Helical tomotherapy quality assurance with ArcCHECK

被引:6
|
作者
Chapman, David [1 ]
Barnett, Rob [1 ]
Yartsev, Slav [1 ]
机构
[1] London Hlth Sci Ctr, London Reg Canc Program, London, ON, Canada
关键词
Helical tomotherapy; Quality assurance; ArcCHECK; QA; UNIT;
D O I
10.1016/j.meddos.2013.12.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
To design a quality assurance (QA) procedure for helical tomotherapy that measures multiple beam parameters with 1 delivery and uses a rotating gantry to simulate treatment conditions. The customized QA procedure was preprogrammed on the tomotherapy operator station. The dosimetry measurements were performed using an ArcCHECK diode array and an AlSL ion chamber inserted in the central holder. The ArcCHECK was positioned 10 cm above the isocenter so that the 21-cm diameter detector array could measure the 40-cm wide tomotherapy beam. During the implementation of the new QA procedure, separate comparative measurements were made using ion chambers in both liquid and solid water, the tomotherapy onboard detector array, and a MapCHECK diode array for a period of 10 weeks. There was good agreement (within 1.3%) for the beam output and cone ratio obtained with the new procedure and the routine QA measurements. The measured beam energy was comparable (0.3%) to solid water measurement during the 10-week evaluation period, excluding 2 of the 10 measurements with unusually high background. The symmetry reading was similarly compromised for those 2 weeks, and on the other weeks, it deviated from the solid water reading by similar to 2.5%. The ArcCHECK phantom presents a suitable alternative for performing helical tomotherapy QA, provided the background is collected properly. The proposed weekly procedure using ArcCHECK and water phantom makes the QA process more efficient. (c) 2014 American Association of Medical Dosimetrists.
引用
收藏
页码:159 / 162
页数:4
相关论文
共 50 条
  • [31] Constancy Check of Longitudinal Beam Profile for Helical Tomotherapy Using ArcCHECK
    Yang, B.
    Wong, W.
    Geng, H.
    Lam, W.
    Cheung, K.
    Yu, S.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2932 - 2932
  • [32] Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass
    Wong, J. H. D.
    Hardcastle, N.
    Tome, W. A.
    Bayliss, A.
    Tolakanahalli, R.
    Lerch, M. L. F.
    Petasecca, M.
    Carolan, M.
    Metcalfe, P.
    Rosenfeld, A. B.
    [J]. MEDICAL PHYSICS, 2011, 38 (04) : 2256 - 2264
  • [33] The effect of the frequency of absolute dose calibration on the performance of ArcCHECK in helical TomoTherapy DQA
    Wong, W. K. R.
    Lo, S. G.
    Cheung, K.
    Yu, S. K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S162 - S163
  • [34] Evaluation of the Use of ArcCHECK™ for RapidArc® Quality Assurance
    Noble, R.
    Edwards, C.
    Handley, J-C
    [J]. CLINICAL ONCOLOGY, 2011, 23 (03) : S47 - S47
  • [35] FILMLESS QUALITY ASSURANCE FOR HELICAL TOMOTHERAPY WITH A CYLINDRICAL STAIR-LIKE ALUMINIUM PHANTOM
    Althof, V.
    Kramer, D.
    Westendorp, R.
    van Haaren, P.
    Wispelweij, M.
    Minken, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S92 - S93
  • [36] Tomotherapy daily quality assurance phantom
    Jaradat, H
    Paliwal, B
    Mehta, M
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1989 - 1989
  • [37] Patient quality assurance analysis for tomotherapy
    Papanikolaou, N
    Shi, CY
    Yan, Y
    Weng, X
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1986 - 1986
  • [38] Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy
    Emilie T. Soisson
    Nick Hardcastle
    Wolfgang A. Tomé
    [J]. Journal of Neuro-Oncology, 2010, 98 : 277 - 285
  • [39] Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy
    Soisson, Emilie T.
    Hardcastle, Nick
    Tome, Wolfgang A.
    [J]. JOURNAL OF NEURO-ONCOLOGY, 2010, 98 (02) : 277 - 285
  • [40] Comparative Evaluation of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy
    Chang, K.
    Lee, J.
    Choi, J.
    Hong, C.
    Jung, D.
    Kim, J.
    Kim, J.
    Han, M.
    Lee, H.
    Kim, D.
    Kim, J.
    Park, K.
    Cho, J.
    Kim, Y.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E611 - E611