Noninvasive Cuffless Estimation of Blood Pressure using Photoplethysmography without Electrocardiograph Measurement

被引:0
|
作者
Samria, Rohan [1 ]
Jain, Ridhi [1 ]
Jha, Ankita [1 ]
Saini, Sandeep [1 ]
Chowdhury, Shubhajit Roy [2 ]
机构
[1] LNM Inst Informat Technol Jaipur, Dept Elect & Commun Engn, Jaipur, Rajasthan, India
[2] IIIT Hyderabad, Ctr VLSI Design & Embedded Syst, Hyderabad, Telangana, India
关键词
blood pressure; photoplethysmography; systolic blood pressure; diastolic blood pressure; systolic upstroke time; diastolic time;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The paper presents a novel approach of measuring blood pressure using Photoplethysmography (PPG). It is a noninvasive, cuffless and painless technique that deploys infrared light to detect small variation in blood volume in the tissues with each cardiac cycle. Few specific features (viz. systolic upstroke time (ST), diastolic time (DT) and the time delay between the systolic and diastolic peak (T1)) of the waveform obtained via this technique were examined and correlated with the arterial blood pressure in 22 subjects of two age groups i) 18-25 years ii) 26-50 years. It was observed that there is a good correlation of blood pressure (both systolic blood pressure (SBP) and diastolic blood pressure (DBP)) with diastolic time and also with the time delay between systolic and diastolic peak.
引用
收藏
页码:254 / 257
页数:4
相关论文
共 50 条
  • [41] Cuffless Blood Pressure Estimation Model Using Clustering Techniques
    Baek, Jin-Hyeok
    Lee, Bongjin
    Rachim, Vega Pradana
    Jang, Jaeseong
    Kim, Sangkyu
    Kim, Heechan
    Park, Sung-Min
    IEEE SENSORS JOURNAL, 2024, 24 (20) : 32444 - 32454
  • [42] Continuous Cuffless Blood Pressure Measurement Using Body Sensors
    Heydari, Fatemeh
    Ebrahim, Malikeh P.
    Wu, Taiyang
    Walkert, Katie
    Joe, Keith
    Redoute, Jean-Michel
    Yuce, Mehmet R.
    2018 IEEE SENSORS, 2018, : 180 - 183
  • [44] Coherence Analysis of Invasive Blood Pressure and Its Noninvasive Indicators for Improvement of Cuffless Measurement Accuracy
    Ding, Xiao-Rong
    Yan, Bryan P.
    Zhang, Yuan-Ting
    Liu, Jing
    Su, Peng
    Zhao, Ni
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 2255 - 2258
  • [45] Is cuffless blood pressure measurement already here?
    van Helmond, Noud
    Martin, Seth S.
    Plante, Timothy B.
    JOURNAL OF HYPERTENSION, 2020, 38 (04) : 774 - 775
  • [46] Continual learning for cuffless blood pressure estimation
    Zhang, Chunlin
    Wang, Wenyan
    Song, Xinyue
    Lin, Yuxuan
    Chen, Yifan
    Ding, Xiaorong
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 92
  • [47] UPR-BP: Unsupervised Photoplethysmography Representation Learning for Noninvasive Blood Pressure Estimation
    Ma C.
    Zhang P.
    Song F.
    Liu Z.
    Feng Y.
    He Y.
    Zhang G.
    IEEE Transactions on Artificial Intelligence, 2024, 5 (09): : 1 - 12
  • [48] Development of a Cuffless Blood Pressure Measurement System
    Shyu, Liang-Yu
    Kao, Yao-Lin
    Tsai, Wen-Ya
    Hu, Weichih
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 2040 - 2043
  • [49] PCA-Based Multi-Wavelength Photoplethysmography Algorithm for Cuffless Blood Pressure Measurement on Elderly Subjects
    Liu, Jing
    Qiu, Shirong
    Luo, Ningqi
    Lau, Sze-Kei
    Yu, Hui
    Kwok, Timothy
    Zhang, Yuan-Ting
    Zhao, Ni
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2021, 25 (03) : 663 - 673
  • [50] The Wavelet Transform of Pulse Wave and Electrocardiogram Improves Accuracy of Blood Pressure Estimation in Cuffless Blood Pressure Measurement
    Yamanaka, Syunsuke
    Morikawa, Koji
    Yamamura, Osamu
    Morita, Hiroshi
    Huh, Ji Young