Dosimetric effect of body contour changes for prostate and head and neck volumetric modulated arc therapy plans

被引:7
|
作者
Sun, Lingyue [1 ,2 ]
Kirkby, Charles [1 ,3 ,4 ]
Smith, Wendy [1 ,2 ,3 ]
机构
[1] Univ Calgary, Dept Phys & Astron, Calgary, AB T2N 1N4, Canada
[2] Torn Baker Canc Ctr, Dept Med Phys, Calgary, AB T2N 4N2, Canada
[3] Univ Calgary, Dept Oncol, Calgary, AB T2N 1N4, Canada
[4] Jack Ady Canc Ctr, Dept Med Phys, Lethbridge, AB T1J 1W5, Canada
来源
关键词
body contour change; head and neck; prostate; VMAT plans; RADIATION-THERAPY; WEIGHT-LOSS; ADAPTIVE RADIOTHERAPY; IMRT; RISK; PROBABILITY; PATIENT; CANCER; ORGANS; SIZE;
D O I
10.1002/acm2.12571
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Body contour changes are commonly seen in prostate and head and neck (H&N) patients undergoing volumetric modulated arc therapy (VMAT) treatments, which may cause a discrepancy between the planned dose and the delivered dose. Dosimetrists, radiation oncologists or medical physicists sometimes are required to visually assess the dosimetric impact of body contour changes and make a judgment call on whether further re-assessment of the plan is needed. However, an intuitive judgment cannot always be made in a timely manner due to the complexity of VMAT plans as well as the complicated forms of body contour changes. This study evaluated the dosimetric effect of body contour changes for prostate and H&N patients to help with clinical decision-making. By analyzing the one-dimensional spatial dose profiles from the original body and the body with different body contour deformations, rules of thumb for dose percentage change and isodose line shift due to body contour changes were ascertained. Moreover, based on dose distribution comparison using three-dimensional gamma analysis, the response of the clinical prostate and H&N VMAT plans to body contour changes was assessed. Within center specific dose deviation tolerances, prostate patients who had less than 2cm single side body contour change or less than 1cm uniform body contour change were unlikely to need plan re-assessment; H&N VMAT plans with less than 1cm uniform body contour change or less than 1cm shoulder superior-inferior positional change were also unlikely to trigger further evaluation. Dose percentage change and isodose line shift were considered independently from the problem of volume changes in this study, but clinically, both aspects must be considered.
引用
收藏
页码:115 / 124
页数:10
相关论文
共 50 条
  • [1] Rules of Thumb for the Dosimetric Effect of Head and Neck and Prostate Patients' Body Contour Changes
    Sun, Lingyue
    Kirkby, Charles
    Smith, Wendy
    [J]. MEDICAL PHYSICS, 2019, 46 (11) : 5384 - 5384
  • [2] Dose distribution comparison in volumetric- modulated arc therapy plans for head and neck cancers with and without an external body contour extended technique
    Tanaka, Yoshihiro
    Monzen, Hajime
    Matsumoto, Kenji
    Inomata, Shinichiro
    Fuse, Toshiaki
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (06) : 576 - 584
  • [3] The dosimetric impact of dental implants on head-and-neck volumetric modulated arc therapy
    Lin, Mu-Han
    Li, Jinsheng
    Price, Robert A., Jr.
    Wang, Lu
    Lee, Chung-Chi
    Ma, C-M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (04): : 1027 - 1040
  • [4] Is there a dosimetric advantage of volumetric modulated arc therapy over intensity modulated radiotherapy in head and neck cancer?
    Buciuman, Nikolett
    Marcu, Loredana G.
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2022, 279 (11) : 5311 - 5321
  • [5] Is there a dosimetric advantage of volumetric modulated arc therapy over intensity modulated radiotherapy in head and neck cancer?
    Nikolett Buciuman
    Loredana G. Marcu
    [J]. European Archives of Oto-Rhino-Laryngology, 2022, 279 : 5311 - 5321
  • [6] Dosimetric comparison of RapidPlan and manually optimised volumetric modulated arc therapy plans in prostate cancer
    Chua, Loyce M. H.
    Pang, Eric P. P.
    Master, Zubin
    Sultana, Rehena
    Tuan, Jeffrey K. L.
    Bragg, Christopher M.
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 257 - 264
  • [7] Dosimetric comparison of RapidPlan and manually optimized plans in volumetric modulated arc therapy for prostate cancer
    Kubo, Kazuki
    Monzen, Hajime
    Ishii, Kentaro
    Tamura, Mikoto
    Kawamorita, Ryu
    Sumida, Iori
    Mizuno, Hirokazu
    Nishimura, Yasumasa
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 44 : 199 - 204
  • [8] Increasing the number of arcs improves head and neck volumetric modulated arc therapy plans
    Tol, Jim P.
    Dahele, Max
    Slotman, Ben J.
    Verbakel, Wilko F. A. R.
    [J]. ACTA ONCOLOGICA, 2015, 54 (02) : 283 - U157
  • [9] Dosimetric Effect On Variation of Patient Size in Prostate Volumetric Modulated Arc Therapy
    Chow, J.
    Jiang, R.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3825 - 3825
  • [10] Verification of Mechanical and Dosimetric Accuracy for Head and Neck Volumetric Modulated Arc Therapy With Jaw Tracking
    Okumura, T.
    Nakashima, T.
    Aita, M.
    Ochi, Y.
    Kawahara, D.
    Masuda, H.
    Hioki, K.
    Ohno, Y.
    Saito, A.
    Nagata, Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E638 - E638