Bench comparative evaluation of a new generation and standard helmet for delivering non-invasive ventilation

被引:37
|
作者
Olivieri, C. [1 ]
Costa, R. [2 ]
Spinazzola, G. [2 ]
Ferrone, G. [2 ]
Longhini, F. [3 ]
Cammarota, G. [1 ]
Conti, G. [2 ]
Navalesi, P. [3 ,4 ,5 ]
机构
[1] Maggiore Carita Univ Hosp, Novara, Italy
[2] Univ Cattolica Sacro Cuore, Dept Anesthesiol & Intens Care Med, Rome, Italy
[3] Eastern Piedmont Univ, Dept Translat Med, I-28100 Novara, Italy
[4] ASL VC, St Andrea Hosp, Vercelli, Italy
[5] CRRF Mons L Novarese, Moncrivello, VC, Italy
关键词
Mechanical ventilation; Non-invasive ventilation; Continuous positive airway pressure; Pressure support ventilation; Helmet; Trigger; OBSTRUCTIVE PULMONARY-DISEASE; POSITIVE-PRESSURE VENTILATION; RESPIRATORY-FAILURE; FACE MASK; MECHANICAL VENTILATION; DIFFERENT INTERFACES; ACUTE EXACERBATION; AIRWAY PRESSURE; COMFORT;
D O I
10.1007/s00134-012-2765-z
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
To evaluate the performance of a new helmet (NH) recently introduced into clinical use relative to that of the standard helmet (SH) in terms of delivering non-invasive continuous positive airway pressure (nCPAP) and pressure support ventilation (nPSV). This was a bench study using a mannequin connected to an active lung simulator. The SH was fastened to the mannequin by armpit braces, which are not needed to secure the NH. The inspiratory and expiratory variations in nCPAP delivered with two different simulated efforts (Pmus), were determined relative to the preset CPAP level. nPSV was applied at two simulated respiratory rates (RR) and two cycling-off flow thresholds. We measured inspiratory trigger delay (Delay(trinsp)), expiratory trigger delay (Delay(trexp)), time of synchrony (Time(sync)), trigger pressure drop (Delta P (trigger)), airway pressure-time product during the triggering phase (PTPt), the initial 200 ms from the onset of the ventilator pressurization (PTP200), and the initial 300 and 500 ms from the onset of the simulated effort; this two latter parameters were expressed as the percentage of the area of ideal pressurization (PTP300-index and PTP500-index, respectively). In nCPAP, at both Pmus, the differences between the two interfaces at both Pmus were small and clinically irrelevant. In nPSV, regardless of the setting, NH resulted in significantly smaller trigger delays, Delta P (trigger), and PTPt. Time(sync), PTP200, PTP300-index, and PTP500-index were also significantly higher with the NH compared to the SH, irrespective of the setting. Compared to the SH, the NH is equally effective in delivering nCPAP and more effective in delivering nPSV, and it is used to avoid the need for armpit braces.
引用
收藏
页码:734 / 738
页数:5
相关论文
共 50 条
  • [31] Non-invasive ventilation
    Heindl, F
    WIENER KLINISCHE WOCHENSCHRIFT, 1999, 111 (19) : 784 - 801
  • [32] Non-invasive ventilation
    Koehnlein, T.
    Welte, T.
    INTERNIST, 2006, 47 (09): : 917 - +
  • [33] Non-invasive ventilation
    Nava, S.
    Navalesi, P.
    Carlucci, A.
    MINERVA ANESTESIOLOGICA, 2009, 75 (1-2) : 31 - 36
  • [34] PHYSIOLOGIC COMPARISON BETWEEN A NEW HELMET AND THE CONVENTIONAL HELMET IN DELIVERING NONINVASIVE VENTILATION
    Longhini, F.
    Olivieri, C.
    Repetto, V.
    Candriella, M.
    Pagni, A. Miziara
    Cammarota, G.
    Della Corte, F.
    Navalesi, P.
    INTENSIVE CARE MEDICINE, 2013, 39 : S407 - S407
  • [35] Improvement of the trigger of a ventilator for non-invasive ventilation in children: bench and clinical study
    Khirani, Sonia
    Louis, Bruno
    Leroux, Karl
    Ramirez, Adriana
    Lofaso, Frederic
    Fauroux, Brigitte
    CLINICAL RESPIRATORY JOURNAL, 2016, 10 (05): : 559 - 566
  • [36] Evaluation of a Volume Targeted NIV Device: Bench Evaluation of the Breathe Technologies Non-Invasive Open Ventilation System (NIOV™)
    Blakeman, Thomas
    Branson, Richard
    COPD-JOURNAL OF CHRONIC OBSTRUCTIVE PULMONARY DISEASE, 2014, 11 (05) : 568 - 574
  • [37] Successful and prolonged application of non-invasive pressure support ventilation by helmet in patients with ARDS
    Antonelli, M
    Conti, G
    Pennisi, MA
    Maggiore, SM
    Gregoretti, C
    Pelosi, P
    Arcangeli, A
    Auricchio, D
    Petrucci, S
    Pratesi, P
    Santilli, F
    Mercurio, G
    INTENSIVE CARE MEDICINE, 2002, 28 : S24 - S24
  • [38] NON-INVASIVE VENTILATION VIA HELMET IN WEANING PATIENTS WITH HYPOXEMIC ACUTE RESPIRATORY FAILURE
    Vaschetto, R.
    Turucz, E.
    Dellapiazza, F.
    Cammarota, G.
    Colombo, D.
    Barra, F.
    Damnotti, F.
    Della Corte, F.
    Antonelli, M.
    Navalesi, P.
    INTENSIVE CARE MEDICINE, 2010, 36 : S347 - S347
  • [39] Effects of non-invasive ventilation with helmet or facial mask on middle ear function of volunteers
    Cavaliere, F
    Suppa, E
    Polidori, L
    Ganz, B
    Toto, R
    Masieri, S
    INTENSIVE CARE MEDICINE, 2002, 28 : S71 - S71
  • [40] Non-invasive pressure support ventilation (NPSV) delivered by helmet for acute exacerbation of COPD
    Antonelli, M
    Conti, G
    Pennisi, M
    Cesare, G
    Rocco, M
    Pelosi, P
    Cavaliere, F
    Polidori, L
    Iacobone, E
    Marchese, M
    INTENSIVE CARE MEDICINE, 2002, 28 : S68 - S68