In vitro evaluation of clot capture efficiency of an absorbable vena cava filter

被引:8
|
作者
Dria, Stephen J. [1 ]
Eggers, Mitchell D. [1 ,2 ,3 ]
机构
[1] Adient Med, Pearland, TX USA
[2] Baylor Coll Med, Dept Orthoped Surg, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Intervent Radiol, Houston, TX 77030 USA
关键词
DEEP-VEIN THROMBOSIS; PULMONARY-EMBOLISM; VENOUS THROMBOEMBOLISM; COMPLICATIONS; PREVENTION; MANAGEMENT; RATES; VIVO;
D O I
10.1016/j.jvsv.2016.05.006
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: The purpose of this study was to determine the in vitro dot capture efficiency (CCE) of an investigational absorbable inferior vena cava filter (IVCF) vs the Greenfield IVCF. Methods: Investigational absorbable and Greenfield filters were challenged with polyacrylamide clot surrogates ranging from 3 x 5 to 10 x 24 mm (diameter x length) in a flow loop simulating the venous system. Filters were challenged with clots until CCE standard error of 5% or less was achieved under binomial statistics. Pressure gradients across the filters were measured for the largest size clot, enabling calculation of forces on the filter. Results: The in vitro CCE of the absorbable IVCF was statistically similar to that of the Greenfield filter for all clot sizes apart from the 3 x 10-mm clot, for which there was statistically significant difference between filter CCEs (absorbable filter, 59%; Greenfield filter, 31%; P =.0001). CCE ranged from an average 32% for the 3 x 5-mm clot to 100% for 7 x 10-mm and larger clots for the absorbable IVCF. Pressure gradient across the absorbable filter with 10 x 24-mm clot averaged 0.14 mm Hg, corresponding to a net force on the filter of 2.1 x 10(-3) N, compared with 0.39 mm Hg or 5.8 x 10(-3) N (P <.001) for the Greenfield filter. Conclusions: CCE of the absorbable filter was statistically similar to or an improvement on that of the Greenfield stainless steel filter for all dot sizes tested. CCE of the Greenfield filter in this study aligned with data from previous studies. Given the efficacy of the Greenfield filter in attenuating the risk of pulmonary embolism, the current study suggests that the absorbable filter may be a viable candidate for subsequent human testing.
引用
收藏
页码:472 / 478
页数:7
相关论文
共 50 条
  • [1] Limitations of using synthetic blood clots for measuring in vitro clot capture efficiency of inferior vena cava filters
    Robinson, Ronald A.
    Herbertson, Luke H.
    Das, Srilekha Sarkar
    Malinauskas, Richard A.
    Pritchard, William F.
    Grossman, Laurence W.
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2013, 6 : 49 - 57
  • [2] Numerical simulation and in vitro experimental study of thrombus capture efficiency of a new retrievable vena cava filter
    Feng, Haiquan
    Li, Changsheng
    Feng, Haoxiang
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2023, 26 (16) : 2034 - 2046
  • [3] Pilot in vivo study of an absorbable polydioxanone vena cava filter
    Eggers, Mitchell D.
    McArthur, Mark J.
    Figueira, Tomas A.
    Abdelsalam, Mohamed E.
    Dixon, Katherine P.
    Pageon, Laura R.
    Wallace, Michael J.
    Huang, Steven Y.
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2015, 3 (04) : 409 - 420
  • [4] Randomized Controlled Study of an Absorbable Vena Cava Filter in a Porcine Model
    Eggers, Mitchell
    Rousselle, Serge
    Urtz, Mark
    Albright, Rhonda
    Will, Alice
    Jourden, Bettina
    Godshalk, Cynthia
    Dria, Stephen
    Huang, Steven
    Steele, Joseph
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2019, 30 (09) : 1487 - 1494
  • [5] In Vitro Clot Trapping Efficiency of the FDA Generic Inferior Vena Cava Filter in an Anatomical Model: An Experimental Fluid–Structure Interaction Benchmark
    J. M. Riley
    N. S. Price
    H. M. Saaid
    B. C. Good
    K. I. Aycock
    B. A. Craven
    K. B. Manning
    Cardiovascular Engineering and Technology, 2021, 12 : 339 - 352
  • [6] Vena cava filter [Vena-cava-filter]
    Helmberger T.
    Der Radiologe, 2007, 47 (5): : 439 - 442
  • [7] Percutaneous management of extensive clot trapped in a temporary vena cava filter
    Vrachliotis, TG
    Tachtaras, A
    Doundoulakis, N
    Sheiman, RG
    Papadopoulos, V
    Stringaris, KA
    JOURNAL OF ENDOVASCULAR THERAPY, 2003, 10 (05) : 1001 - 1005
  • [8] Hemodynamic effects of clot entrapment in the TrapEase inferior vena cava filter
    Leask, RL
    Johnston, KW
    Ojha, M
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (05) : 485 - 490
  • [9] In Vitro Evaluation of Capturing and Dissolving the Clot by Magnetic Nanoparticles Carrying a Thrombolytic Agents Instead of Temporary Inferior Vena Cava (IVC) Filter
    Moghanizadeh, Abbas
    Ashrafizadeh, Fakhreddin
    Varshosaz, Jaleh
    Kharaziha, Mahshid
    JOURNAL OF BIOMEDICAL NANOTECHNOLOGY, 2020, 16 (11) : 1623 - 1632
  • [10] In Vitro Clot Trapping Efficiency of the FDA Generic Inferior Vena Cava Filter in an Anatomical Model: An Experimental Fluid-Structure Interaction Benchmark
    Riley, J. M.
    Price, N. S.
    Saaid, H. M.
    Good, B. C.
    Aycock, K. I.
    Craven, B. A.
    Manning, K. B.
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2021, 12 (03) : 339 - 352