Assessing the Performance of Handheld Glucose Testing for Critical Care

被引:15
|
作者
Kost, Gerald J. [1 ,2 ]
Tran, Nam K. [1 ,2 ]
Louie, Richard F. [1 ,2 ]
Gentile, Nicole L. [1 ,2 ]
Abad, Victor J. [3 ]
机构
[1] Univ Calif Davis, Point Care Testing Ctr Teaching & Res, Sch Med, Davis, CA 95616 USA
[2] Univ Calif Davis, Univ Calif Lawrence Livermore Natl Lab, Ctr Point Care Technol, Davis, CA 95616 USA
[3] Epsilon Grp LLC, Charlottesville, VA USA
关键词
D O I
10.1089/dia.2008.0049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: We assessed the performance of a point-of-care (POC) glucose meter system (GMS) with multitasking test strip by using the locally-smoothed (LS) median absolute difference (MAD) curve method in conjunction with a modified Bland-Altman difference plot and superimposed International Organization for Standardization (ISO) 15197 tolerance bands. We analyzed performance for tight glycemic control (TGC). Methods: A modified glucose oxidase enzyme with a multilayer-gold, multielectrode, four-well test strip (StatStrip (TM), NOVA Biomedical, Waltham, MA) was used. There was no test strip calibration code. Pragmatic comparison was done of GMS results versus paired plasma glucose measurements from chemistry analyzers in clinical laboratories. Venous samples (n = 1,703) were analyzed at 35 hospitals that used 20 types of chemistry analyzers. Erroneous results were identified using the Bland-Altman plot and ISO 15197 criteria. Discrepant values were analyzed for the TGC interval of 80-110 mg/dL. Results: The GMS met ISO 15197 guidelines; 98.6% (410 of 416) of observations were within tolerance for glucose <75 mg/dL, and for >= 75 mg/dL, 100% were within tolerance. Paired differences (handheld. minus reference) averaged -2.2 (SD 9.8) mg/dL; the median was -1 (range, -96 to 45) mg/dL. LS MAD curve analysis revealed satisfactory performance below 186 mg/dL; above 186 mg/dL, the recommended error tolerance limit (5 mg/dL) was not met. No discrepant values appeared. All points fell in Clarke Error Grid zone A. Linear regression showed y = 1.018x - 0.716 mg/dL, and r(2) = 0.995. Conclusions: LS MAD curves draw on human ability to discriminate performance visually. LS MAD curve and ISO 15197 performance were acceptable for TGC. POC and reference glucose calibration should be harmonized and standardized.
引用
收藏
页码:445 / 451
页数:7
相关论文
共 50 条
  • [41] Informatics requirements for critical care testing
    Blick, KE
    CLINICAL CHEMISTRY, 1999, 45 (06) : S37 - S37
  • [42] SFBC guidelines on critical care testing
    Vaubourdolle, Michel
    Alvarez, Jean-Claude
    Barbe, Francoise
    Beaudeux, Jean-Louis
    Boissier, Elodie
    Caillon, Helene
    Chatron, Philippe
    Joly-Guillou, Marie-Laure
    Mailloux, Agnes
    ANNALES DE BIOLOGIE CLINIQUE, 2016, 74 (02) : 130 - 155
  • [43] Using Point of Care Glucose Meters in the Critically Ill Assessing Meter Performance in the Clinical Context
    Schmolze, Daniel B.
    Horowitz, Gary L.
    Tolan, Nicole V.
    POINT OF CARE, 2016, 15 (04): : 137 - 143
  • [44] Critical care testing in the new millennium: the integration of point-of-care testing
    Okorodudu, AO
    Jacobs, E
    Fogh-Andersen, N
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 1 - 2
  • [45] Assessing the Performance of Handheld Laser Scanning for Individual Tree Mapping in an Urban Area
    Yang, Jinming
    Yuan, Wenwen
    Lu, Huicui
    Liu, Yuehan
    Wang, Yongkang
    Sun, Letong
    Li, Shimei
    Li, Haifang
    FORESTS, 2024, 15 (04):
  • [46] Performance and Limitations of Point-of-Care Glucose Meters for Measuring Blood Glucose Concentration in Patients with Critical Medical Conditions
    Zhang, Lei
    Schwartz, Carol
    Wang, Diping
    El Zein, Dima
    Krabill, Keith
    Howard, Susan
    Tomaszewski, John
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2015, 144 : A48 - A48
  • [47] Assessing pain, agitation and delirium on critical care
    Sayes, K.
    Lim, C.
    ANAESTHESIA, 2022, 77 : 19 - 19
  • [48] A critical view of critical velocity testing to predict performance
    Jolly, Stuart
    JOURNAL OF SPORTS SCIENCES, 2013, 31 (07) : 688 - 689
  • [49] Assessing glucose monitor performance - A standardized approach
    Mahoney, John J.
    Ellison, John M.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2007, 9 (06) : 545 - 552
  • [50] Blood glucose determination: Point of care testing
    Bennett, BD
    SOUTHERN MEDICAL JOURNAL, 1997, 90 (07) : 678 - 680