Expression of the uncertainty of measurement in the calibration of thermometers - Part I: Standard platinum resistance thermometers

被引:10
|
作者
Lira, I
Camarano, D
Villalobos, JP
Santiago, F
机构
[1] Pontificia Univ Catolica Chile, Escuela Ingn, Santiago 22, Chile
[2] Natl Metrol Inst Brazil, BR-25250020 Xerem, RJ, Brazil
关键词
D O I
10.1088/0026-1394/36/2/5
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A model is proposed for the calibration of a standard platinum resistance thermometer (SPRT) in accordance with the BIPM recommended procedures to realize the International Temperature Scale of 1990 (ITS-90). The model is used to obtain the standard uncertainty of measurement associated with the temperature indicated by the thermometer. The difference between calibration uncertainty and user's uncertainty is established. Equations are illustrated by an example that considers the ITS-90 subrange from 273.15 K to 692.677 K. The approach closely follows the rules in the ISO Guide to the Expression of Uncertainty in Measurement.