The repeatability of corneal and corneal epithelial thickness measurements using optical coherence tomography

被引:86
|
作者
Sin, Sokpheaktra [1 ]
Simpson, Trefford L. [1 ]
机构
[1] Univ Waterloo, Sch Optometry, Ctr Contact Lens Res, Waterloo, ON N2L 3G1, Canada
关键词
optical coherence tomographer; OCT; pachymetry; pachometry; corneal thickness; corneal epithelial thickness; corneal epithelial; repeatability;
D O I
10.1097/01.opx.0000221388.26031.23
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. The purpose of this study is to examine within and between session repeatability of clinical optical coherence tomography (OCT) imaging for anterior segment morphometry. Methods. Images of the corneal apex of each eye in 18 subjects were obtained using a Humphrey Zeiss OCT imager. Subjects viewed a target positioned to ensure that scans were orthogonal to the ocular surface and each image, consisting of 100 adjacent sagittal scans, analyzed using custom software. Repeatability data were analyzed using intraclass correlation coefficients (ICCs), correlation coefficient of concordance (CCC, perfect test-retest agreement ICC or CCC = 1.0), and coefficients of repeatability (COR, 95% confidence interval of test-retest differences). To account for each eye, the multivariate repeatability statistic Iota was estimated. Results. Mean central corneal and epithelial thickness of 32 eyes (OD and OS combined) is 536 +/- 26 mu m (standard deviation [SD]) and 52 +/- 3 mu m (SD) with 5th and 95th percentile thicknesses of 507 and 591 mu m for central cornea and 48 and 57 mu m for central epithelial. Worst case within session repeatability was defined as repeatability between images with greatest differences in mean thickness within a session. Corneal thickness worst case ICC was 0.95 and COR was +/- 9.98 mu m. Epithelium worst case ICC was 0.36, CCC was 0.12, and COR was +/- 11.11 mu m. First image between session corneal thickness had an ICC = 0.98 and a COR = 10.83 mu m, whereas epithelium ICC = 0.38, CCC = 0.37, and COR was +/- 12.84 mu m. When we compared the average of the first three tests with the first three retest images, corneal ICC was 0.98 and COR was 10.64 mu m and epithelium ICC = 0.73, CCC = 0.72, and COR was +/- 6.53 mu m. Iota (multivariate repeatability, using eye as a factor) for the cornea was at least 0.96 (worst case) and increased to at least 0.98 when within-session image data were averaged. Iota for epithelium measures ranged from 0.29 when first images were compared with 0.57 when within-session image data were averaged. Conclusions. There is very good repeatability of corneal thickness measurement using OCT; even the worst case measurements are similar between sessions. On the other hand, this is not the case for epithelium measurements, and if multiple images within a session are acquired, the worst case results demonstrate how important it is to optimize each OCT scan and also average multiple scans to maximize intersession repeatability.
引用
收藏
页码:360 / 365
页数:6
相关论文
共 50 条
  • [11] Assessment of central corneal thickness using optical coherence tomography
    Fishman, GR
    Pons, ME
    Seedor, JA
    Liebmann, JM
    Ritch, R
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04): : 707 - 711
  • [12] Corneal layer thickness in keratoconus using optical coherence tomography
    Heidari, Zahra
    Mohammadpour, Mehrdad
    Hajizadeh, Fedra
    Fotouhi, Akbar
    Hashemi, Hassan
    [J]. CLINICAL AND EXPERIMENTAL OPTOMETRY, 2024, 107 (01) : 32 - 39
  • [13] Wide Corneal Epithelial Mapping Using an Optical Coherence Tomography
    Hashmani, Nauman
    Hashmani, Sharif
    Saad, Choudhry M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (03) : 1652 - 1658
  • [14] Age-related differences in corneal epithelial thickness measurements with anterior segment optical coherence tomography
    Bong Jun Kim
    Ik-Hee Ryu
    Sun Woong Kim
    [J]. Japanese Journal of Ophthalmology, 2016, 60 : 357 - 364
  • [15] Age-related differences in corneal epithelial thickness measurements with anterior segment optical coherence tomography
    Kim, Bong Jun
    Ryu, Ik-Hee
    Kim, Sun Woong
    [J]. JAPANESE JOURNAL OF OPHTHALMOLOGY, 2016, 60 (05) : 357 - 364
  • [16] Corneal epithelial thickness measured by Fourier-domain optical coherence tomography
    Wang, Ting
    Li, Fengjie
    Wang, Shuting
    Shi, Weiyun
    [J]. SPEKTRUM DER AUGENHEILKUNDE, 2016, 30 (01) : 3 - 9
  • [17] Measuring Corneal Epithelial Thickness Change after LASIK with Optical Coherence Tomography
    Tang, Maolong
    Li, Yan
    Huang, David
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [18] Corneal epithelial thickness analysis of forme fruste keratoconus with optical coherence tomography
    Yang, Xiao-Long
    Wang, Yun
    Luo, Bao-Gen
    Xu, Yue
    Zhang, Xiao-Feng
    [J]. INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2021, 14 (01) : 89 - 96
  • [19] Corneal epithelial thickness analysis of forme fruste keratoconus with optical coherence tomography
    Xiao-Long Yang
    Yun Wang
    Bao-Gen Luo
    Yue Xu
    Xiao-Feng Zhang
    [J]. International Journal of Ophthalmology, 2021, 14 (01) : 89 - 96
  • [20] In vitro measurement of rabbit corneal epithelial thickness using ultrahigh resolution optical coherence tomography
    Reiser, BJ
    Ignacio, TS
    Wang, YM
    Taban, M
    Graff, JM
    Sweet, P
    Chen, ZP
    Chuck, RS
    [J]. VETERINARY OPHTHALMOLOGY, 2005, 8 (02) : 85 - 88