Performance evaluation of the microINR® point-of-care INR-testing system

被引:5
|
作者
Joubert, J. [1 ,2 ]
van Zyl, M. C. [2 ]
Raubenheimer, J. [3 ,4 ]
机构
[1] Univ Hosp, Haematol Acad Labs, NHLS, Bloemfontein, South Africa
[2] Univ Free State, Dept Haematol & Cell Biol, Fac Hlth Sci, Bloemfontein, South Africa
[3] Univ Free State, Dept Biostat, Fac Hlth Sci, Bloemfontein, South Africa
[4] Univ Sydney, Discipline Pharmacol, Sydney Med Sch, Sydney, NSW, Australia
关键词
international normalised ratio; micro international normalised ratio((R)); oral anticoagulant therapy; performance evaluation; point-of-care; COAGULATION MONITOR; PROTHROMBIN TIME; ACCURACY; PRECISION;
D O I
10.1111/ijlh.12749
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionPoint-of-care International Normalised Ratio (INR) testing is used frequently. We evaluated the microINR((R)) POC system for accuracy, precision and measurement repeatability, and investigated instrument and test chip variability and error rates. MethodsVenous blood INRs of 210 patients on warfarin were obtained with Thromborel((R)) S on the Sysmex CS-2100i((R)) analyser and compared with capillary blood microINR((R)) values. Precision was assessed using control materials. Measurement repeatability was calculated on 51 duplicate finger-prick INRs. Triplicate finger-prick INRs using three different instruments (30 patients) and three different test chip lots (29 patients) were used to evaluate instrument and test chip variability. ResultsLinear regression analysis of microINR((R)) and Sysmex CS2100i((R)) values showed a correlation coefficient of 0.96 (P<.0001) and a positive proportional bias of 4.4%. Dosage concordance was 93.8% and clinical agreement 95.7%. All acceptance criteria based on ISO standard 17593:2007 system accuracy requirements were met. Control material coefficients of variation (CV) varied from 6.2% to 16.7%. The capillary blood measurement repeatability CV was 7.5%. No significant instrument (P=.93) or test chip (P=.81) variability was found, and the error rate was low (2.8%). ConclusionThe microINR((R)) instrument is accurate and precise for monitoring warfarin therapy.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 50 条
  • [1] Point-of-care testing INR: an overview
    Barcellona, Doris
    Fenu, Lara
    Marongiu, Francesco
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (06) : 800 - 805
  • [2] Evaluation of the analytical and clinical performance of the Micropoint qLabs INR point-of-care system
    van den Heuvel, J. K.
    Kena, N.
    van Hattum, T.
    van Dolder, D.
    van Wijnen, M.
    [J]. THROMBOSIS RESEARCH, 2019, 176 : 104 - 107
  • [3] INR Management of an Antiphospholipid Syndrome Patient With Point-of-Care INR Testing
    Dush, Aaron
    Erdeljac, H. Paige
    [J]. JOURNAL OF PHARMACY PRACTICE, 2020, 33 (03) : 390 - 391
  • [4] The microINR portable coagulometer: analytical quality and user-friendliness of a PT (INR) point-of-care instrument
    Larsen, Pia Bukmann
    Storjord, Elin
    Bakke, Asne
    Bukve, Tone
    Christensen, Mikael
    Eikeland, Joakim
    Haugen, Vegar Engeland
    Husby, Kristin
    McGrail, Rie
    Mikaelsen, Solveig Meier
    Monsen, Grete
    Moller, Mette Fogh
    Nybo, Jan
    Revsholm, Jesper
    Risoy, Aslaug Johanne
    Skalsvik, Unni Marie
    Strand, Heidi
    Teruel, Reyes Serrano
    Theodorsson, Elvar
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2017, 77 (02): : 115 - 121
  • [5] Evaluating the Effect of Timing on Point-of-Care Testing of INR
    Son, Joanne
    Kim, Elise M. M.
    [J]. JOURNAL OF PHARMACY PRACTICE, 2024, 37 (03) : 703 - 707
  • [6] Point-of-care testing of INR in acute ischemic stroke
    Newcommon, Nancy
    Hill, Michael
    Green, Teri
    [J]. STROKE, 2007, 38 (02) : 610 - 610
  • [7] Reliability of point-of-care testing of INR in acute stroke
    Green, Theresa L.
    Mansoor, Adnan
    Newcommon, Nancy
    Stephenson, Caroline
    Stewart, Eileen
    Hill, Michael D.
    [J]. CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 2008, 35 (03) : 348 - 351
  • [8] Potential impact of point-of-care INR testing on intravenous thrombolysis
    Conforto, Adriana Bastos
    [J]. ARQUIVOS DE NEURO-PSIQUIATRIA, 2014, 72 (07) : 485 - 486
  • [9] INR management of an anti-phospholipid syndrome patient with point-of-care INR testing
    Dush, Aaron D.
    Stewart, H. Paige
    Klaczak, Kendra
    [J]. JOURNAL OF THROMBOSIS AND THROMBOLYSIS, 2015, 39 (03) : 412 - 412
  • [10] Implementation of point-of-care INR testing to improve effectiveness of anticoagulation clinic
    Mohammad, I. N.
    [J]. EUROPEAN HEART JOURNAL, 2023, 44 : 24 - 24