Performance evaluation of the microINR® point-of-care INR-testing system

被引:5
|
作者
Joubert, J. [1 ,2 ]
van Zyl, M. C. [2 ]
Raubenheimer, J. [3 ,4 ]
机构
[1] Univ Hosp, Haematol Acad Labs, NHLS, Bloemfontein, South Africa
[2] Univ Free State, Dept Haematol & Cell Biol, Fac Hlth Sci, Bloemfontein, South Africa
[3] Univ Free State, Dept Biostat, Fac Hlth Sci, Bloemfontein, South Africa
[4] Univ Sydney, Discipline Pharmacol, Sydney Med Sch, Sydney, NSW, Australia
关键词
international normalised ratio; micro international normalised ratio((R)); oral anticoagulant therapy; performance evaluation; point-of-care; COAGULATION MONITOR; PROTHROMBIN TIME; ACCURACY; PRECISION;
D O I
10.1111/ijlh.12749
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionPoint-of-care International Normalised Ratio (INR) testing is used frequently. We evaluated the microINR((R)) POC system for accuracy, precision and measurement repeatability, and investigated instrument and test chip variability and error rates. MethodsVenous blood INRs of 210 patients on warfarin were obtained with Thromborel((R)) S on the Sysmex CS-2100i((R)) analyser and compared with capillary blood microINR((R)) values. Precision was assessed using control materials. Measurement repeatability was calculated on 51 duplicate finger-prick INRs. Triplicate finger-prick INRs using three different instruments (30 patients) and three different test chip lots (29 patients) were used to evaluate instrument and test chip variability. ResultsLinear regression analysis of microINR((R)) and Sysmex CS2100i((R)) values showed a correlation coefficient of 0.96 (P<.0001) and a positive proportional bias of 4.4%. Dosage concordance was 93.8% and clinical agreement 95.7%. All acceptance criteria based on ISO standard 17593:2007 system accuracy requirements were met. Control material coefficients of variation (CV) varied from 6.2% to 16.7%. The capillary blood measurement repeatability CV was 7.5%. No significant instrument (P=.93) or test chip (P=.81) variability was found, and the error rate was low (2.8%). ConclusionThe microINR((R)) instrument is accurate and precise for monitoring warfarin therapy.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 50 条
  • [41] Reliability of patient self-testing and point-of-care testing with the Harmony INR Monitoring System: A two-year study.
    Earp, BC
    Jacobson, A
    Ruybalid, L
    Howe, B
    Gunneman, T
    Peterson, M
    Buenjemia, J
    [J]. CLINICAL CHEMISTRY, 2002, 48 (06) : A199 - A199
  • [42] Artificial intelligence performance in testing microfluidics for point-of-care
    Doganay, Mert Tunca
    Chakraborty, Purbali
    Bommakanti, Sri Moukthika
    Jammalamadaka, Soujanya
    Battalapalli, Dheerendranath
    Madabhushi, Anant
    Draz, Mohamed S.
    [J]. Lab on a Chip, 2024, 24 (21) : 4998 - 5008
  • [43] The Impact of Point-of-Care INR Checks on Endoscopy Cancellations
    Sharma, Ayush
    Mukewar, Saurabh
    Krishnamoorthi, Rajesh
    Rustagi, Tarun
    Majumder, Shounak
    Okafor, Philip N.
    [J]. AMERICAN JOURNAL OF GASTROENTEROLOGY, 2016, 111 : S177 - S177
  • [44] Evaluation of the Meritas® BNP Test for Point-of-Care Testing
    Olausson, Johan
    Peterson, Christer
    Bergstrom, Mats
    Larsson, Anders
    [J]. CLINICAL LABORATORY, 2015, 61 (07) : 727 - 730
  • [45] INR Point of Care Testing Trial Project
    Abby, Sereda
    Beth, Luhowy
    Olga, Norrie
    Jo-Anne, Kilgour
    [J]. INTERNATIONAL JOURNAL OF STROKE, 2017, 12 : 45 - 45
  • [46] Comparison of emergency department point-of-care international normalised ratio (INR) testing with laboratory-based testing
    Celenza, Antonio
    Skinner, Kirsty
    [J]. EMERGENCY MEDICINE JOURNAL, 2011, 28 (02) : 136 - 140
  • [47] Magnitude of INR variation attributable to differences between point-of-care prothrombin time testing devices
    Jacobson, AK
    Peterson, M
    Gunneman, T
    Westengard, J
    Ruybalid, L
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (05) : 234A - 235A
  • [48] Efficacy of a Point-of-Care INR Device Compared to Laboratory Testing In Patients Starting on a Warfarin Nomogram
    Lewis, Geoffrey M.
    Evans, Mark
    Hutton, Brian
    Rodger, Marc
    [J]. CANADIAN JOURNAL OF HOSPITAL PHARMACY, 2007, 60 : 65 - 65
  • [49] Point-of-care testing: Administration within a health system
    Main, RI
    Kiechle, FL
    [J]. LABORATORY MEDICINE, 2000, 31 (08) : 453 - 459
  • [50] Evaluation of point of care INR in children
    Greenway, A. L.
    Ignjatovic, V
    Summerhayes, R.
    Newall, F.
    Burgess, J.
    DeRosa, L.
    Monagle, P. T.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 : 466 - 466