOSIUM ON POWER LINE COMMUNICATIONS AND ITS APPLICATIONS (ISPLC), 2015, : 195 - 198
  • [50] Assessment of the permeability and pore size of quartz-based consolidated sedimentary rocks by dielectric spectroscopy in the frequency range of 1 kHz-500 MHz
    Bobrov, P. P.
    Golikov, N. A.
    Kroshka, E. S.
    Repin, A. V.
    [J]. JOURNAL OF APPLIED GEOPHYSICS, 2022, 204