Amplitude Doppler US: Slow blood flow detection tested with a flow phantom

被引:53
|
作者
Weskott, HP
机构
[1] Department of Internal Medicine II, Städtisches Krankenhaus Siloah, Hannover 30449
关键词
blood; flow dynamics; blood vessels; US; ultrasound; (US); Doppler studies; experimental;
D O I
10.1148/radiology.202.1.8988201
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the in vitro sensitivity for the detection of slow blood flow in tiny tubes with an amplitude Doppler ultrasound (US) technique at different angles and pulse repetition frequencies. MATERIALS AND METHODS: Human blood treated with heparin was pumped through silicone tubes (flow phantoms made by the author) with lumina of 3.00 mm, 0.30 mm, and 0.05 mm. Slow flow was imaged by using amplitude and color Doppler US techniques with two US devices at different pulse repetition frequencies and angles and in different sound propagation media. RESULTS: Under ideal conditions, the lowest detectable velocity was 0.04-0.06 cm/sec in the 3.00-mm tube. Detection was about three to four times less sensitive in the 0.30-mm and 0.05-mm tubes. Amplitude Doppler US sensitivity is only slightly dependent on the angle of incidence and propagation medium. CONCLUSION: The fractional volume flow determined the sensitivity of the amplitude Doppler technique for detecting low flow. Amplitude Doppler US is nearly angle independent; thus, it is superior to color Doppler US. Amplitude Doppler US is three to four times more sensitive when tubes have very tiny lumina and when probes of a lower frequency are used.
引用
收藏
页码:125 / 130
页数:6
相关论文
共 50 条
  • [41] Manufacture of Flow Phantom with Stenosis and Imaging Evaluation of Power Doppler
    Park, Hee-Young
    Bae, Jong-Rim
    Kim, Jeong-Koo
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF KOREA, 2009, 28 (08): : 732 - 739
  • [42] DEVELOPMENT OF A FLOW PHANTOM FOR TRANSCRANIAL DOPPLER ULTRASOUND QUALITY ASSURANCE
    Alablani, Fatmah
    Janus, Justyna
    Pallett, Edward
    Mullins, Toni M.
    Almudayni, Alanoud
    Chung, Emma M. L.
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2022, 48 (11): : 2302 - 2309
  • [43] A VERSATILE PHANTOM FOR HEMODYNAMIC MEASUREMENTS BY ULTRASOUND DOPPLER FLOW INSTRUMENTS
    VONBOETTICHER, H
    DELEBINSKI, R
    RISCH, U
    LUSKA, G
    [J]. ULTRASCHALL IN DER MEDIZIN, 1994, 15 (05): : 264 - 268
  • [44] A ROTATING CYLINDER PHANTOM FOR FLOW AND TISSUE COLOR DOPPLER TESTING
    Walker, Andrew
    Henriksen, Egil
    Ringqvist, Ivar
    Ask, Per
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (11): : 1892 - 1898
  • [45] Blood Flow Imaging with Ultrafast Doppler
    Baranger, Jerome
    Mertens, Luc
    Villemain, Olivier
    [J]. JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2020, (164): : 1 - 13
  • [46] DOPPLER BLOOD-FLOW IN PREGNANCY
    MCPARLAND, P
    PEARCE, JM
    [J]. PLACENTA, 1988, 9 (04) : 427 - 450
  • [47] SOURCE OF DOPPLER EFFECT IN BLOOD FLOW
    Wang Weiqi
    [J]. Chinese Journal of Biomedical Engineering, 1996, (04) : 139 - 146
  • [48] DOPPLER ANALYSIS OF BRACHIAL BLOOD FLOW
    GONZALEZ, RR
    HILL, L
    SPENCER, MP
    [J]. FEDERATION PROCEEDINGS, 1967, 26 (02) : 549 - &
  • [49] Measuring blood flow dynamics using DOCT and Doppler amplitude optical coherence tomography (DAOCT)
    Moger, J
    Matcher, SJ
    Winlove, CP
    Shore, A
    [J]. COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE VII, 2003, 4956 : 366 - 373
  • [50] ECHOGENIC BLOOD DURING SLOW FLOW
    HAMILTON, G
    [J]. CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 1994, 45 (06): : 487 - 487