Graphite furnace atomic absorption spectrometric determination of some noble metals using flow injection on-line clean-up

被引:18
|
作者
Su, XG
Wang, MJ
Zhang, YH
Zhang, JH
Zhang, HQ
Jin, QH [1 ]
机构
[1] Jilin Univ, Dept Chem, Changchun 130023, Peoples R China
[2] Chinese Acad Sci, Lab Excited State Proc, Changchun 130021, Peoples R China
关键词
D O I
10.1039/b105306j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, a flow injection on-line clean-up system was developed for determination of trace Pt, Pd and Au in the presence of Na+, K+, Ca2+, Mg2+, Fe3+ and Al3+ by electrothermal atomic absorption spectrometry (ETAAS). A column packed with strongly acidic cation-exchange resin (Type 732) was used as a clean-up column. As Pt2+, Pd2+ and Au3+ can form strong complex anions with Cl- in HCl medium, they can be separated from coexisting cations by passing through the clean-up column. The experimental results obtained for trace amounts of gold in reference materials were in good agreement with the certified values.