Maximizing kinetic performance in supercritical fluid chromatography using state-of-the-art instruments

被引:84
|
作者
Perrenoud, Alexandre Grand-Guillaume [1 ]
Hamman, Chris [2 ]
Goel, Meenakshi [3 ]
Veuthey, Jean-Luc [1 ]
Guillarme, Davy [1 ]
Fekete, Szabolcs [1 ]
机构
[1] Univ Lausanne, Univ Geneva, Sch Pharmaceut Sci, CH-1211 Geneva 4, Switzerland
[2] Genentech Inc, Discovery Chem, San Francisco, CA 94080 USA
[3] Genentech Inc, Small Mol Analyt Chem Qual Control, San Francisco, CA 94080 USA
关键词
Extra-column band broadening; Kinetic performance; Kinetic plot methodology; Waters Acquity UPC2; Agilent 1260 SFC system; High resolution; BAND-BROADENING CONTRIBUTIONS; ULTRA-HIGH PRESSURE; LIQUID-CHROMATOGRAPHY; MOBILE-PHASE; CARBON-DIOXIDE; COLUMN EFFICIENCY; STATIONARY PHASES; POROUS PARTICLES; SOLUTE RETENTION; SEPARATIONS;
D O I
10.1016/j.chroma.2013.09.039
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Recently, there has been a renewed interest in supercritical fluid chromatography (SFC), due to the introduction of state-of-the-art instruments and dedicated columns packed with small particles. However, the achievable kinetic performance and practical possibilities of such modern SFC instruments and columns has not been described in details until now. The goal of the present contribution was to provide some information about the optimal column dimensions (i.e. length, diameter and particle size) suitable for such state-of the-art systems, with respect to extra-column band broadening and system upper pressure limit. In addition, the reliability of the kinetic plot methodology, successfully applied in RPLC, was also evaluated under SFC conditions. Taking into account the system variance, measured at similar to 85 mu L-2, on modern SFC instruments, a column of 3 mm I.D. was ideally suited for the current technology, as the loss in efficiency remained reasonable (i.e. less than 10% decrease for k > 6). Conversely, these systems struggle with 2.1 mm I.D. columns (55% loss in N for k = 5). Regarding particle size, columns packed with 5 mu m particles provided unexpectedly high minimum reduced plate height values (h(min) =3.0-3.4), while the 3.5 and 1.7 mu m packing provided lower reduced plate heights h(min) = 2.2-2.4 and h(min) = 2.7-3.2, respectively. Considering the system upper pressure limit, it appears that columns packed with 1.7 mu m particles give the lowest analysis time for efficiencies up to 40,000-60,000 plates, if the mobile phase composition is in the range of 2-19% MeOH. The 3.5 mu m particles were attractive for higher efficiencies, particularly when the modifier percentage was above 20%, while 5 mu m was never kinetically relevant with modern SFC instruments, due to an obvious limitation in terms of upper flow rate value. The present work also confirms that the kinetic plot methodology could be successfully applied to SFC, without the need for isopycnic measurements, as the difference in plate count between predicted and experimental values obtained by coupling several columns in series (up to 400 mm) was on average equal to 3-6% and with a maximum of 13%. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:288 / 297
页数:10
相关论文
共 50 条
  • [21] Magnetorheological fluid in prostheses: A state-of-the-art review
    Dutra, Rina Mariane Alves
    de Andrade, Rafhael Milanezi
    Soares, Alcimar Barbosa
    Thakor, Nitish Vyomesh
    Vimieiro, Claysson Bruno Santos
    [J]. JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 2024, 35 (05) : 485 - 516
  • [22] ANALYSIS OF SULFONAMIDES USING SUPERCRITICAL FLUID CHROMATOGRAPHY AND SUPERCRITICAL FLUID CHROMATOGRAPHY MASS-SPECTROMETRY
    PERKINS, JR
    GAMES, DE
    STARTIN, JR
    GILBERT, J
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 540 (1-2): : 239 - 256
  • [23] Nanoparticles in drilling fluid: A review of the state-of-the-art
    Cheraghian, Goshtasp
    [J]. JOURNAL OF MATERIALS RESEARCH AND TECHNOLOGY-JMR&T, 2021, 13 : 737 - 753
  • [24] Nanoparticles in drilling fluid: A review of the state-of-the-art
    Cheraghian, Goshtasp
    [J]. Journal of Materials Research and Technology, 2021, 13 : 737 - 753
  • [25] State-of-the-art fluid management in the operating room
    Miller, Timothy E.
    Raghunathan, Karthik
    Gan, Tong J.
    [J]. BEST PRACTICE & RESEARCH-CLINICAL ANAESTHESIOLOGY, 2014, 28 (03) : 261 - 273
  • [26] FUNDAMENTALS OF COLUMN PERFORMANCE IN SUPERCRITICAL FLUID CHROMATOGRAPHY
    NOVOTNY, M
    SPRINGSTON, SR
    [J]. JOURNAL OF CHROMATOGRAPHY, 1983, 279 (NOV): : 417 - 422
  • [27] PERFORMANCE OF CAPILLARY RESTRICTORS IN SUPERCRITICAL FLUID CHROMATOGRAPHY
    SMITH, RD
    FULTON, JL
    PETERSEN, RC
    KOPRIVA, AJ
    WRIGHT, BW
    [J]. ANALYTICAL CHEMISTRY, 1986, 58 (09) : 2057 - 2064
  • [28] INTEGRATED INTELLIGENT INSTRUMENTS - SUPERCRITICAL FLUID EXTRACTION, DESORPTION, REACTION, AND CHROMATOGRAPHY
    LIEBMAN, SA
    LEVY, EJ
    LURCOTT, S
    ONEILL, S
    GUTHRIE, J
    RYAN, T
    YOCKLOVICH, S
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1989, 27 (03) : 118 - 126
  • [29] Kinetic performance of a 50 mm long 1.8 μm chiral column in supercritical fluid chromatography
    Berger, Terry A.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2016, 1459 : 136 - 144
  • [30] Design and evaluation of various methods for the construction of kinetic performance limit plots for supercritical fluid chromatography
    Delahaye, Sander
    Broeckhoven, Ken
    Desmet, Gert
    Lynen, Frederic
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2012, 1258 : 152 - 160