Realization of a radiation temperature scale from 0 °C to 232 °C by a thermal infrared thermometer based on a multiple-fixed-point technique

被引:9
|
作者
Yoo, Yong Shim [1 ]
Kim, Bong-Hak [1 ]
Lim, Sun Do [1 ]
Park, Seung-Nam [1 ]
Park, Seongchong [1 ]
机构
[1] Korea Res Inst Stand & Sci, Div Phys Metrol, Taejon 305340, South Korea
关键词
CALIBRATION;
D O I
10.1088/0026-1394/50/4/409
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The radiation temperature scale for a pyroelectric detector based thermal infrared thermometer with its spectral response from 8 mu m to 14 mu m was realized in the temperature range from 0 degrees C to 232 degrees C by using four fixed-point blackbodies (ice, Ga, In and Sn). The Planck version of the Sakuma-Hattori equation was used to interpolate the scale between the fixed-point temperatures that are corrected by considering a size-of-source effect (SSE). The expanded uncertainties (k = 2) of the scale were estimated to be 108 mK for ice, 99 mK for Ga, 175 mK for In and 234 mK for Sn.
引用
收藏
页码:409 / 416
页数:8
相关论文
共 14 条
  • [1] An InGaAs Detector Based Radiation Thermometer and Fixed-point Blackbodies for Temperature Scale Realization at NIM
    Hao, X.
    Yuan, Z.
    Wang, J.
    Lu, X.
    [J]. TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 : 643 - 648
  • [2] Middle Temperature Scale for Infrared Radiation Thermometer Calibrated Against Multiple Fixed Points
    Y. Shimizu
    J. Ishii
    [J]. International Journal of Thermophysics, 2008, 29 : 1014 - 1025
  • [3] Middle temperature scale for infrared radiation thermometer calibrated against multiple fixed points
    Shimizu, Y.
    Ishii, J.
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (03) : 1014 - 1025
  • [4] Bilateral Comparison Between CMI and CEM in Radiance Temperature Scale Realization from 232 °C to 1085 °C
    Martin, M. J.
    Knazovicka, L.
    del Campo, D.
    Strnad, R.
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2014, 35 (3-4) : 475 - 484
  • [5] Validation of the fixed-point realization of the thermocouple scale using a radiation thermometer in the temperature range between the freezing points of Ag and Pd
    Kim, Yong-Gyoo
    Kim, Bong Hak
    Yang, Inseok
    [J]. METROLOGIA, 2010, 47 (03) : 239 - 247
  • [6] FIXED-POINT TECHNIQUE FOR APPROXIMATING THE ITS-90 BETWEEN 420-DEGREES-C AND 1085-DEGREES-C WITH AN INFRARED THERMOMETER
    BATTUELLO, M
    LANZA, F
    RICOLFI, T
    [J]. METROLOGIA, 1990, 27 (02) : 75 - 82
  • [7] Realization of radiance temperature scale from 500K to 1,250K by a radiation thermometer with a thermal detector
    Park, Seung-Nam
    Kim, Bong-Hak
    Park, Chul-Woung
    Lee, Dong-Hoon
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (01) : 301 - 311
  • [8] Realization of Radiance Temperature Scale from 500K to 1,250K by a Radiation Thermometer with a Thermal Detector
    Seung-Nam Park
    Bong-Hak Kim
    Chul-Woung Park
    Dong-Hoon Lee
    [J]. International Journal of Thermophysics, 2008, 29 : 301 - 311
  • [9] Realization of the temperature scale in the range from 234.3 K (Hg triple point) to 1084.62°C (Cu freezing point) in croatia
    Zvizdic, Davor
    Veliki, Tomislav
    Bermanec, Lovorka Grgec
    [J]. INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (03) : 984 - 990
  • [10] Realization of the Temperature Scale in the Range from 234.3 K (Hg Triple Point) to 1084.62°C (Cu Freezing Point) in Croatia
    Davor Zvizdic
    Tomislav Veliki
    Lovorka Grgec Bermanec
    [J]. International Journal of Thermophysics, 2008, 29 : 984 - 990