Air dissipation in saline breast implants

被引:3
|
作者
Schanzer, A [1 ]
Wong, GB [1 ]
机构
[1] Univ Calif Davis, Davis Med Ctr, Div Plast Surg, Sacramento, CA 95817 USA
关键词
D O I
10.1097/00006534-200206000-00010
中图分类号
R61 [外科手术学];
学科分类号
摘要
Residual air within saline breast implants can cause patient discomfort due to the mechanical and auditor, effects of sloshing. Small amounts of air have no clinical significance, but if larger quantities are present and audible the patient is reassured that the implant shell is gas-permeable and that the air ill dissipate/diffuse. This study examined the time necessary for air dissipation in saline breast implants. Twelve McGhan style #68 saline breast implants were divided into two groups: group A, which included six implants with a size of 240 cc. and group B, which included six implants with a size of 270 cc. The implants were filled with room-temperature, intravenous. normal saline to their designated volumes, Plus 30 cc of overfill. All air was evacuated, and each implant was inoculated with 5 cc of air, The implants were then submerged in a single tank of normal saline at 37degreesC, at a depth of 20.4 cut to replicate capillary pressure. Intragroup analysis showed the air bubble was absent in group A at an average of 35.3 days (variance = 4.13) and in group B at an average of 38.0 days (variance = 0). If audible intraluminal air is present in the clinical setting, the patient can be reassured that the problem will resolve in approximately 30 class or less.
引用
收藏
页码:2238 / 2241
页数:4
相关论文
共 50 条