Non-Invasive Hemodynamic Assessment of Aortic Coarctation: Validation with In Vivo Measurements

被引:51
|
作者
Itu, Lucian [1 ,2 ]
Sharma, Puneet [1 ]
Ralovich, Kristof [1 ,3 ]
Mihalef, Viorel [1 ]
Ionasec, Razvan [1 ]
Everett, Allen [4 ]
Ringel, Richard [4 ]
Kamen, Ali [1 ]
Comaniciu, Dorin [1 ]
机构
[1] Siemens Corp Res & Technol, Princeton, NJ 08540 USA
[2] Transilvania Univ Brasov, Brasov 500024, Romania
[3] Tech Univ Munich, D-85748 Garching, Germany
[4] Johns Hopkins Univ, Sch Med, Baltimore, MD 21287 USA
关键词
Pressure gradient; CFD; Reduced-order models; Coarctation of aorta; Non-invasive; PC-MRI; PEAK-TO-PEAK; BLOOD-FLOW; ARTERIAL; PRESSURE; SIMULATION; GRADIENT; MODELS;
D O I
10.1007/s10439-012-0715-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We propose a CFD-based approach for the non-invasive hemodynamic assessment of pre- and post-operative coarctation of aorta (CoA) patients. Under our approach, the pressure gradient across the coarctation is determined from computational modeling based on physiological principles, medical imaging data, and routine non-invasive clinical measurements. The main constituents of our approach are a reduced-order model for computing blood flow in patient-specific aortic geometries, a parameter estimation procedure for determining patient-specific boundary conditions and vessel wall parameters from non-invasive measurements, and a comprehensive pressure-drop formulation coupled with the overall reduced-order model. The proposed CFD-based algorithm is fully automatic, requiring no iterative tuning procedures for matching the computed results to observed patient data, and requires approximately 6-8 min of computation time on a standard personal computer (Intel Core2 Duo CPU, 3.06 GHz), thus making it feasible for use in a clinical setting. The initial validation studies for the pressure-drop computations have been performed on four patient datasets with native or recurrent coarctation, by comparing the results with the invasively measured peak pressure gradients recorded during routine cardiac catheterization procedure. The preliminary results are promising, with a mean absolute error of less than 2 mmHg in all the patients.
引用
收藏
页码:669 / 681
页数:13
相关论文
共 50 条
  • [21] NON-INVASIVE ASSESSMENT OF VALVAR AORTIC-STENOSIS
    RITTER, S
    BOXER, R
    HUTCHEON, G
    GERSONY, W
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1982, 49 (04): : 974 - 974
  • [22] NON-INVASIVE ASSESSMENT OF AORTIC-STENOSIS IN CHILDREN
    ARENSMAN, FW
    KAPLAN, S
    JAMES, FW
    MEYER, RA
    SCHWARTZ, DC
    [J]. CIRCULATION, 1979, 60 (04) : 252 - 252
  • [23] Future of invasive and non-invasive hemodynamic assessment for coronary artery disease management
    Scala, Antonella
    Marchini, Federico
    Meossi, Sofia
    Zanarelli, Luca
    Sanguettoli, Federico
    Frascaro, Federica
    Bianchi, Nicola
    Cocco, Marta
    Erriquez, Andrea
    Tonet, Elisabetta
    Campo, Gianluca
    Pavasini, Rita
    [J]. MINERVA CARDIOLOGY AND ANGIOLOGY, 2024, 72 (04) : 385 - 404
  • [24] Validation of non-invasive hemodynamic monitoring with Nexfin in critically ill patients
    K Van de Vijver
    A Verstraeten
    C Gillebert
    U Maniewski
    M Gabrovska
    D Viskens
    N Van Regenmortel
    I De laet
    K Schoonheydt
    H Dits
    M Malbrain
    [J]. Critical Care, 15 (Suppl 1):
  • [25] Non-invasive hemodynamic assessment in LVAD patients: validation of a bed-side echocardiographic approach.
    Giordana, F.
    Frea, S.
    Pidello, S. Stefano
    Bovolo, V.
    Galluzzo, A.
    Bergerone, S.
    Baronetto, A.
    Centofanti, P.
    Rinaldi, M.
    Gaita, F.
    [J]. EUROPEAN JOURNAL OF HEART FAILURE, 2016, 18 : 82 - 82
  • [26] MRI as a tool for non-invasive vascular profiling: a pilot study in patients with aortic coarctation
    Kelm, Marcus
    Goubergrits, Leonid
    Fernandes, Joao Filipe
    Biocca, Lucio
    Pongiglione, Giacomo
    Muthurangu, Vivek
    Khushnood, Abbas
    Secinaro, Aurelio
    Chinali, Marcello
    Schubert, Stephan
    Berger, Felix
    Kuehne, Titus
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (01) : 103 - 112
  • [27] A Non-Invasive in vivo Assessment of the Akimba Mouse Eye
    Lai, M. C.
    Rahman, I. S. Ali
    Li, C.
    Vagaja, N.
    Rakoczy, E. P.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [28] NON-INVASIVE MEASUREMENTS
    ROUGHTON, JE
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1982, 15 (12): : 1257 - 1270
  • [29] Non-invasive measurements of cerebral carbohydrate metabolism by in vivo NMR
    Gruetter, R
    Pfeuffer, J
    Tkac, I
    Choi, IY
    Ugurbil, K
    [J]. JOURNAL OF NEUROCHEMISTRY, 1999, 73 : S102 - S102
  • [30] 3.2 A Comparison of the Vicorder Apparatus with Sphygmocor Device for the Non-Invasive Assessment of Aortic Blood Pressure: An Invasive Validation Study
    G. Pucci
    J. Cheriyan
    A. Hubsch
    S. S. Hickson
    T. Watson
    G. Schillaci
    I. B. Wilkinson