Accuracy of a feedback device for cardiopulmonary resuscitation on a dental chair

被引:11
|
作者
Segal, Nicolas [1 ]
Laurent, Florian [2 ]
Maman, Louis [2 ]
Plaisance, Patrick
Augustin, Pascal [3 ,4 ]
机构
[1] Univ Paris 07, Hosp Lariboisiere, Serv Accueil Urgences, Emergency Dept, F-75010 Paris, France
[2] Univ Paris 05, Dept Oral Med & Oral Surg, Fac Dent, Paris, France
[3] Univ Paris 07, Bichat Claude Bernard Hosp, Dept Anesthesiol, F-75010 Paris, France
[4] Univ Paris 07, Bichat Claude Bernard Hosp, Surg Intens Care Unit, F-75010 Paris, France
关键词
HOSPITAL CARDIAC-ARREST; CHEST-WALL DECOMPRESSION; EMERGENCY MANAGEMENT; CPR PERFORMANCE; LIFE-SUPPORT; COMPRESSION; EFFICACY; PREVALENCE; GUIDELINES; EXPERIENCE;
D O I
10.1136/emermed-2011-200003
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background Conflicting studies exist about the effectiveness of cardiopulmonary resuscitation (CPR) on a dental chair. In some situations, dental surgeons are obliged to perform CPR with the patient on the chair. Feedback devices are supposed to guide the compression depth in order to improve the quality of CPR, but some devices are based on an accelerometer that can theoretically report erroneous results because of the lack of rigidity of a dental chair. Objective The aim of this study was to evaluate the accuracy of these devices to guide chest compressions on a dental chair. Methods A prospective, randomised, crossover, equivalence/non-inferiority study was conducted to compare the values of compression depths provided by the feedback device (Real CPR Help (R), delivered by Zoll (c) Medical Corporation, Chelmsford, MA, USA) with the real measurements provided by the manikin (Resusci Anne (R) Advanced SkillTrainer, Laerdal Medical AS (c), Norway). Chest-compression-only CPR was performed by 15 Basic Life Support instructors who carried out two rounds of continuous CPR for 2 min each. Data were analysed with a correlation test, a Bland-Altman method and a Wilcoxon test. Statistical significance was defined as p<0.05. Results A significant difference was found between the mean depths of compression measured by the feedback device and the manikin on a dental chair and on the floor (p<0.0001). The feedback device overestimated the depth of chest compressions, and Bland-Altman analysis demonstrated poor agreement. Conclusion This study indicates that feedback devices with accelerometer technology are not sufficiently reliable to ensure adequate chest compression on dental chairs.
引用
收藏
页码:890 / 893
页数:4
相关论文
共 50 条
  • [1] Efficacy of cardiopulmonary resuscitation performed in a dental chair
    Lepere, AJ
    Finn, J
    Jacobs, I
    AUSTRALIAN DENTAL JOURNAL, 2003, 48 (04) : 244 - 247
  • [2] Design of a new feedback device for Cardiopulmonary Resuscitation
    Lhoucine, Ben Taleb
    Elmaati, Essoukaki
    Aissam, Lyazidi
    Azeddine, Mouhsen
    Abdelhadi, Assir
    Mourad, Rattal
    Mohammed, Harmouchi
    2018 IEEE 4TH MIDDLE EAST CONFERENCE ON BIOMEDICAL ENGINEERING (MECBME), 2018, : 90 - 95
  • [3] Using a Feedback Device to Improve Quality of Cardiopulmonary Resuscitation
    Kupchik, Nicole
    Nehmer, Maria
    CRITICAL CARE NURSE, 2016, 36 (02) : E42 - E42
  • [4] Development of a Cardiopulmonary Resuscitation Feedback Learning Device Using Maltisensor
    Hayashi, Hidehiko
    Minazuki, Akinori
    2017 14TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING/ELECTRONICS, COMPUTER, TELECOMMUNICATIONS AND INFORMATION TECHNOLOGY (ECTI-CON), 2017, : 473 - 476
  • [5] THE SHEFFIELD CHAIR FOR CARDIOPULMONARY-RESUSCITATION
    EDBROOKE, DL
    WARDROPE, J
    CROSBY, AC
    MURRAY, R
    FERGUSON, DG
    ARCHIVES OF EMERGENCY MEDICINE, 1987, 4 (03): : 151 - 154
  • [6] CARDIOPULMONARY RESUSCITATION IN DENTAL PRACTICE
    ODAY, RA
    DRIGGS, RL
    DENTAL CLINICS OF NORTH AMERICA, 1973, 17 (02) : 329 - 341
  • [7] Feedback during cardiopulmonary resuscitation
    Bohn, Andreas
    Gude, Philipp
    CURRENT OPINION IN ANESTHESIOLOGY, 2008, 21 (02) : 200 - 203
  • [8] Effect of Audiovisual Cardiopulmonary Resuscitation Feedback Device on Improving Chest Compression Quality
    Chia-Ying Lin
    Shao-Hsuan Hsia
    En-Pei Lee
    Oi-Wa Chan
    Jainn-Jim Lin
    Han-Ping Wu
    Scientific Reports, 10
  • [9] Effect of Audiovisual Cardiopulmonary Resuscitation Feedback Device on Improving Chest Compression Quality
    Lin, Chia-Ying
    Hsia, Shao-Hsuan
    Lee, En-Pei
    Chan, Oi-Wa
    Lin, Jainn-Jim
    Wu, Han-Ping
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [10] Hand injuries caused by feedback device usage during cardiopulmonary resuscitation training
    Hong, Jun Young
    Oh, Je Hyeok
    Kim, Chan Woong
    Lee, Dong Hoon
    RESUSCITATION, 2016, 107 : E3 - E4