Comparative measurements of 60Co 'hot particle' depth-dose distributions and Monte Carlo calculations

被引:15
|
作者
Villarreal-Barajas, JE [1 ]
Darley, PJ [1 ]
Charles, MW [1 ]
机构
[1] Univ Birmingham, Sch Phys & Astron, Birmingham B15 2TT, W Midlands, England
关键词
D O I
10.1093/oxfordjournals.rpd.a032908
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The present study attempts to compare several possible measurement techniques and MCNP Monte Carlo calculations of beta/gamma doses from test 'bench-mark' Co-60 hot particle sources. Depth-dose distributions have been obtained by ionmetric measurements using ionisation extrapolation chamber (ECH) and densitometric measurements using radiation induced colour changes in radiochromic dye films (RDF). The ECH provides absorbed dose information averaged over specific areas dictated by the measuring electrode diameter (1 mm, 3 mm, 1 cm and 3 cm). The RDF displays full two-dimensional dose distributions in a single measurement with a high spatial resolution down to 25 mu m. In general there was good agreement between the two measurement techniques for the doses measured close to the surface after correction due to the non-linear response of the ionisation response relative to electrode separation (mainly due to the spatially non-uniform radiation field). However the agreement is markedly worse at greater depths where the dose is dominated by the gamma ray component. The measurements and MCNP4B calculations agree within a few per cent for the surface dose rate and to within similar to 30% at 30 mg.cm(-2). The reasons for these differences are discussed.
引用
收藏
页码:507 / 512
页数:6
相关论文
共 50 条
  • [1] Electron spectra measurements and Monte Carlo calculations for a 60Co standardised hot particle source
    Villarreal-Barajas, JE
    Bowler, MG
    Charles, MW
    [J]. RADIATION PROTECTION DOSIMETRY, 2003, 104 (02) : 113 - 119
  • [3] HEAVY PARTICLE COMPARATIVE-STUDY .1. DEPTH-DOSE DISTRIBUTIONS
    RAJU, MR
    AMOLS, HI
    DICELLO, JF
    HOWARD, J
    LYMAN, JT
    KOEHLER, AM
    GRAVES, R
    SMATHERS, JB
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1978, 51 (609): : 699 - 703
  • [4] Impact of photon cross section uncertainties on Monte Carlo-determined depth-dose distributions
    Aguirre, E.
    David, M.
    deAlmeida, C. E.
    Bernal, M. A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (09): : 1065 - 1071
  • [5] MONTE-CARLO CALCULATIONS OF MONOENERGETIC ELECTRON DEPTH-DOSE DISTRIBUTIONS IN LIF CHIPS - SKIN DOSE CORRECTION FACTORS FOR BETA-RAYS
    HOROWITZ, YS
    HIRNING, CR
    YUEN, P
    WONG, P
    [J]. HEALTH PHYSICS, 1994, 67 (04): : 330 - 337
  • [6] DEPTH-DOSE DISTRIBUTION MEASUREMENTS AND CALCULATIONS IN AN ELLIPTICAL PHANTOM
    PALFALVI, J
    KOBLINGER, L
    SZABO, PP
    [J]. HEALTH PHYSICS, 1983, 44 (01): : 35 - 44
  • [7] MONTE-CARLO METHOD FOR DEPTH-DOSE DISTRIBUTIONS FROM DIAGNOSTIC X-RAY BEAMS
    GIBBS, SJ
    KING, RJ
    PUJOL, A
    [J]. RADIATION RESEARCH, 1976, 67 (03) : 615 - 615
  • [8] Monte Carlo Dosimetry of the 60Co BEBIG High Dose Rate for Brachytherapy
    Campos, Luciana Tourinho
    Veloso de Almeida, Carlos Eduardo
    [J]. PLOS ONE, 2015, 10 (09):
  • [9] Dose assessment for ingestion of a 330 kilobecquerel 60Co hot particle
    Whillans, D. W.
    Chase, W. J.
    Wolodarsky, W. H.
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 127 (1-4) : 90 - 92
  • [10] Dose mapping of the 60Co gamma irradiation facility and a real irradiated product - Measurements and Monte Carlo simulation
    Majer, Marija
    Pasaricek, Luka
    Knezevic, Zeljka
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2024, 214