A Monte Carlo based analytic model of the in-room neutron ambient dose equivalent for a Mevion gantry-mounted passively scattered proton system

被引:5
|
作者
Baradaran-Ghahfarokhi, Milad [1 ]
Reynoso, Francisco [1 ]
Darafsheh, Arash [1 ]
Sun, Baozhou [1 ]
Prusator, Michael T. [1 ]
Mutic, Sasa [1 ]
Zhao, Tianyu [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
neutron ambient dose equivalent; Monte Carlo simulation; analytic model; passively scattered proton; shielding; RADIATION; RADIOTHERAPY; PHOTON;
D O I
10.1088/1361-6498/abaff4
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The goal of this study was to develop a Monte Carlo (MC)-based analytical model that can predict the in-room ambient dose equivalent from a Mevion gantry-mounted passively scattered proton system. The Mevion S250 and treatment vault were simulated using the MCNPX MC code. The results of the in-room neutron dose measurements, using an FHT 762 WENDI-II detector, were employed to benchmark the MC-derived values. After tuning the MCNPX MC code, for the same beam delivery parameters, the code was used to calculate the neutron spectra and ambient dose equivalent in the vault and at varying angles from the isocenter. Then, based on the calculations, an analytical model was reconstructed and data were fitted to derive the model parameters at 95% confidence intervals (CI). The MCNPX codes were tuned to within about 19% of the measured values for most of the measurements in the vault. For the maze, up to 0.08 mSv Gy(-1)discrepancies were found between the experimental measurements and MCNPX calculated results. The analytical model showed up to 18% discrepancy for distances between 100 and 600 cm from the isocenter compared to the MC calculations. The model may underestimate the neutron ambient dose equivalent up to 21% for distances less than 100 cm from the isocenter. The proposed analytical model can be used to estimate the contribution of the secondary neutron dose from the Mevion S250 for the design of local shielding inside the proton therapy treatment vault.
引用
收藏
页码:980 / 996
页数:17
相关论文
共 13 条
  • [1] A Monte Carlo-Based Analytic Model of Neutron Dose Equivalent for a Mevion Gantry-Mounted Passively Scattered Proton System
    BaradaranGhahfarokhi, M.
    Reynoso, F.
    Sun, B.
    Darafsheh, A.
    Prusator, M.
    Mutic, S.
    Zhao, T.
    MEDICAL PHYSICS, 2020, 47 (06) : E682 - E682
  • [2] A Monte Carlo-based analytic model of neutron dose equivalent for a mevion gantry-mounted passively scattered proton system for craniospinal irradiation
    Baradaran-Ghahfarokhi, Milad
    Reynoso, Francisco
    Sun, Baozhou
    Darafsheh, Arash
    Prusator, Michael T.
    Mutic, Sasa
    Zhao, Tianyu
    MEDICAL PHYSICS, 2020, 47 (09) : 4509 - 4521
  • [3] Sensitivity analysis of Monte Carlo model of a gantry-mounted passively scattered proton system
    Baradaran-Ghahfarokhi, Milad
    Reynoso, Francisco
    Prusator, Michael T.
    Sun, Baozhou
    Zhao, Tianyu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (02): : 26 - 37
  • [4] Monte Carlo Simulation and Sensitivity Analysis of a Gantry-Mounted Passively Scattered Proton System
    Baradaran, M.
    Reynoso, F.
    Prusator, M.
    Zhao, T.
    MEDICAL PHYSICS, 2019, 46 (06) : E600 - E600
  • [5] Monte Carlo evaluation of organ doses from a proton gantry-mounted CBCT system
    Henry, T.
    Ardenfors, O.
    Gudowska, I.
    Poludniowski, G.
    Dasu, A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S904 - S904
  • [6] Monte Carlo simulations of neutron spectral fluence, radiation weighting factor and ambient dose equivalent for a passively scattered proton therapy unit
    Zheng, Yuanshui
    Fontenot, Jonas
    Taddei, Phil
    Mirkovic, Dragan
    Newhauser, Wayne
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (01): : 187 - 201
  • [7] An analytic model of neutron ambient dose equivalent and equivalent dose for proton radiotherapy
    Zhang, Rui
    Perez-Andujar, Angelica
    Fontenot, Jonas D.
    Taddei, Phillip J.
    Newhauser, Wayne D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23): : 6975 - 6985
  • [8] Measured Neutron Spectra and Dose Equivalents From a Mevion Single-Room, Passively Scattered Proton System Used for Craniospinal Irradiation
    Howell, Rebecca M.
    Burgett, Eric A.
    Isaacs, Daniel
    Hedrick, Samantha G. Price
    Reilly, Michael P.
    Rankine, Leith J.
    Grantham, Kevin K.
    Perkins, Stephanie
    Klein, Eric E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01): : 249 - 257
  • [9] Monte Carlo study of neutron-ambient dose equivalent to patient in treatment room
    Mohammadi, A.
    Afarideh, H.
    Davani, F. Abbasi
    Ghergherehchi, M.
    Arbabi, A.
    APPLIED RADIATION AND ISOTOPES, 2016, 118 : 140 - 148
  • [10] Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design
    Titt, Uwe
    Pera, Enzo
    Gillin, Michael T.
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2020, 6 (04) : 29 - 37